Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking
Technical notes | 1998 | Agilent TechnologiesInstrumentation
The accurate timing of peak elution is critical in gas chromatographic screening of complex samples, especially for pesticide residues and suspected endocrine disruptors. Method translation combined with retention time locking (RTL) enables laboratories to adapt a single capillary GC method across different detectors, instruments, speeds, and column formats while preserving retention order and minimizing revalidation effort.
This work illustrates how Agilent’s method translation software and RTL tools were used to:
The goal is to extend the utility of an existing 567-compound pesticide screening method, reducing development time and ensuring precise retention time matching for library searching and confirmation.
Method translation calculates new inlet pressure and oven ramp settings for a desired speed gain based on original column dimensions. RTL then uses calibration runs of a reference compound at several pressures to generate a retention time versus pressure curve. The steps are:
Analyses were performed on Agilent 6890 Series GC systems in constant pressure mode with:
Scaling from GC-AED to GC-MS (scale factor 1) achieved retention time matching within ±0.1 minute for a 30 minute run. Predicted inlet pressures (17.93 psi) agreed with measured pressures after RTL calibration to within 0.03 psi (0.006 minute RT shift).
Increasing speed threefold on the same GC-AED platform reduced run times by a factor of three with acceptable resolution loss. RTL calibration pressures spanned only a 0.11 psi range (0.003 minute RT) demonstrating the method’s robustness.
Adapting the method to a 0.1 mm id micro-ECD column with 3× speed initially showed larger deviations due to assumed column length. Iterative adjustment of the column length in the translation software (from 10 m to 10.56 m) restored RT matching to within 0.03 minute, confirming the importance of accurate column dimensions for translations.
• Laboratories can reuse validated GC methods on different systems without full redevelopments
• Pesticide libraries remain consistent in retention time matching, streamlining library searches and reducing false positives
• Speed gains increase sample throughput while maintaining sufficient chromatographic separation for selective detectors
• Method translation and RTL integration yield predictable performance, improving method transfer between sites
• Integration of real-time method translation into instrument control software for automated scaling
• Expansion to new high-speed and low-thermal‐mass GCs enabling even faster analyses
• Application to emerging detectors and multidimensional GC methods
• Use of machine learning to predict optimal scale factors and calibration settings from historical data
Combining method translation with retention time locking offers a reliable framework for precise time scaling of capillary GC methods across instruments, detectors, and operational speeds. This approach maximizes the return on investment in method development, supports consistent library searching, and enables flexible adaptation to diverse analytical needs.
GC
IndustriesManufacturerAgilent Technologies
Summary
Significance of the Topic
The accurate timing of peak elution is critical in gas chromatographic screening of complex samples, especially for pesticide residues and suspected endocrine disruptors. Method translation combined with retention time locking (RTL) enables laboratories to adapt a single capillary GC method across different detectors, instruments, speeds, and column formats while preserving retention order and minimizing revalidation effort.
Objectives and Overview of the Study
This work illustrates how Agilent’s method translation software and RTL tools were used to:
- Scale the HP RTL Pesticide Library method between GC-AED and GC-MS with a scale factor of 1
- Increase analysis speed threefold on the same GC-AED setup
- Adapt the method to a high-speed 0.1 mm id micro-ECD column at constant pressure
The goal is to extend the utility of an existing 567-compound pesticide screening method, reducing development time and ensuring precise retention time matching for library searching and confirmation.
Methodology
Method translation calculates new inlet pressure and oven ramp settings for a desired speed gain based on original column dimensions. RTL then uses calibration runs of a reference compound at several pressures to generate a retention time versus pressure curve. The steps are:
- Define the scale factor (speed gain) in the translation software and enter original method parameters
- Calculate translated inlet pressures and oven ramps for constant-pressure operation
- Perform or predict RTL calibration runs to obtain locking pressures against target retention times
- Export and scale the original retention time table by the factor
- Import the new table and validate the scaled method with a standard mixture
Instrumentation
Analyses were performed on Agilent 6890 Series GC systems in constant pressure mode with:
- GC-AED detector (G2350A) controlled by GC-AED ChemStation
- 5973 MSD for GC-MS confirmation
- Micro-ECD detector for split 100:1 injections on a 0.1 mm id column
- HP-5MS capillary columns: 30 m×0.25 mm×0.25 µm and 10 m×0.1 mm×0.1 µm
- RTL software modules G2080AA and G2081AA
Main Results and Discussion
Scaling from GC-AED to GC-MS (scale factor 1) achieved retention time matching within ±0.1 minute for a 30 minute run. Predicted inlet pressures (17.93 psi) agreed with measured pressures after RTL calibration to within 0.03 psi (0.006 minute RT shift).
Increasing speed threefold on the same GC-AED platform reduced run times by a factor of three with acceptable resolution loss. RTL calibration pressures spanned only a 0.11 psi range (0.003 minute RT) demonstrating the method’s robustness.
Adapting the method to a 0.1 mm id micro-ECD column with 3× speed initially showed larger deviations due to assumed column length. Iterative adjustment of the column length in the translation software (from 10 m to 10.56 m) restored RT matching to within 0.03 minute, confirming the importance of accurate column dimensions for translations.
Benefits and Practical Applications
• Laboratories can reuse validated GC methods on different systems without full redevelopments
• Pesticide libraries remain consistent in retention time matching, streamlining library searches and reducing false positives
• Speed gains increase sample throughput while maintaining sufficient chromatographic separation for selective detectors
• Method translation and RTL integration yield predictable performance, improving method transfer between sites
Future Trends and Opportunities
• Integration of real-time method translation into instrument control software for automated scaling
• Expansion to new high-speed and low-thermal‐mass GCs enabling even faster analyses
• Application to emerging detectors and multidimensional GC methods
• Use of machine learning to predict optimal scale factors and calibration settings from historical data
Conclusion
Combining method translation with retention time locking offers a reliable framework for precise time scaling of capillary GC methods across instruments, detectors, and operational speeds. This approach maximizes the return on investment in method development, supports consistent library searching, and enables flexible adaptation to diverse analytical needs.
References
- Wylie PL, Quimby BD. A Method Used to Screen for 567 Pesticides and Suspected Endocrine Disrupters. Hewlett-Packard Application Note 228-402, 1998.
- Klee M, Giarrocco V. Predictable Translation of Capillary GC Methods for Fast GC. Hewlett-Packard Application Note 228-373, 1997.
- Giarrocco V, Quimby BD, Klee MS. Retention Time Locking: Concepts and Applications. Hewlett-Packard Application Note 228-392, 1997.
- Capillary Column Method Translator. Agilent user-contributed software. www.hp.com/go/mts.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking
2006|Agilent Technologies|Applications
Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking Application Gas Chromatography May 1998 Authors Key Words B. D. Quimby, L. M. Blumberg, M. S. Klee, and P. L. Wylie Agilent Technologies, Inc. 2850 Centerville Road…
Key words
rtl, rtlaed, aedpressure, pressurelocking, lockingmethod, methodpsi, psiretention, retentionpoints, pointscalculated, calculatedtranslation, translationret, retcalibration, calibrationtime, timepressures, pressuresscaled
Excellent choices for environmental applications
2009|Agilent Technologies|Guides
Solutions that meet your demands for: speed accuracy productivity Excellent choices for environmental applications Productivity Tools Applications > Return to Table of Contents > Search entire document Integrated Second Peristaltic Pump for Improved Sample Throughput and Reduced Matrix Effects Technical…
Key words
rtl, rtlpressure, pressuretime, timecolumn, columnaed, aedmethod, methodlocking, lockingagilent, agilentrtlock, rtlockretention, retentionicp, icpinlet, inletcan, canflow, flowmsd
Retention Time Locking of Organochlorine Pesticides on an Agilent 8860 GC System Using the OpenLab Retention Time Locking Wizard
2019|Agilent Technologies|Applications
Application Note Food Testing Retention Time Locking of Organochlorine Pesticides on an Agilent 8860 GC System Using the OpenLab Retention Time Locking Wizard Author Ian Eisele Agilent Technologies, Inc. Abstract The retention time locking (RTL) wizard (a feature of Agilent…
Key words
locking, lockingbhc, bhcwizard, wizardrtl, rtlretention, retentionrelocked, relockedafter, afteropenlab, openlabpremaintenance, premaintenanceheptachlor, heptachlortrimming, trimminginlet, inletepoxide, epoxidetime, timeendrin
Excellent choices for environmental applications - Water
2009|Agilent Technologies|Guides
Solutions that meet your demands for: speed accuracy productivity Excellent choices for environmental applications Water Volatiles Applications > Return to Table of Contents > Search entire document Ultra-Fast Total Petroleum Hydrocarbons (TPH) Analysis with Agilent Low Thermal Mass (LTM) GC…
Key words
icp, icpwater, wateragilent, agilenttime, timesample, samplewere, weremsd, msdcolumn, columnmass, massanalysis, analysisgas, gasusa, usamethod, methodion, ionmode