Oxygenates - Analysis of trace methanol in 1,3-butadiene
Applications | 2011 | Agilent TechnologiesInstrumentation
Trace analysis of methanol in 1,3-butadiene is critical for ensuring product purity and preventing catalyst poisoning in downstream processes. Methanol, even at concentrations as low as a few ppm, can compromise polymerization efficiency and affect the quality of synthetic rubbers and plastics.
This application note demonstrates a robust gas chromatography method for quantifying methanol at low ppm levels in 1,3-butadiene. The goal is to achieve reliable separation from co-eluting hydrocarbons, particularly 4-vinyl-1-cyclohexene, and to maintain quantification reproducibility within 5%.
The analysis employs a wide-bore GC configuration with the following key components:
The Lowox column provides high retention of oxygenates, allowing methanol to elute after n-C14 and well ahead of 4-vinyl-1-cyclohexene. Chromatograms show clear baseline separation, supporting quantification down to 5 ppm with better than 5% reproducibility. The column’s high thermal stability (up to 350 °C) and low bleed ensure reliable performance across a range of oxygenated C1–C5 compounds.
Advances may include coupling the Lowox stationary phase with fast temperature programming or multidimensional GC to push detection limits into the sub-ppm and ppb range. Integration with automated sample handling and advanced data processing will further streamline routine trace oxygenate screening.
This GC–MS method using the Agilent Lowox column delivers reliable, low-level methanol quantification in 1,3-butadiene with excellent reproducibility and selectivity. Its robustness and adaptability make it an effective tool for quality assurance in hydrocarbon processing.
GC/MSD, GC columns, Consumables
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Trace analysis of methanol in 1,3-butadiene is critical for ensuring product purity and preventing catalyst poisoning in downstream processes. Methanol, even at concentrations as low as a few ppm, can compromise polymerization efficiency and affect the quality of synthetic rubbers and plastics.
Study Objectives and Overview
This application note demonstrates a robust gas chromatography method for quantifying methanol at low ppm levels in 1,3-butadiene. The goal is to achieve reliable separation from co-eluting hydrocarbons, particularly 4-vinyl-1-cyclohexene, and to maintain quantification reproducibility within 5%.
Methodology and Instrumentation
The analysis employs a wide-bore GC configuration with the following key components:
- Column: Agilent Lowox 0.53 mm fused silica PLOT (CP8587)
- Temperature Program: 175 °C hold for 2 min, ramp at 10 °C/min to 275 °C
- Carrier Gas: Helium at 420 kPa (4.2 bar)
- Injection: Split via Valco valve, 0.1 μL of commercial 1,3-butadiene
- Detection: Mass selective detector (MSD)
- Optional Pre-column: CP-Sil 5 CB for trace-level (ppb) pre-separation
Results and Discussion
The Lowox column provides high retention of oxygenates, allowing methanol to elute after n-C14 and well ahead of 4-vinyl-1-cyclohexene. Chromatograms show clear baseline separation, supporting quantification down to 5 ppm with better than 5% reproducibility. The column’s high thermal stability (up to 350 °C) and low bleed ensure reliable performance across a range of oxygenated C1–C5 compounds.
Benefits and Practical Applications
- Accurate trace methanol measurement enhances quality control in petrochemical and polymer industries.
- Broad applicability to ethylene, propylene, and other hydrocarbon streams.
- High selectivity reduces interference from co-eluting species without extensive sample preparation.
Future Trends and Possibilities
Advances may include coupling the Lowox stationary phase with fast temperature programming or multidimensional GC to push detection limits into the sub-ppm and ppb range. Integration with automated sample handling and advanced data processing will further streamline routine trace oxygenate screening.
Conclusion
This GC–MS method using the Agilent Lowox column delivers reliable, low-level methanol quantification in 1,3-butadiene with excellent reproducibility and selectivity. Its robustness and adaptability make it an effective tool for quality assurance in hydrocarbon processing.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of trace methanol in propylene
2011|Agilent Technologies|Applications
Oxygenates Analysis of trace methanol in propylene Application Note Energy & Fuels Authors Introduction Agilent Technologies, Inc. The Agilent Lowox adsorbent provides very high retention for oxygenated compounds. The methanol elutes after n-C14 allowing this component to be measured at…
Key words
lowox, lowoxoxygenated, oxygenatedoxygenates, oxygenatesadsorbent, adsorbentpropylene, propylenefuels, fuelsprinted, printedinjector, injectorusa, usamethanol, methanolenergy, energyauthors, authorstrace, tracesize, sizevery
Oxygenates - Separation of oxygenates in a C1-C5 hydrocarbon matrix
2011|Agilent Technologies|Applications
Oxygenates Separation of oxygenates in a C1-C5 hydrocarbon matrix Application Note Energy & Fuels Authors Introduction Agilent Technologies, Inc. For the separation of oxygenates in light hydrocarbons a highly polar stationary phase has to be used. TCEP is used widely…
Key words
ether, etheroxygenates, oxygenatestert, terthydrocarbon, hydrocarbonbutyl, butylhydrocarbons, hydrocarbonspropionaldehyde, propionaldehydelowox, lowoxpropanal, propanaloxygenated, oxygenatedisobutanol, isobutanoldiisopropyl, diisopropylsize, sizestreams, streamsadsorbent
PLOT Columns Separation Solutions for Light Hydrocarbons & Gases
2021|Agilent Technologies|Presentations
PLOT Columns Separation Solutions for Light Hydrocarbons & Gases Porous Layer Open Tubulars Johan Kuipers Training & Development Sept 2021 1 Webinar Agilent PLOT Columns DE44454.2362731481 Table of Boiling Point Fractions Carbon # C1 C2 C3 C4 C5 C6 C7…
Key words
plot, plotwebinar, webinarcolumns, columnsagilent, agilentporous, porousoxygenates, oxygenatesalumina, aluminaalcohol, alcoholpolymer, polymerparticle, particlehydrocarbons, hydrocarbonspolar, polarether, etherselectivity, selectivitycolumn
Solutions for light hydrocarbons and gasses: PLOT columns
2019|Agilent Technologies|Presentations
Solutions for light hydrocarbons and gasses: PLOT columns 1 April 17, 2019 Separation Superhero Seminar Tour Solutions for Light Hydrocarbons and Gasses: PLOT Columns Allen K. Vickers Agilent Technologies Folsom Technology Center California, USA April 18, 2019 2 Table of…
Key words
superhero, superherotour, tourseminar, seminarseparation, separationplot, plotporous, porousoxygenates, oxygenatesparticle, particlecolumns, columnscolumn, columnalcohol, alcoholalumina, aluminahydrocarbons, hydrocarbonsgas, gasether