Enhanced Metabolite Profiling: Hard and Soft Ionization High Resolution Time-of-Flight Mass Spectrometry
Applications | 2013 | LECOInstrumentation
Metabolomic profiling of biological tissues is critical for understanding cellular processes, biomarker discovery and disease mechanisms. Gas chromatography-mass spectrometry (GC–MS) is a cornerstone technique for separating and detecting volatile or derivatized metabolites. Combining electron ionization (EI) with soft chemical ionization (CI) on a high resolution time-of-flight mass spectrometer (GC–HRT-TOFMS) enhances sensitivity, mass accuracy and confidence in compound identification, especially for labile or low-abundance analytes.
The study aimed to demonstrate how complementary EI and CI high-resolution TOFMS analysis improves metabolite annotation in complex extracts. A mouse liver sample was derivatized, analyzed by GC–HRT-TOFMS under both ionization modes, and data were evaluated for mass accuracy, library matching and detection of molecular ions.
The analytical ion chromatogram revealed over ten well-resolved peaks from mouse liver extract. Representative TMS derivatives (e.g., lactic acid, alanine, glycine, malic acid, proline, erythronic acid, glucose, arachidonic acid, cholesterol) showed retention times from 324 s to 977 s and EI library similarity scores of 831–919 out of 1000. Mass accuracies near 1 ppm enabled robust formula assignments for molecular, fragment and adduct ions. Comparison of EI spectra with NIST library hits demonstrated high spectral match, while CI spectra provided intact protonated molecules for low-abundance or labile analytes. For example, the EI spectrum of 1,5-anhydroglucitol TMS yielded a marginal match (679/1000) without a molecular ion, whereas CI-HRT data detected a protonated ion at m/z 453.23334 (–1.1 ppm), conclusively confirming the formula C₁₈H₄₅O₅Si₄.
Advances in ionization techniques and high-resolution instrumentation will further enhance metabolomic workflows. Integration with automated data processing, expanding high-quality spectral libraries for CI data, coupling to multidimensional separations and real-time pathway mapping are promising directions. Continuous improvements in reagent gases and calibration strategies will push detection limits and broaden compound classes amenable to GC-MS profiling.
The combination of EI and CI on a GC–HRT-TOFMS platform delivers a powerful tool for comprehensive metabolite profiling. High resolution, accurate mass measurements enable confident identification of both abundant and trace analytes, addressing challenges in complex biological samples and supporting quantitative systems biology studies.
GC/MSD, GC/HRMS, GC/TOF
IndustriesMetabolomics, Clinical Research
ManufacturerAgilent Technologies, LECO
Summary
Significance of the Topic
Metabolomic profiling of biological tissues is critical for understanding cellular processes, biomarker discovery and disease mechanisms. Gas chromatography-mass spectrometry (GC–MS) is a cornerstone technique for separating and detecting volatile or derivatized metabolites. Combining electron ionization (EI) with soft chemical ionization (CI) on a high resolution time-of-flight mass spectrometer (GC–HRT-TOFMS) enhances sensitivity, mass accuracy and confidence in compound identification, especially for labile or low-abundance analytes.
Objectives and Study Overview
The study aimed to demonstrate how complementary EI and CI high-resolution TOFMS analysis improves metabolite annotation in complex extracts. A mouse liver sample was derivatized, analyzed by GC–HRT-TOFMS under both ionization modes, and data were evaluated for mass accuracy, library matching and detection of molecular ions.
Methodology and Instrumentation
- Sample Preparation
Two-step trimethylsilyl (TMS) derivatization: methoximation (20 mg/mL in pyridine, 90 min at 30 °C) followed by silylation (BSTFA + 1% TMCS, 37 °C, 30 min). - Chromatography and Mass Spectrometry
Agilent 7890A GC with LECO Pegasus® GC–HRT (high resolution TOFMS, R≈25 000).
Column: Rxi-5Sil MS (30 m × 0.25 mm × 0.25 µm), helium carrier at 1 mL/min.
Temperature: 50 °C (1 min) to 320 °C (5 min) at 20 °C/min.
Ionization: EI at 70 eV and CI at 140 eV using 5% NH₃ in CH₄ reagent gas.
Mass range: m/z 30–510 (EI), 50–650 (CI); acquisition rate 6 sps.
Internal calibration with PFTBA; source at 250 °C (EI) or 180 °C (CI).
Main Results and Discussion
The analytical ion chromatogram revealed over ten well-resolved peaks from mouse liver extract. Representative TMS derivatives (e.g., lactic acid, alanine, glycine, malic acid, proline, erythronic acid, glucose, arachidonic acid, cholesterol) showed retention times from 324 s to 977 s and EI library similarity scores of 831–919 out of 1000. Mass accuracies near 1 ppm enabled robust formula assignments for molecular, fragment and adduct ions. Comparison of EI spectra with NIST library hits demonstrated high spectral match, while CI spectra provided intact protonated molecules for low-abundance or labile analytes. For example, the EI spectrum of 1,5-anhydroglucitol TMS yielded a marginal match (679/1000) without a molecular ion, whereas CI-HRT data detected a protonated ion at m/z 453.23334 (–1.1 ppm), conclusively confirming the formula C₁₈H₄₅O₅Si₄.
Benefits and Practical Applications
- Increased identification confidence via complementary EI and CI spectra.
- High mass accuracy (<1 ppm) supports unambiguous elemental composition determination.
- Expanded coverage of labile, low-level metabolites not easily detected by EI alone.
- Improved spectral library matching and peak deconvolution in complex matrices.
Future Trends and Potential Applications
Advances in ionization techniques and high-resolution instrumentation will further enhance metabolomic workflows. Integration with automated data processing, expanding high-quality spectral libraries for CI data, coupling to multidimensional separations and real-time pathway mapping are promising directions. Continuous improvements in reagent gases and calibration strategies will push detection limits and broaden compound classes amenable to GC-MS profiling.
Conclusion
The combination of EI and CI on a GC–HRT-TOFMS platform delivers a powerful tool for comprehensive metabolite profiling. High resolution, accurate mass measurements enable confident identification of both abundant and trace analytes, addressing challenges in complex biological samples and supporting quantitative systems biology studies.
Instrumentation Used
- LECO Pegasus® GC–HRT high resolution time-of-flight mass spectrometer.
- Agilent 7890A gas chromatograph with 7693 autosampler.
- Rxi-5Sil MS column (30 m × 0.25 mm × 0.25 µm).
- CI reagent gas: 5% ammonia in methane; internal PFTBA calibrant.
References
- Kind T. et al. FiehnLib: Mass spectral and retention index libraries for metabolomics. Anal. Chem. 2009, 81, 100038–10048.
- Lisec J. et al. GC–MS-based metabolite profiling in plants. Nat. Protoc. 2006, 1(1), 387–396.
- Warren C.R. Use of chemical ionization for GC–MS metabolite profiling. Metabolomics 2011, 1–11.
- Wang Y. et al. Association of serum 1,5-anhydroglucitol levels and hyperglycaemic excursions measured by CGMS. Diabetes Metab. Res. Rev. 2012, 28(4), 357–362.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Improved Methodology for Metabolomics Data Acquisition Workflow: Utilization of Electron Impact and Chemical Ionization High Resolution Time-of-Flight Mass Spectrometry
2013|Agilent Technologies|Posters
Improved Methodology for Metabolomics Data Acquisition Workflow: Utilization of Electron Impact and Chemical Ionization High Resolution Time-of-Flight Mass Spectrometry David E. Alonso*, Joe Binkley, John Heim and Jeff Patrick | LECO Corporation, St. Joseph, Michigan USA Experimental 500 1000 1500…
Key words
hrt, hrtheight, heightcounts, countstms, tmsdata, datatrue, truenist, nistpegasus, pegasusformula, formulapeak, peakacid, acidconfident, confidentosi, osiparameters, parametersconfirmation
Solutions for Metabolomics Analysis
2009|LECO|Brochures and specifications
Solutions for Metabolomics Analysis EMPOWERING RESULTS The Established Leader in GC-MS Metabolomics Metabolomics presents challenges for both the analytical methods used and the data reduction required to interpret the results. No single analytical technique can be used for complete characterization…
Key words
identi, identigcxgc, gcxgcpegasus, pegasustofms, tofmsmetabolomics, metabolomicscon, concholesterol, cholesterolour, ourdeconvolution, deconvolutionleco, lecodence, dencewhat, whatcation, cationtms, tmsyour
Strategic Utilization of Gas Chromatography with Both Nominal and High Resolution Time-of-Flight Mass Spectrometers for Metabolomic Studies
2014|LECO|Posters
Strategic Utilization of Gas Chromatography with Both Nominal and High Resolution Time-of-Flight Mass Spectrometers for Metabolomic Studies David E. Alonso, Jeffrey S. Patrick, and Joe Binkley | LECO Corporation, St. Joseph, MI USA Results and Discussion – Instrumental Platforms Introduction…
Key words
tms, tmshrt, hrtpegasus, pegasusacid, acidpalmitoleic, palmitoleictemp, tempsimilarity, similaritycholesterol, cholesterolphenotype, phenotypeoctadecanoic, octadecanoicarachidonic, arachidonicmass, massdiscovery, discoverymonosaccharides, monosaccharidesoleic
Electron Impact and Chemical Ionization High Resolution Time-of-Flight Mass Spectrometry Analyses of Blood Plasma Samples
2013|Agilent Technologies|Posters
Electron Impact and Chemical Ionization High Resolution Time-of-Flight Mass Spectrometry Analyses of Blood Plasma Samples David E. Alonso, Jeffrey S. Patrick, Joe Binkley, and John Heim | LECO Corporation, St. Joseph, MI USA 0.0e0 M/Z 800 AIC Fatty 1 900…
Key words
hrt, hrtheight, heightcounts, countshexatms, hexatmsobese, obesetms, tmslean, leanpegasus, pegasusfatty, fattyoxime, oximeaic, aiccholestadiene, cholestadienesinh, sinhgenedata, genedatadeconvolution