The Future of Environmental Analysis by GC/MS: Combining New Deactivation Chemistries, Microfluidics, and Precision Pneumatics

Presentations | 2017 | Agilent TechnologiesInstrumentation
GC, GC/MSD
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Environmental trace analysis by GC/MS is critical for assessing pollutants in soil, water, and air. Matrix interferences and system contamination limit sensitivity and throughput, so innovations in flowpath design, surface inertness, and fluidic control can greatly enhance analytical performance and operational efficiency.

Objectives and Study Overview


This study examines the Agilent Intuvo 9000 GC/MS platform, focusing on its modular guard chip technology, microfluidic-enabled pneumatic control, and new deactivation chemistries. It aims to characterize guard chip trapping modes, demonstrate selective analyte retention and release strategies, and integrate guard chip pulsing with post-column backflush for improved environmental analyses.

Methodology and Instrumentation


Instrumentation
  • Agilent Intuvo 9000 GC with disposable, ferrule-free guard chips and flow chips
  • Intuvo ultra inert surface treatment and direct resistive heating
  • 6th-generation microfluidic-enabled electronic pneumatic control (EPC) modules
  • Mass spectrometer detector (MSD) at the column outlet

Methodological Steps
  • Evaluate guard chip trapping under isothermal (40 °C and 350 °C) and temperature-ramped conditions
  • Test rapid temperature pulsing (up to 300 °C/min) to selectively trap or release n-alkanes and polycyclic aromatic hydrocarbons (PAHs)
  • Combine guard chip pulsing with post-column backflush sequences to clear high-boiling fraction
  • Demonstrate pseudo heart-cutting by sequential low- and high-volatility fraction analysis

Main Results and Discussion


Guard Chip Trapping Modes
  • Low-temperature isothermal trapping retains both analytes and matrix, causing peak broadening.
  • Temperature ramping following the oven program improves peak shape and reproducibility.
  • Fast temperature pulses distinguish compounds by boiling point with ~30 °C resolution (~2–4 carbon units).

Selective Matrix Removal
  • Guard chip pulsing at tailored temperatures traps heavy matrix while passing target analytes.
  • Optimization of pulse endpoint and hold time minimizes carryover (<1%) for PAHs.

Backflush Integration
  • Rapid guard chip backflush completes in ~0.5 min vs. 3.5 min for traditional post-column methods—a 7× time improvement.
  • Matrix background is effectively reduced for both n-alkanes and PAHs.

Pseudo Heart-Cutting
  • Two sequential runs separate low-boiling (C10–C18) and high-boiling (C28–C40) fractions.
  • Precise temperature and pressure control enable selective trapping and transfer without hardware changes.

Benefits and Practical Applications


  • Enhanced system inertness reduces adsorption and contamination.
  • Automated, disposable guard chips streamline maintenance and minimize downtime.
  • Selective trapping and backflush accelerate analysis cycles and improve throughput.
  • Pseudo heart-cutting provides targeted fractionation without complex valve systems.

Future Trends and Opportunities


Further integration of microfluidic flow control could enable real-time adaptive trapping protocols based on analyte characteristics. Combining guard chip pulsing with multi-dimensional GC (GC×GC) and automated sample preparation in environmental laboratories may expand capabilities for ultratrace detection and high-throughput screening. Development of new surface chemistries and miniaturized detectors could further enhance selectivity and sensitivity.

Conclusion


The Intuvo 9000 GC with guard chip technology leverages precise thermal and pneumatic control to selectively trap, release, and backflush analytes and matrix components. Fast temperature pulsing and microfluidic EPC modules deliver improved peak shapes, reduced carryover, and significantly accelerated backflush cycles. Pseudo heart-cutting workflows illustrate potential for targeted fractionation without additional hardware, positioning this platform as a powerful tool for environmental trace analysis.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Characterization of a Temperature and Flow-Programmable Microfluidic Precolumn for Gas Chromatography
Application Note Environmental Characterization of a Temperature and Flow-Programmable Microfluidic Precolumn for Gas Chromatography Author Matthew Giardina, Ph.D. Agilent Technologies, Inc. Wilmington, USA Abstract This Application Note investigates the temperature control of Agilent Intuvo Guard Chips as a means of…
Key words
chip, chipguard, guardpulsing, pulsingbackflush, backflushtemperature, temperatureprogram, programmin, mintime, timeretention, retentionprograms, programsmicrofluidic, microfluidicoven, ovenexperiments, experimentsflow, flowprograming
Characterization of a Temperature and Flow Programmable Microfluidic Pre-Column for Gas Chromatography
Characterization of a Temperature and Flow Programmable Microfluidic Pre-Column for Gas Chromatography. Introduction Matthew Giardina, James D. McCurry, Rebecca A. Veeneman, Abbey Fausett, Joseph L. Hedrick 42nd ISCC / 15th GC×GC Agilent Technologies – Gas Phase Separations Division May 13-18,…
Key words
chip, chipguard, guardpulsing, pulsingbackflush, backflushtemperature, temperaturevolatilized, volatilizedpulsed, pulsedexperiments, experimentsflow, flowmatrix, matrixprogram, programmin, mincan, canbackflushing, backflushingbegin
Transferring Methods to Intuvo: Six Practical Examples
Transferring Methods to Intuvo: Six Practical Examples
2018|Agilent Technologies|Technical notes
Technical Overview Transferring Methods to Intuvo: Six Practical Examples Introduction This Technical Overview demonstrates different examples of method transfer from an Agilent 7890 GC system to an Agilent Intuvo 9000 GC system. More information on concepts related to transferring methods…
Key words
same, sameintuvo, intuvobus, busdefault, defaulttemperature, temperaturechip, chipendrin, endrindiesel, dieselguard, guardagilent, agilentoven, ovenbreakdown, breakdownsplitless, splitlesssplit, splitconventional
Trim Time Not Columns –Multiresidue Pesticides Analysis with an Intuvo GC/TQ
Trim Time Not Columns – Multiresidue Pesticides Analysis with an Intuvo GC/TQ Courtney Milner, Ph.D. Agilent Technologies Santa Clara, CA ASMS 2018 VIP Session 1 6/17/2018 Agenda Pesticide analysis trends Targeted vs. untargeted analysis What makes the Intuvo 9000 GC…
Key words
targeted, targetedscreening, screeningpesticides, pesticidesbackflush, backflushintuvo, intuvomrm, mrmtrim, trimdatabase, databasesystem, systempcdl, pcdlrobustness, robustnessdmrm, dmrmrepeatability, repeatabilitytea, teapesticide
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike