Analyze Trace Amounts of Solvents in Water Using Packed Column GC
Guides | 1996 | MerckInstrumentation
Analysis of trace organic solvents in aqueous matrices is critical for environmental monitoring, quality control in industrial processes and regulatory compliance. Direct analysis by packed column gas chromatography offers a rapid approach without extensive sample preparation. The method addresses challenges such as water-induced column degradation and coelution, providing reliable detection at ppm and sub-ppm levels.
The study evaluated glass and stainless steel columns packed with 80/120 Carbopack B/3% SP-1500 for trace-level quantitation of common solvents in water. Key aims included establishing minimum detection limits, maximum sample capacity, column stability under repeated aqueous injections, and assessing potential memory effects. Recommendations to optimize the analytical protocol were also developed.
Packed columns with Carbopack B/3% SP-1500 permit direct aqueous injection for rapid trace solvent screening in environmental, pharmaceutical and QC laboratories. The inert packing ensures prolonged column life and reproducible results without derivatization or extraction. Simple conditioning procedures and inline gas purification support high sensitivity and minimal maintenance.
Advances in detector technology, such as micro-FID or mass spectrometric interfaces, could push detection limits below 0.1 ppm. Coupling packed column GC with automated sampling or on-line monitoring platforms may enable real-time water quality assessment. Expanded application to volatile organic compounds in complex matrices will further extend this method’s utility.
The packed column GC method using 80/120 Carbopack B/3% SP-1500 provides a robust, sensitive and straightforward approach for trace solvent analysis in water. Its high stability, linearity and low memory effects make it suitable for routine monitoring. With proper conditioning, sample handling and gas purification, reliable quantitation at ppm levels is readily achieved.
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerMerck
Summary
Significance of the Topic
Analysis of trace organic solvents in aqueous matrices is critical for environmental monitoring, quality control in industrial processes and regulatory compliance. Direct analysis by packed column gas chromatography offers a rapid approach without extensive sample preparation. The method addresses challenges such as water-induced column degradation and coelution, providing reliable detection at ppm and sub-ppm levels.
Objectives and Study Overview
The study evaluated glass and stainless steel columns packed with 80/120 Carbopack B/3% SP-1500 for trace-level quantitation of common solvents in water. Key aims included establishing minimum detection limits, maximum sample capacity, column stability under repeated aqueous injections, and assessing potential memory effects. Recommendations to optimize the analytical protocol were also developed.
Instrumental Setup
- Column packing: 80/120 Carbopack B/3% SP-1500
- Column formats: 2 m×2 mm ID TightSpec™ glass; 10 ft×1/8 in OD stainless steel
- Oven programming: 100 °C isothermal for trace tests; 70–235 °C at 4 °C/min for broader screening
- Carrier gas: Nitrogen at 20–24 mL/min
- Detectors: Flame ionization (FID) at 250 °C; Thermal conductivity (TCD) for comparison
- Injector temperature: 200 °C; Injection volume: 1–4 µL aqueous sample
- Calibration: Peak area vs concentration, R² = 0.999 for alcohols and ketones
- Gas purification: Inline OMI™ purifiers and Supelcarb™ traps to remove O₂, H₂O and hydrocarbons
Main Results and Discussion
- Detection Range: 0.9 ng to 65 µg on-column, limited by FID sensitivity (~2×10⁻¹⁴ A)
- Linearity and Precision: Excellent linear response for alcohols and ketones with 1–4 % RSD
- Column Stability: No loss of efficiency or shift in relative retention (MEK/isopropanol) after 200 aqueous injections
- Water Interference: FID suppressed the water peak, enabling clear methanol quantitation; TCD showed masking of low-MW peaks
- Memory Effects: No ghost peaks observed when switching from high-level or low-level solvent samples to water blanks
- Sample Volume Limits: Best retention reproducibility up to 2 µL on glass columns and 4 µL on stainless steel; larger volumes caused retention shifts
- Sample Handling: Avoid glass wool at the inlet for trace work; use deactivated liners, dedicated syringes and pre-conditioning with water injections to stabilize the baseline
Benefits and Practical Applications
Packed columns with Carbopack B/3% SP-1500 permit direct aqueous injection for rapid trace solvent screening in environmental, pharmaceutical and QC laboratories. The inert packing ensures prolonged column life and reproducible results without derivatization or extraction. Simple conditioning procedures and inline gas purification support high sensitivity and minimal maintenance.
Future Trends and Potential Applications
Advances in detector technology, such as micro-FID or mass spectrometric interfaces, could push detection limits below 0.1 ppm. Coupling packed column GC with automated sampling or on-line monitoring platforms may enable real-time water quality assessment. Expanded application to volatile organic compounds in complex matrices will further extend this method’s utility.
Conclusion
The packed column GC method using 80/120 Carbopack B/3% SP-1500 provides a robust, sensitive and straightforward approach for trace solvent analysis in water. Its high stability, linearity and low memory effects make it suitable for routine monitoring. With proper conditioning, sample handling and gas purification, reliable quantitation at ppm levels is readily achieved.
Reference
- Gill JM, Hartmann CH. J Gas Chromatogr. 1967;5:605.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Bulletin 824C Resolving Industrial Solvent Mixtures on Packed and Capillary GC Columns Nonpolar SPB-1 and polar SUPELCOWAX 10 capillary columns separate complex mixtures of most commonly used solvents with good resolution. Resolution on 0.53mm ID columns is similar to that…
Key words
acetate, acetatecellosolve, cellosolvelate, latebutyl, butylethyl, ethylisobutyl, isobutylalcohol, alcoholmethyl, methylxylene, xyleneketone, ketonepropyl, propylisopropyl, isopropyltoluene, toluenesolvents, solventsmesityl
Packed Column GC Application Guide
1999|Merck|Guides
Packed Column GC Application Guide Bulletin 890A Packed Column GC Application Guide This guide to packed column applications provides information on the best approach to designing your application. It includes helpful information about packings, phases, tubing, and different types of…
Key words
min, minfid, fidstock, stockacetate, acetateacid, acidalcohol, alcoholcellosolve, cellosolveavailable, availableethyl, ethylxylene, xylenemethyl, methylbutyl, butylethylene, ethylenecolumn, columnmethanol
Supelco™ Gas Purifiers for Total Protection of GC Systems
1998|Merck|Guides
Bulletin 848D Supelco™ Gas Purifiers for Total Protection of GC Systems Contaminants entering a carrier gas stream can adversely affect the performance of a GC system. To prevent damage and interference caused by water, oxygen, and other critical contaminants, purification…
Key words
purifier, purifiersupelco, supelcoremover, removeromi, omioxygen, oxygengas, gasmoisture, moisturesieve, sievecarrier, carriercapacity, capacitytube, tubeindicating, indicatingwater, watermolecular, molecularpurifiers
Petroleum/Chemicals Application Guide
2004|Merck|Guides
595 North Harrison Road Bellefonte, PA 16823-0048 USA Telephone 800-247-6628 ● 814-359-3441 Fax 800-447-3044 ● 814-359-3044 email: [email protected] sigma-aldrich.com/supelco Bulletin 858D Petroleum/Chemicals Application Guide This guide contains up-to-date information about Supelco products and technology for separating hydrocarbons by chromatographic methods.…
Key words
xylene, xylenehayesep, hayesepsupelco, supelcoscotty, scottymin, minfid, fidpetrocol, petrocolmix, mixmixes, mixestert, tertbenzene, benzenetoluene, toluenegasoline, gasolinebutane, butanehydrocarbons