All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?

We, 31.12.2025
| Original article from: Phenomenex
How can neutral analytes be efficiently concentrated in the presence of ionized acidic or basic contaminants? This SPE tip explains how mixed-mode cartridges improve selectivity and cleanup.
<p><strong>Phenomenex:</strong> All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?</p>

Phenomenex: All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?

Welcome to our new sample prep technical tip. This month we examine how to extract neutral analytes with ionized interferences.

When extracting neutral analytes from an aqueous-based sample matrix (urine, plasma, etc.), it is easy to reach for a reversed phase cartridge (C18, SDBV polymer, etc.). 

The analytes are retained on the C18 chains via Van der Waals interactions. Polar interferences, such as salts, can be washed from the cartridge using water; less polar interferences can be removed using a small percentage of Methanol (typically around 10 %), again in water. Analytes are typically eluted using 100 % Methanol.

There will be occasions where your sample matrix contains acidic or basic impurities that have a relatively high hydrophobicity. Such interferences would be retained at loading, may not be removed during washing, and will probably elute along with your analytes of interest. They may also be found to elute in a similar elution window as your analytes of interest, potentially causing issues with method specificity.

Phenomenex: All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?Phenomenex: All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?

If you make use of a mixed mode cartridge instead of using a “straight reversed phase” cartridge, it is possible to remove the charged interferences during the SPE procedure. If we consider neutral compounds in the presence of a hydrophobic base, the first step would be to acidify the sample to ensure that the basic compound is protonated. This can be achieved by the addition of a strong acid such as Hydrochloric Acid (HCl).

The pH-adjusted sample can then be loaded onto a conditioned and equilibrated cartridge such as Strata™-X-C, shown above. The basic interference will be retained by coulombic attractions on the Sulfonic Acid group shown in red; the analytes of interest by Van der Waals interactions (as seen in a straight reversed phase extraction) through interactions with the rest of the polymer material shown in black.

The wash steps should also contain acid, such as dilute HCl, keeping the pH low. In this case, we would wash with 0.1 M HCl to remove polars, and 0.1 M HCl in 10 % Methanol to remove more hydrophobic impurities. The final elution would then be in Methanol. The hydrophobic base will remain inside the cartridge during the elution step.

A similar approach, using Strata-X-A can be utilized if the removal of acidic compounds is desired. In this case however, rather than using dilute HCl, we would want a buffer around pH 8 for sample adjustment and for using during the wash steps.

Phenomenex: All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?Phenomenex: All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?

We hope this tip was useful! Check out our resources on solid phase extraction. See you next month for another tip!

Phenomenex
LinkedIn Logo
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?
Article | Academy

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?
Phenomenex
tag
share
more
Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
I am getting low recovery in my SPE method, how do I fix the problem?
Article | Academy

I am getting low recovery in my SPE method, how do I fix the problem?

Practical troubleshooting guide for low recovery in SPE methods. Learn how to identify problems with solvents, flow rate, pH, or sorbent choice step by step.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?
Article | Academy

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?
Phenomenex
tag
share
more
Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
I am getting low recovery in my SPE method, how do I fix the problem?
Article | Academy

I am getting low recovery in my SPE method, how do I fix the problem?

Practical troubleshooting guide for low recovery in SPE methods. Learn how to identify problems with solvents, flow rate, pH, or sorbent choice step by step.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?
Article | Academy

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?
Phenomenex
tag
share
more
Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
I am getting low recovery in my SPE method, how do I fix the problem?
Article | Academy

I am getting low recovery in my SPE method, how do I fix the problem?

Practical troubleshooting guide for low recovery in SPE methods. Learn how to identify problems with solvents, flow rate, pH, or sorbent choice step by step.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?
Article | Academy

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?

How is normal phase (NP) SPE different from reversed phase (RP) SPE? How do I decide which one to use?
Phenomenex
tag
share
more
Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
I am getting low recovery in my SPE method, how do I fix the problem?
Article | Academy

I am getting low recovery in my SPE method, how do I fix the problem?

Practical troubleshooting guide for low recovery in SPE methods. Learn how to identify problems with solvents, flow rate, pH, or sorbent choice step by step.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike