GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Applications | 2013 | Agilent TechnologiesInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Industries
Environmental, Clinical Research
Manufacturer
Agilent Technologies

Summary

Significance of Topic


Polychlorinated biphenyl mixtures (Aroclors) are persistent organic pollutants with strict regulatory limits and challenging analytical requirements. Traditional gas chromatography with electron capture detection (GC/ECD) often fails to distinguish individual Aroclors in complex matrices due to coeluting interferences. Developing a sensitive, selective, and rapid approach is crucial for environmental monitoring, food safety, and toxicological studies.

Objectives and Study Overview


This application note presents a novel method enabling simultaneous identification and quantitation of individual Aroclors in a single analytical run without prior cleanup. The approach leverages MS/MS pattern matching and combined calibration curves, based on guidelines from EPA Methods 8082 and 8270. Validation was demonstrated using two customer-provided samples: an environmental soil extract and human blood extracts.

Methodology and Instrumentation


The analysis was performed on an Agilent 7000B Triple Quadrupole GC/MS coupled to an Agilent 7890A Gas Chromatograph. Key parameters included:
  • Column: Agilent HP-5MS UI (30 m × 0.25 mm, 0.25 µm film)
  • Inlet: Agilent MMI splitless liner without glass wool at 250 °C
  • Carrier gas: Helium at 1 mL/min
  • Oven program: 70 °C (1.5 min), ramp 16 °C/min to 200 °C (1.5 min)
  • MS/MS mode: Electron ionization with multiple reaction monitoring (MRM), collision gas flow 2.5 mL/min
The method employed ten Aroclor standards (1016–1268) at known concentrations in isooctane, with acenaphthene-d10 as internal standard. Serial dilutions generated calibration points at 25, 50, 100, 200, and 400 ng/mL. Unique MRM transitions were selected for each homologue class, and a combined calibration curve table stored qualifier/quantifier ion ratios and acceptable ranges.

Main Results and Discussion


Comparison with GC/ECD on a spiked environmental sample showed unresolvable interferences in the ECD chromatogram. In contrast, the GC/MS/MS method clearly identified Aroclor 1248 at 1.407 ppb by matching four characteristic transitions and their qualifier ratios, despite matrix effects. In spiked blood extracts containing Aroclors 1254 and 1260 at 5 and 50 ppb, the method delivered measured concentrations of 4.24 and 43.14 ppb for Aroclor 1254, and 4.89 and 48.98 ppb for Aroclor 1260. Coeluting peaks were resolved by selecting unique congener transitions and evaluating qualifier ion ratios, with expert review confirming positive identifications.

Benefits and Practical Applications


  • One-run analysis without sample cleanup streamlines workflow.
  • Enhanced selectivity and sensitivity in complex matrices at low ppb levels.
  • Reliable pattern recognition using combined calibration tables reduces false positives.
  • Applicability to environmental, biological, and industrial quality-control laboratories.

Future Trends and Opportunities


Advancements may include integration of machine-learning algorithms for automated pattern recognition and qualifier ratio assessment, expansion to additional persistent organic pollutants, and further miniaturization of triple-quadrupole platforms. Software enhancements could enable dynamic calibration updates and remote data processing for high-throughput monitoring networks.

Conclusion


The Agilent 7000B Triple Quadrupole GC/MS method in MS/MS mode offers a powerful solution for rapid, accurate, and sensitive identification and quantitation of PCB Aroclor mixtures in a single run. By combining tailored MRM transitions with a unified calibration table and expert data interpretation, this approach overcomes the limitations of GC/ECD and streamlines analysis of challenging environmental and biological samples.

Reference


1. Frame M., Cochran J.W., Bøwadt S.S. Complete PCB congener distributions for 17 Aroclor mixtures determined by three HRGC systems optimized for comprehensive, quantitative, congener-specific analysis. J. High Res. Chromatog. 19, 657–668 (1996).

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Determination of Polychlorinated Biphenyls (PCBs) in Soils and Solid Waste by Accelerated Solvent Extraction and GC-MS/MS
CONSULENZE AMBIENTALI SPA Determination of Polychlorinated Biphenyls (PCBs) in Soils and Solid Waste by Accelerated Solvent Extraction and GC-MS/MS Fabrizio Galbiati, Thermo Fisher Scientific (Schweiz) AG, Reinach, Switzerland Luca Teli, Consulenze Ambientali SpA, Scanzorosciate, Italy Custom e r A ppli…
Key words
extraction, extractionaccelerated, acceleratedsolvent, solventpcb, pcbase, asepcbs, pcbsbiphenyls, biphenylspolychlorinated, polychlorinatedheptachlorobiphenyls, heptachlorobiphenylshexachlorobiphenyls, hexachlorobiphenylspentachlorobiphenyls, pentachlorobiphenylstetrachlorobiphenyls, tetrachlorobiphenylscleanup, cleanupfish, fishconsulenze
Analysis of Aroclor and Organochlorine Pesticides in Ground Water using Triple Quadrupole GC-MS/MS
C146-E389 Technical Report Analysis of Aroclor and Organochlorine Pesticides in Ground Water using Triple Quadrupole GC-MS/MS Riki Kitano1, Yuka Fujito1, Ruth Marfil-Vega1 A b s tra c t: USEPA method 608 is GC-ECD method to determine Aroclor and organochlorine pesticides.…
Key words
nonachlorobiphenyl, nonachlorobiphenylhexachlorobiphenyl, hexachlorobiphenylbhc, bhcmonochlorobiphenyl, monochlorobiphenyldichlorobiphenyl, dichlorobiphenyloctachlorobiphenyl, octachlorobiphenyltetrachlorobiphenyl, tetrachlorobiphenyltrichlorobiphenyl, trichlorobiphenylheptachlorobiphenyl, heptachlorobiphenylpentachlorobiphenyl, pentachlorobiphenylsed, sedendosulfan, endosulfanratio, ratioheptachlor, heptachloraroclor
Determination of Polychlorinated Biphenyl Congeners in Foodstuffs and Animal Feed using a Triple Quadrupole GC-MS/MS Instrument
Determination of Polychlorinated Biphenyl Congeners in Foodstuffs and Animal Feed using a Triple Quadrupole GC-MS/MS Instrument Chris Sandy, Agilent Technologies Ltd., 610 Wharfedale Road, IQ Winnersh, Wokingham, Berkshire RG41 5TP, UK Peter Fürst, Thorsten Bernsmann, Dominik Baumeister, CVUA-MEL, Joseph-König-Strasse 40,…
Key words
congeners, congenersfoodstuffs, foodstuffspcb, pcbanimal, animalfeed, feedndl, ndlhrms, hrmsortho, orthodwell, dwellpcbs, pcbsproduct, producttef, tefexpressed, expressedprecursor, precursorpercentage
Determination of Dioxin-Like and Non-Dioxin-Like Polychlorinated Biphenyl Congeners in Foodstuffs and Animal Feed Using the Agilent 7000 Triple Quadrupole GC/MS System
Determination of Dioxin-Like and Non-Dioxin-Like Polychlorinated Biphenyl Congeners in Foodstuffs and Animal Feed Using the Agilent 7000 Triple Quadrupole GC/MS System Application Note Food Safety Authors Abstract Prof. Dr. Peter Fürst Two methods have been developed on the Agilent 7000…
Key words
pcb, pcbcongeners, congenersfoodstuffs, foodstuffsortho, orthoanimal, animalfeed, feedndl, ndlmeat, meatpolychlorinated, polychlorinatedmono, monopcdd, pcddtef, tefdioxin, dioxindioxins, dioxinsproducts
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike