Solid Phase Microextraction of Volatile Compounds
Applications | 1998 | MerckInstrumentation
The development of solid phase microextraction (SPME) has transformed the analysis of volatile organic compounds (VOCs) by offering a rapid, solvent-free, and cost-effective alternative to traditional purge-and-trap and liquid extraction methods. Its compatibility with GC-MS and ability to achieve regulatory detection limits make SPME a pivotal tool in environmental monitoring and quality control.
This work evaluates the use of a 100 µm polydimethylsiloxane (PDMS) SPME fiber to extract the full suite of VOCs defined by US EPA Method 524.2 in drinking water. Key aims include comparing headspace versus direct immersion sampling, assessing chromatographic resolution, and exploring the applicability of SPME for both trace-level and high-concentration analyses.
Samples containing 50 ppb of each VOC were prepared in 20 mL water vials. The PDMS fiber was immersed for 5 minutes, then thermally desorbed at 220 °C for 6 minutes in a split/splitless inlet. Separation employed a 60 m × 0.25 mm ID, 1.5 µm film VOCOL capillary column with helium carrier at 2 mL/min, and detection by mass spectrometry scanning m/z 35–260 at 0.6 s per scan.
SPME using PDMS fibers offers a versatile, efficient, and sensitive approach for VOC analysis in water, meeting regulatory requirements while reducing solvent use and sample preparation time. Its adaptability to a range of concentrations and straightforward integration with existing GC-MS systems position it as a preferred technique for environmental monitoring.
GC/MSD, SPME, Consumables
IndustriesEnvironmental
ManufacturerMerck
Summary
Significance of the Topic
The development of solid phase microextraction (SPME) has transformed the analysis of volatile organic compounds (VOCs) by offering a rapid, solvent-free, and cost-effective alternative to traditional purge-and-trap and liquid extraction methods. Its compatibility with GC-MS and ability to achieve regulatory detection limits make SPME a pivotal tool in environmental monitoring and quality control.
Objectives and Study Overview
This work evaluates the use of a 100 µm polydimethylsiloxane (PDMS) SPME fiber to extract the full suite of VOCs defined by US EPA Method 524.2 in drinking water. Key aims include comparing headspace versus direct immersion sampling, assessing chromatographic resolution, and exploring the applicability of SPME for both trace-level and high-concentration analyses.
Methodology and Instrumentation
Samples containing 50 ppb of each VOC were prepared in 20 mL water vials. The PDMS fiber was immersed for 5 minutes, then thermally desorbed at 220 °C for 6 minutes in a split/splitless inlet. Separation employed a 60 m × 0.25 mm ID, 1.5 µm film VOCOL capillary column with helium carrier at 2 mL/min, and detection by mass spectrometry scanning m/z 35–260 at 0.6 s per scan.
Main Results and Discussion
- Cryofree operation yielded well-resolved early-eluting peaks; cooling the column to 10 °C further sharpened these signals and lowered detection limits.
- Comparison of sampling modes showed immersion favored the lightest, most volatile species, while headspace sampling improved recovery of heavier aromatics and chlorinated benzenes, reflecting differences in water solubility and density.
- Detection limits depended on analyte polarity and fiber distribution constants: nonpolar compounds achieved low-ppt sensitivity, whereas highly polar analytes required higher concentrations for reliable detection.
- SPME demonstrated linear responses over several orders of magnitude, enabling quick screening of samples with contaminant levels from tens of ppt up to thousands of ppb, reducing the need for repeated GC-MS runs and instrument maintenance.
Benefits and Practical Applications
- Solventless extraction minimizes waste and operating costs.
- Streamlined sample preparation accelerates turnaround times in environmental and QA/QC laboratories.
- Wide dynamic range supports both drinking water compliance testing and preliminary screening of industrial wastewater.
- Integration with dual-detector setups (e.g., FID and ECD) enhances analytical flexibility and throughput.
Future Trends and Opportunities
- Design of advanced fiber coatings tailored for polar and semi-volatile compounds to broaden analyte coverage.
- Greater adoption of automated SPME platforms to improve reproducibility and laboratory efficiency.
- Integration of miniaturized, portable GC-SPME systems for field and on-site monitoring applications.
- Coupling with high-resolution and tandem mass spectrometry to enhance sensitivity and compound identification.
Conclusion
SPME using PDMS fibers offers a versatile, efficient, and sensitive approach for VOC analysis in water, meeting regulatory requirements while reducing solvent use and sample preparation time. Its adaptability to a range of concentrations and straightforward integration with existing GC-MS systems position it as a preferred technique for environmental monitoring.
Used Instrumentation
- SPME fiber: 100 µm PDMS coating
- Sampling: 5 min immersion
- Desorption: 6 min at 220 °C in split/splitless inlet
- GC column: 60 m × 0.25 mm ID, 1.5 µm VOCOL fused silica capillary
- Carrier gas: helium at 2 mL/min
- Detector: mass spectrometer scanning m/z 35–260 at 0.6 s per scan
References
- Arthur C.L. et al. LC/GC 10 (9): 656–661 (1992).
- Arthur C.L. et al. Environ. Sci. Technol. 26: 979–983 (1992).
- Arthur C.L. et al. J. High Res. Chromatogr. 15: 741–744 (1992).
- Arthur C.L. et al. Environmental Lab Dec ’92/Jan ’93, 10–14.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Solid Phase Microextraction: Theory and Optimization of Conditions
1998|Merck|Guides
Bulletin 923 Solid Phase Microextraction: Theory and Optimization of Conditions Solid phase microextraction, a simple, effective adsorption/ desorption technique, eliminates the need for solvents or complicated apparatus for concentrating volatile or nonvolatile compounds in liquid samples or headspace. SPME is…
Key words
spme, spmefiber, fibercoating, coatingholder, holderanalyte, analytepdms, pdmsdesorption, desorptionsampling, samplingsupelco, supelcohplc, hplcdescription, descriptionextraction, extractionpolar, polarnonpolar, nonpolarcarboxen
Fast Analysis of Volatile Organic Compounds by Solid Phase Microextraction/Capillary GC
1998|Merck|Applications
Application Note 56 Fast Analysis of Volatile Organic Compounds by Solid Phase Microextraction/Capillary GC Solid phase microextraction is a fast, solventless alternative to conventional sample extraction techniques. Because no solvent is injected, and the analytes are rapidly desorbed onto the…
Key words
vocol, vocolmicroextraction, microextractionspme, spmecolumn, columnfactors, factorsanalytes, analytesdesorbed, desorbedstirring, stirringaql, aqlsolid, solidresponse, responsechloride, chloridepurge, purgepolarity, polaritytrap
Air Sampling of VOCs by SPME for Analysis by Capillary GC
1998|Merck|Applications
Application Note 141 Air Sampling of VOCs by SPME for Analysis by Capillary GC The disposable SPME portable field sampler and the Carboxen/ PDMS fiber are an ideal combination for monitoring VOCs in air samples. The ability to collect air…
Key words
pdms, pdmsspme, spmefiber, fiberchloride, chloridecarboxen, carboxenfield, fieldvinyl, vinylvocs, vocsacrylonitrile, acrylonitriletetrachloroethene, tetrachloroethenetrichloroethene, trichloroetheneair, airportable, portablesampler, samplermethylene
Monitor Organic Volatile Impurities (OVIs) in Pharmaceutical Products, Using Solid Phase Microextraction/Capillary GC
1997|Merck|Applications
Application Note 62 Monitor Organic Volatile Impurities (OVIs) in Pharmaceutical Products, Using Solid Phase Microextraction/Capillary GC A 100µm polydimethylsiloxane-coated SPME fiber provided higher sensitivity for less polar analytes (i.e., organic volatile impurities) and a polyacrylate-coated fiber provided higher sensitivity for…
Key words
spme, spmevolatile, volatileheadspace, headspaceovis, ovisimmersion, immersionfiber, fiberimpurities, impuritiestrichloroethylene, trichloroethylenemicroextraction, microextractionorganic, organicmanual, manualsampling, samplingpharmaceutical, pharmaceuticalisopropanol, isopropanolaqx