N2O/CO/CO2/CH4 analysis system (TCD) Nexis GC-2030NCCC3 GC-2014NCCC3
Applications | 2017 | ShimadzuInstrumentation
The accurate quantification of nitrous oxide and other permanent gases released from soil is essential for environmental monitoring, greenhouse gas inventory assessments and biogeochemical research. Multiplex gas chromatography systems equipped with both electron capture and thermal conductivity detectors enable simultaneous sensitive detection of N2O, CO, CO2, CH4, N2 and O2, improving analytical throughput and data reliability.
This application note describes a method for the determination of soil-emitted N2O using gas chromatography with electron capture detection and simultaneous analysis of permanent gases via thermal conductivity detection. The system utilizes multiple packed columns and sample loops to separate a complex gas mixture into key target components.
The analysis is performed on a Shimadzu Nexis GC-2030N (or GC-2014N) system configured with four switching valves, seven packed columns and two detectors (one ECD and one TCD). The flow path is divided into two channels:
Samples are introduced via two sample loops. The complementary use of ECD and TCD detectors covers trace to percent-level concentration ranges.
Typical chromatograms demonstrate baseline-resolved peaks for N2O (ECD) with a retention time around 4 min and well-separated peaks for CH4, N2, O2, CO2 and CO (TCD) between 1 and 9 min. The system achieves detection limits down to 50 ppb for N2O and 0.01% for permanent gases with excellent repeatability. Concentration ranges extend from trace to high-level percentages, accommodating diverse sample matrices.
Advances in micro-detector technology and miniaturized GC systems could further enhance field deployability. Integration with automated soil gas sampling modules and remote data acquisition platforms will enable real-time greenhouse gas flux monitoring. Coupling with mass spectrometry or laser-based detectors may improve compound identification and isotopic analysis capabilities.
This method demonstrates a versatile, high-throughput GC system for concurrent quantification of N2O and permanent gases from soil samples. The dual-detector approach and multi-column configuration deliver reliable, sensitive and reproducible results, meeting the demands of environmental monitoring and research.
No references provided in the original text.
GC
IndustriesEnergy & Chemicals
ManufacturerShimadzu
Summary
Importance of the topic
The accurate quantification of nitrous oxide and other permanent gases released from soil is essential for environmental monitoring, greenhouse gas inventory assessments and biogeochemical research. Multiplex gas chromatography systems equipped with both electron capture and thermal conductivity detectors enable simultaneous sensitive detection of N2O, CO, CO2, CH4, N2 and O2, improving analytical throughput and data reliability.
Objectives and Study Overview
This application note describes a method for the determination of soil-emitted N2O using gas chromatography with electron capture detection and simultaneous analysis of permanent gases via thermal conductivity detection. The system utilizes multiple packed columns and sample loops to separate a complex gas mixture into key target components.
Methods and Instrumentation
The analysis is performed on a Shimadzu Nexis GC-2030N (or GC-2014N) system configured with four switching valves, seven packed columns and two detectors (one ECD and one TCD). The flow path is divided into two channels:
- Channel 1: HayeSep-D column for N2O separation, detected by ECD.
- Channel 2: Two Porapak-N columns serve as pre-column and separation columns to remove C2 hydrocarbons and resolve Air/CH4/CO/CO2. Air, CH4 and CO are further separated on a MS-13X column and detected by TCD. CO2 is directed through Porapak-Q to the TCD.
Samples are introduced via two sample loops. The complementary use of ECD and TCD detectors covers trace to percent-level concentration ranges.
Main Results and Discussion
Typical chromatograms demonstrate baseline-resolved peaks for N2O (ECD) with a retention time around 4 min and well-separated peaks for CH4, N2, O2, CO2 and CO (TCD) between 1 and 9 min. The system achieves detection limits down to 50 ppb for N2O and 0.01% for permanent gases with excellent repeatability. Concentration ranges extend from trace to high-level percentages, accommodating diverse sample matrices.
Benefits and Practical Applications
- Simultaneous multi-component analysis reduces total run time and sample consumption.
- High sensitivity for N2O monitoring supports environmental and agricultural studies.
- Robust configuration with packed columns ensures long-term stability and reproducibility.
- Wide dynamic range meets requirements for research laboratories, QA/QC and field studies.
Future Trends and Possibilities
Advances in micro-detector technology and miniaturized GC systems could further enhance field deployability. Integration with automated soil gas sampling modules and remote data acquisition platforms will enable real-time greenhouse gas flux monitoring. Coupling with mass spectrometry or laser-based detectors may improve compound identification and isotopic analysis capabilities.
Conclusion
This method demonstrates a versatile, high-throughput GC system for concurrent quantification of N2O and permanent gases from soil samples. The dual-detector approach and multi-column configuration deliver reliable, sensitive and reproducible results, meeting the demands of environmental monitoring and research.
References
No references provided in the original text.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
N2O/CO/CO2/CH4 analysis system (TCD/FID) Nexis GC-2030NCCC1 GC-2014NCCC1
2017|Shimadzu|Applications
SGC-ADS-0057A System Gas Chromatograph N2O/CO/CO2/CH4 analysis system (TCD/FID) Nexis GC-2030NCCC1 GC-2014NCCC1 57 This method provides for the determination of Nitrous oxide (N2O) , in atmospheric air, by gas chromatography (GC) with Electron Capture detector (ECD) using Porapak-N and HayeSep-D packed…
Key words
tcd, tcdecd, ecdfid, fidmtn, mtnone, onedetector, detectorporapak, porapakhayesep, hayesepmin, minsystem, systemversatile, versatilecapture, capturevalves, valvesseven, sevenpacked
Shimadzu’s System GC Solutions - Data Sheets
2018|Shimadzu|Brochures and specifications
C10G-E064 System Gas Chromatography (GC) Data Sheets Shimadzu’s System GC Solutions Shimadzu provides proven System GC Solutions designed to meet the demands of your business and industry. Our System GC Analyzers are built and tested to meet the specific analysis…
Key words
fid, fidreturn, returntable, tabletcd, tcdsystem, systemanalyzer, analyzerone, onechromatograph, chromatographname, nametypical, typicalchromatograms, chromatogramspacked, packedfeatures, featuresshimadzu, shimadzugas
N2O/CO/CO2/CH4 analysis system (ECD/FID) Nexis GC-2030NCCC2 GC-2014NCCC2
2017|Shimadzu|Applications
SGC-ADS-0058A System Gas Chromatograph N2O/CO/CO2/CH4 analysis system (ECD/FID) Nexis GC-2030NCCC2 GC-2014NCCC2 58 This method provides for the determination of nitrous oxide (N2O) , in atmospheric air, by gas chromatography (GC) with Electron Capture Detector (ECD) using Porapak-N and HayeSep-D packed…
Key words
fid, fidecd, ecdone, onedetector, detectorporapak, porapakhayesep, hayesepaldehyde, aldehydesystem, systempropylene, propyleneversatile, versatilecapture, capturevalves, valvesseven, sevenpacked, packedchannel
NGA Analysis with ISO6974-3 Nexis Nexis GC-2030ISO6974-3 GC-2014ISO6974-3
2017|Shimadzu|Applications
SGC-ADS-0060A System Gas Chromatograph NGA Analysis with ISO6974-3 Nexis Nexis GC-2030ISO6974-3 GC-2014ISO6974-3 60 This method is for determining the chemical composition of natural gases within the composition range shown in the specification sheet. This test method provides data for calculating…
Key words
fid, fidcyclohexane, cyclohexanepropane, propanetcd, tcdmetylcyclohexane, metylcyclohexaneneopentane, neopentanetwo, twocomposition, compositionisopentane, isopentaneisobutane, isobutaneoctane, octanebutane, butaneethane, ethaneheptane, heptanesystem