Permanent gases and CO2 - Fast analysis of permanent gases and CO2 using coupled tandem PLOT columns
Permanent gases and CO
2
Fast analysis of permanent gases
and CO
2 using coupled tandem PLOT
columns
Application Note
Environmental
Authors
Agilent Technologies, Inc.
Introduction
A parallel setup of two PLOT columns is tuned for separation of permanent gases
in a short time. The sample is injected via a normal injection port and is split into
the parallel setup of two columns. In this application a short Agilent PoraBOND Q
is used to separate the CO
2 (and methane) from the permanent gases (first peak
composite) before the first peak (helium) elutes from the Agilent CP-Molsieve PLOT
column. After helium, all other permanent gases will be separated which include
helium, argon, oxygen, xenon, CO and methane. If water is present, it will appear on
the PoraBond and will elute after the CO
2 peak.
This analysis is done isothermally and requires a long CP-Molsieve column to
separate peakpairs argon-oxygen and helium-neon. The CO
2 and eventually water
that enters the Molsieve column will be adsorbed. If the amount of CO
2 or water
accumulated on the CP-Molsieve causes a shift of the retention time of the inert
gases out of the integration window, the Agilent Select Permanent Gases/CO
2 and
water adsorption has very little impact on the retention and many analysis can be
done before regeneration is required. As methane elutes from both systems the split
ratio between the columns can be calculated by the ratio of the methane peaks.
If heavier compounds are present, for instance ethane, ethylene, propane, these
components will elute later between the peaks that elute from the CP-Molsieve
column. If such a component interferes with a compound that elutes from the
Molsieve, the oven temperature must be changed by a few degrees.
www.agilent.com/chem
This information is subject to change without notice.
© Agilent Technologies, Inc. 2011
Printed in the USA
31 October, 2011
First published prior to 11 May, 2010
A02028
Conditions
Technique
: GC
Column
: Agilent Select Permanent Gases/CO
2
Part no. CP7429
Temperature
: 35 °C
Carrier Gas
: Helium, 100 kPa
Injector
: Split 50 mL/min
Detector
: TCD
Sample Size
: 20 μL
Concentration Range : % level
Courtesy
: C. Duvekot, Agilent Application Laboratory,
Middelburg, The Netherlands
Peak identification
oxygen
nitrogen
methane
carbon monoxide
carbon dioxide