Analysis of Oxygenates and Aromatics Using ASTM Method D4815 and D5580
Applications | 2018 | PerkinElmerInstrumentation
The accurate analysis of oxygenates and aromatic hydrocarbons in gasoline is critical for meeting environmental regulations, ensuring fuel performance, and protecting engine components. Traditional single-column gas chromatography methods often struggle to separate complex matrix interferences, leading to increased analysis time and instrument downtime.
This application note demonstrates a unified gas chromatographic solution that performs both ASTM D4815 (oxygenates in gasoline) and ASTM D5580 (aromatic hydrocarbons) using a single column set and programmable valving. The main goal is to enhance sample flexibility, minimize column changes, and improve overall laboratory productivity.
The method couples two capillary columns—a polar 20% TCEP column and a nonpolar Elite 1™ column—with timed valve switching and backflushing to remove unwanted light and heavy hydrocarbon fractions.
The dual-column approach achieved baseline resolution for all target analytes:
This streamlined GC configuration delivers several advantages:
Emerging developments will further enhance fuel analysis workflows:
The PerkinElmer dual-column, valve-switching GC approach effectively combines ASTM D4815 and D5580 into a single, robust method. This solution enhances analytical flexibility, reduces downtime, and delivers reliable quantitation of key oxygenate and aromatic compounds in gasoline, optimizing laboratory efficiency and return on investment.
GC
IndustriesEnergy & Chemicals
ManufacturerPerkinElmer
Summary
Importance of the Topic
The accurate analysis of oxygenates and aromatic hydrocarbons in gasoline is critical for meeting environmental regulations, ensuring fuel performance, and protecting engine components. Traditional single-column gas chromatography methods often struggle to separate complex matrix interferences, leading to increased analysis time and instrument downtime.
Study Objectives and Overview
This application note demonstrates a unified gas chromatographic solution that performs both ASTM D4815 (oxygenates in gasoline) and ASTM D5580 (aromatic hydrocarbons) using a single column set and programmable valving. The main goal is to enhance sample flexibility, minimize column changes, and improve overall laboratory productivity.
Methodology and Instrumentation
The method couples two capillary columns—a polar 20% TCEP column and a nonpolar Elite 1™ column—with timed valve switching and backflushing to remove unwanted light and heavy hydrocarbon fractions.
- Column 1: 20% TCEP, 56 cm × 1/16″
- Column 2: 30 m × 0.53 mm × 5 µm Elite 1™
- Detector: Wide-range flame ionization detector (FID)
- Instrumentation: PerkinElmer Model Arnel 4004 analyzer and Clarus 690 GC
- D4815 Conditions: Isothermal 60 °C, 30 min; valve on at 0.13 min, off at 8 min
- D5580 Conditions: Initial 60 °C (6 min), ramp to 115 °C; valve on at 0.25 min, off at 10 min
Key Results and Discussion
The dual-column approach achieved baseline resolution for all target analytes:
- Oxygenates (methanol, ethanol, propanols, MTBE, ETBE, etc.) separated efficiently in an isothermal run without detector saturation.
- Aromatics (benzene, toluene, ethylbenzene, xylenes, 1,2,4-trimethylbenzene) were quantified with clear peak separation and internal standard recovery.
- Backflushing eliminated matrix interferences, reducing run times and preventing column overload.
Benefits and Practical Applications
This streamlined GC configuration delivers several advantages:
- Single column set for both methods reduces instrument downtime associated with manual column changes.
- High sample throughput and minimal cool-down times boost productivity.
- Wide dynamic range FID ensures accurate quantitation across varying concentration levels.
- Flexible programming of valve events allows adaptation to additional fuel analyses.
Future Trends and Applications
Emerging developments will further enhance fuel analysis workflows:
- Automated valve control and software integration for remote monitoring and data handling.
- Expansion of method scope to include oxygenated and non-conventional fuel additives.
- Implementation of advanced detectors and miniaturized GC systems for in-field applications.
- Integration with laboratory information management systems (LIMS) for seamless quality control.
Conclusion
The PerkinElmer dual-column, valve-switching GC approach effectively combines ASTM D4815 and D5580 into a single, robust method. This solution enhances analytical flexibility, reduces downtime, and delivers reliable quantitation of key oxygenate and aromatic compounds in gasoline, optimizing laboratory efficiency and return on investment.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Benzene Toluene and Aromatic Analysis Nexis GC-2030_3606-4815-5580_1
2017|Shimadzu|Applications
SGC-ADS-0050A System Gas Chromatograph Benzene Toluene and Aromatic Analysis 50 Nexis GC-2030_3606-4815-5580_1 This system contains two analysis lines, one is for benzene, toluene analysis (ASTM-D3606), the other is for aromatic component analysis (ASTM-D5580). Flow diagram of ASTM-D4815 is exactly same…
Key words
toluene, toluenebenzene, benzenecolumn, columnback, backaromatic, aromaticbutanol, butanolswitched, switchedcompound, compoundwcot, wcotflush, flushboiling, boilinglighter, lighterflame, flameelute, eluteportions
Running ASTM Methods D4815 and D5580 on a Single Agilent 6890N Gas Chromatograph with Nitrogen Carrier Gas
2003|Agilent Technologies|Applications
Running ASTM Methods D4815 and D5580 on a Single Agilent 6890N Gas Chromatograph with Nitrogen Carrier Gas Application Petrochemical Author James D. McCurry Agilent Technologies, Inc. 2850 Centerville Rd. Wilmington DE 19808 USA Abstract An Agilent 6890N GC system is…
Key words
benzene, benzenetcep, tcepepc, epcgasoline, gasolinevalve, valvecapillary, capillarycolumn, columnpcm, pcmmtbe, mtbesplit, splitprecision, precisionnonaromatic, nonaromaticfid, fidtoluene, toluenehydrocarbons
Determining Aromatics in Finished Gasoline Using the Agilent 6820 GC System
2004|Agilent Technologies|Applications
Determining Aromatics in Finished Gasoline Using the Agilent 6820 GC System Application Gasoline Analysis and Environmental Protection Authors ChunXiao Wang Agilent Technologies Co., Ltd. (Shanghai) 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai 200131, P.R.C. Zhenxi Guan Agilent Technologies Co.,…
Key words
tcep, tceparomatics, aromaticsfid, fidvalve, valvebenzene, benzenefinished, finishedswitch, switchnds, ndsgasoline, gasolinerepeatability, repeatabilityxylene, xyleneinlet, inletcapillary, capillaryintercept, intercepttoluene
Agilent Reformulated Fuel Analyzers
2014|Agilent Technologies|Others
PRODUCE ACCURATE, RELIABLE DATA AND MAXIMIZE COMBUSTION EFFICIENCY Agilent Reformulated Fuel Analyzers Fuels such as gasoline undergo reformulation to improve combustion efficiency, and to meet strict U.S. EPA limits for volatile organic compounds (VOCs) and other emissions. For example, catalytic…
Key words
reformulated, reformulatedcheckout, checkoutfuel, fuelanalyzers, analyzerslvo, lvooven, ovendme, dmefactory, factoryvalve, valveyour, youragilent, agilentgasoline, gasolineanalyzer, analyzerbenzene, benzenesupport