USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler

- Photo: EST Analytical: USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler
- Video: EST Analytical: EVOLUTION 2 | EST Analytical
USEPA Method 524.2 calls for purge and trap sampling coupled with a Gas Chromatograph (GC) and Mass Spectrometer (MS) for separation and analysis. It has an extensive list of compounds and requires lower detection limits due to its use in drinking water testing. A 20% relative standard deviation for compound response factors is required across the calibration curve while the calibration check standard analytes have to be within 30% of the expected concentration. Currently, this method calls for a 5mL or 25mL purge volume and a four minute desorb time. These method requirements can pose a challenge for the analyst due to the amount of moisture that is created.
The negative effects of the increased water vapor include instability of the chromatography, band eluting and poor peak shape to name a few. A side effect of this instability is sample re-runs and lost laboratory productivity. In addition, a common practice for mitigating moisture is increasing the GC inlet split rate. This poses yet another difficulty as the increased split rate can result in reduced sensitivity. EST Analytical’s eight port valve in the Evolution purge and trap concentrator aids in limiting the water exposure because the water management of the system is not part of the desorb pathway. This feature also enables laboratories to use a smaller split ratio at the GC inlet thus increasing sensitivity.
Experimental:
The sampling system used for this study was the EST Analytical Evolution purge and trap concentrator and the Centurion WS autosampler. The concentrator was affixed with a Vocarb 3000 trap and connected to an Agilent 7890A GC and 5975C inert XL MS. The GC was configured with a Restek Rxi-624 Sil MS 30m x 0.25mm x 1.4µm column. The purge and trap and GC/MS parameters used for this study are listed below.
Purge and Trap Parameters
- Concentrator: EST Evolution
- Trap type: Vocarb 3000
- Valve oven temperature: 130 °C
- Transfer line temperature: 130 °C
- Trap temperature: 35 °C
- Moisture Reduction Trap (MoRT) temperature: 39 °C
- Purge time: 11 min
- Purge flow: 40 mL/min
- Dry purge Temp.: Ambient
- Dry Purge Flow: 40mL/min
- Dry Purge Time: 1.0 min
- Desorb pressure Control: ON
- Desorb Pressure: 5psi
- Desorb time: 4 min
- Desorb preheat delay: 15 sec
- Desorb temperature: 250 °C
- MoRT bake temperature: 210 °C
- Bake temperature: 260 °C
- Sparge vessel bake temperature: 110 °C
- Bake time: 6 min
- Bake flow: 70 mL/min
Auto-Sampler (EST Centurion WS):
- Sample type: Water
- Water volume: 25 mL
- Internal standard volume: 5 µL
GC/MS Parameters
- System: Agilent 7890A GC / 5975C inert XL MS
- Inlet: Split mode, 220 °C, 40:1 split ratio, 12.153 psi head pressure
- Column: Restek Rxi-624Sil MS, 30 m × 0.25 mm × 1.4 µm
- Oven program: 45 °C (1 min) → ramp 15 °C/min → 220 °C, hold 1.33 min (total run ~14 min)
- Column flow rate: 1 mL/min helium
- Total flow: 44 mL/min
- Source temperature: 230 °C
- Quadrupole temperature: 150 °C
- Transfer line temperature: 180 °C
- Scan range: m/z 35–300, 5.2 scans/sec
- Solvent delay: 0.7 min
A calibration curve was established with a linear range of 0.25 to 40µg/L using the USEPA Method 524.2 standards from Restek. After the curve was determined, method detection limits were ascertained by running a series of seven replicate 0.25 µg/L standards and calculated using the procedure outlined in 40 CFR Part 136 Appendix B. Finally, seven 20 µg/L standards were examined in order to verify the precision and accuracy of the analysis. The results of this study are displayed in Tables 1 and 2 while Figure 1 presents a 20 µg/L standard GC/MS chromatogram.
EST Analytical: Table 1 - Compound Linearity and Response Factor Table
EST Analytical: Table 2 - MDL and Precision and Percent Recovery Table
EST Analytical: Figure 1 - 20 µg/L Standard Chromatogram
Conclusions:
This study demonstrated the effectiveness of the Evolution/Centurion purge and trap sampling system. The results passed all of the USEPA Method 524.2 criteria for both linearity and detection limits. Since the Evolution bypasses the water management trap during the four minute desorb process, the amount of water effecting the experiments was limited. Furthermore, the forty to one split ratio at the GC inlet provided excellent detection limits as the system was more sensitive to the low level standard requirements of the method. The lower split rate has an added benefit of reducing laboratory helium consumption when setting it against other system’s higher split ratios. Finally, the precision and percent recovery results showed the reliability of the system in producing accurate experimental findings.
- USEPA Method 524.2, “Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry, Revision 4.1, 1995.




