Polycyclic aromatic hydrocarbons

Applications | 2011 | Agilent TechnologiesInstrumentation
GC/MSD, GC/IT, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Polycyclic aromatic hydrocarbons (PAHs) are persistent environmental pollutants with known carcinogenic and mutagenic properties. Reliable measurement of trace-level PAHs is essential for environmental monitoring, regulatory compliance, and protection of human health.
  • High environmental persistence and bioaccumulation potential
  • Stringent regulatory limits for water, soil, and air quality
  • Demand for sensitive, reproducible analytical methods

Objectives and Study Overview


This application note evaluates the performance of the Agilent FactorFour VF-Xms GC column for the separation and quantification of 16 priority PAHs at trace levels. Key aims include:
  • Assessing chromatographic resolution and peak identification
  • Evaluating column bleed and thermal stability up to 340 °C
  • Demonstrating sensitivity with an ion trap mass spectrometer

Methodology


The study employed a capillary gas chromatography–mass spectrometry (GC-MS) approach under the following conditions:
  • Injection: split mode at 275 °C, 1 µL sample in hexane
  • Carrier Gas: helium at 60 kPa
  • Temperature Program: 100 °C to 320 °C at 20 °C/min, hold 8 min
  • Analyte Load: ~3 ng per PAH on column

Instrumentation


  • GC Column: Agilent FactorFour VF-Xms (30 m × 0.25 mm i.d., 0.10 µm film, Part No. CP8805)
  • Detector: ion trap mass spectrometer
  • Data Courtesy: J. Peene, Agilent R&D Laboratories, Middelburg, The Netherlands

Main Results and Discussion


The VF-Xms column demonstrated exceptional thermal stability up to at least 340 °C, yielding very low bleed and a stable baseline. Its higher arylene content imparted slightly greater polarity compared to VF-5ms, enhancing separation of isomeric PAHs. The phase fragmentation produced a low-intensity m/z 281 fragment, facilitating precise quantitation of heavier PAHs. All 16 target compounds, ranging from naphthalene to benzo(g,h,i)perylene, were baseline-separated and detected at approximately 3 ng levels.

Benefits and Practical Applications


  • Trace-level quantitation of PAHs in environmental samples (water, soil, air)
  • Reduced MS contamination and maintenance due to low column bleed
  • Fast method stabilization and high sample throughput enabled by robust temperature programming

Future Trends and Potential Applications


Ongoing developments in stationary-phase chemistry and surface treatments promise further improvements in thermal resilience and inertness for ultratrace analyses. Coupling low-bleed columns with high-resolution MS and automated sample preparation will expand applications in environmental monitoring, food safety testing, and forensic investigations.

Conclusion


The Agilent FactorFour VF-Xms column offers a reliable, low-bleed solution for GC-MS analysis of PAHs, delivering high thermal stability, enhanced polarity, and sensitive detection at trace levels. Its performance supports rigorous environmental assessment and compliance with regulatory standards.

References


No literature references were provided in the original document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Polycyclic aromatic hydrocarbons
Polycyclic aromatic hydrocarbons
2011|Agilent Technologies|Applications
Polycyclic aromatic hydrocarbons Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC with an Agilent VF-Xms column separates 13 polycyclic aromatic hydrocarbons in 30 minutes. The FactorFour VF-Xms bonded phase has an optimized stabilization structure. Combined with fused silica surface…
Key words
stabilization, stabilizationpolycyclic, polycycliccourtesy, courtesyfused, fusednetherlands, netherlandsaromatic, aromatictemperature, temperatureleast, leasthydrocarbons, hydrocarbonssilica, silicatreatment, treatmentprinted, printedcontamination, contaminationinjector, injectorcombined
PAHs - Analysis of EPA 610 polyaromatic hydrocarbons
PAHs - Analysis of EPA 610 polyaromatic hydrocarbons
2011|Agilent Technologies|Applications
PAHs Analysis of EPA 610 polyaromatic hydrocarbons Application Note Environmental Authors Introduction Agilent Technologies, Inc. For the analysis of trace PAH often the on-column injection technique is used because in hot injection ports discrimination, adsorption or decomposition can occur. Best…
Key words
injector, injectorenvironmental, environmentalaccessed, accessedtechnique, techniquecarrier, carriercolumn, columnpah, pahpahs, pahsdetector, detectorinside, insideinjection, injectiontemperature, temperaturediameter, diameterhydrocarbons, hydrocarbonsgas
Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms
Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms Application Note Author Introduction Laura Provoost Agilent Technologies, Inc. In multi-residue analysis different groups of compounds are separated in a single operation, saving time and increasing productivity. This…
Key words
heptachlor, heptachlorfluoranthene, fluorantheneepoxide, epoxidehch, hchbenzo, benzotrans, transxms, xmscis, cisminutes, minutesendosulfan, endosulfanresidue, residuechlordane, chlordanepcbs, pcbsendrin, endrinpyrene
Polyaromatic hydrocarbons
Polyaromatic hydrocarbons
2011|Agilent Technologies|Applications
Polyaromatic hydrocarbons Application Note Environmental Authors Introduction Agilent Technologies, Inc. Determination of 18 polyaromatic hydrocarbons using Agilent VF-35ms and GC/MS. Conditions Technique : GC-capillary Column : Agilent FactorFour VF-35ms, 0.25 mm x 30 m fused silica (df = 0.25 μm)…
Key words
polyaromatic, polyaromaticcourtesy, courtesyconc, conchydrocarbons, hydrocarbonsprinted, printedinjector, injectorcomponent, componentusa, usaauthors, authorsenvironmental, environmentaltechnique, techniquecarrier, carrieridentification, identificationsolvent, solventdetector
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike