Evaluation of a Liquid Cryogen-Free Thermal Modulator for GCxGC
Posters | 2009 | LECOInstrumentation
The development of GCxGC thermal modulators that do not rely on liquid cryogens addresses significant operational and cost challenges in analytical laboratories. Eliminating LN2 dewars reduces consumable expenses, frees up benchtop space, and simplifies instrument control, enhancing accessibility and sustainability of comprehensive two-dimensional gas chromatography.
This study evaluates a consumable-free (CF) dual-stage quad-jet modulator using a closed-loop chiller with a cold probe to generate cooled nitrogen for GCxGC. Performance is compared against a traditional liquid nitrogen (LN2) modulator for the analysis of pump diesel fuel, focusing on volatility range, peak widths, and operational flexibility.
Advances in closed-loop cooling may extend the usable volatility range further toward lighter compounds. Integration with adaptive modulation schemes and enhanced data processing could improve resolution and throughput. Coupling CF modulators with novel stationary phases may expand applications into environmental and life-science analyses.
The consumable-free GCxGC modulator delivers performance on par with traditional LN2 systems for mid-to-high volatility compounds while streamlining operations and reducing costs. Its flexible control and sustainable design establish it as a viable alternative for routine and advanced comprehensive two-dimensional gas chromatography.
No external references were provided in the source document.
GCxGC, GC/MSD, GC/TOF
IndustriesManufacturerLECO
Summary
Importance of the Topic
The development of GCxGC thermal modulators that do not rely on liquid cryogens addresses significant operational and cost challenges in analytical laboratories. Eliminating LN2 dewars reduces consumable expenses, frees up benchtop space, and simplifies instrument control, enhancing accessibility and sustainability of comprehensive two-dimensional gas chromatography.
Objectives and Study Overview
This study evaluates a consumable-free (CF) dual-stage quad-jet modulator using a closed-loop chiller with a cold probe to generate cooled nitrogen for GCxGC. Performance is compared against a traditional liquid nitrogen (LN2) modulator for the analysis of pump diesel fuel, focusing on volatility range, peak widths, and operational flexibility.
Methodology and Instrumentation
- Sample: Neat pump diesel injection (200 nL) in split mode (100:1) at 250 °C.
- Primary column: 50 m × 0.20 mm ID × 0.5 µm df DB-PONA
- Secondary column: 1.7 m × 0.10 mm ID × 0.1 µm df DB-Wax
- Carrier gas: Helium at 1.5 mL/min constant flow
- Modulation: 5 s period with variable hot pulse duration (0.6–1.0 s)
- Temperature program: Ramp to 240 °C (1.5 °C/min for LN2, 2.0 °C/min for CF), hold 20 min.
Main Results and Discussion
- The CF modulator reliably trapped and desorbed compounds from n-C8 to n-C40, matching the LN2 system’s range.
- Peak width analysis (FWHH for biphenyl) showed comparable performance when modulation occurred at the start of the secondary column.
- Shifting modulation to the end of the primary column enhanced trapping of lighter compounds (< n-C8) but slightly broadened desorption peaks due to reduced carrier gas velocity.
- The CF system’s higher minimum temperature minimized overcooling and maintained consistent peak shapes for less volatile analytes.
Advantages and Practical Applications
- Elimination of liquid nitrogen reduces operating costs and laboratory clutter.
- Piggy-back electronics free up GC auxiliary controls and enable dynamic hot pulse programming during runs.
- Well suited for complex hydrocarbon mixtures, such as fuels and petrochemicals, within the C8–C40 range.
Future Trends and Perspectives
Advances in closed-loop cooling may extend the usable volatility range further toward lighter compounds. Integration with adaptive modulation schemes and enhanced data processing could improve resolution and throughput. Coupling CF modulators with novel stationary phases may expand applications into environmental and life-science analyses.
Conclusion
The consumable-free GCxGC modulator delivers performance on par with traditional LN2 systems for mid-to-high volatility compounds while streamlining operations and reducing costs. Its flexible control and sustainable design establish it as a viable alternative for routine and advanced comprehensive two-dimensional gas chromatography.
Reference
No external references were provided in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
LECO's GCxGC Utilizing a Consumable- Free Modulator and Second Column Modulation
2009|LECO|Technical notes
® LECO's GCxGC Utilizing a ConsumableFree Modulator and Second Column Modulation LECO Corporation; Saint Joseph, Michigan USA NOTE: This document is intended to provide an introduction to LECO's GCxGC system utilizing the Consumable-Free Thermal Modulator. This work will cover differences…
Key words
cfm, cfmmodulation, modulationmodulated, modulatedband, bandgcxgc, gcxgchvcc, hvccbrief, briefdelivering, deliveringscc, sccvolatilities, volatilitiesdewar, dewarcold, coldright, rightusable, usablearm
LECO's GCxGC Utilizing a Consumable- Free Modulator and Second Column Modulation
2009|LECO|Technical notes
® LECO's GCxGC Utilizing a ConsumableFree Modulator and Second Column Modulation LECO Corporation; Saint Joseph, Michigan USA NOTE: This document is intended to provide an introduction to LECO's GCxGC system utilizing the Consumable-Free Thermal Modulator. This work will cover differences…
Key words
cfm, cfmmodulation, modulationmodulated, modulatedband, bandgcxgc, gcxgchvcc, hvccbrief, briefdelivering, deliveringscc, sccvolatilities, volatilitiesdewar, dewarcold, coldright, rightusable, usablearm
Petroleum Analysis Using the LECO Pegasus 4D GCxGC-TOFMS System
2008|Agilent Technologies|Applications
® Petroleum Analysis Using the LECO® Pegasus® 4D GCxGC-TOFMS System LECO Corporation; Saint Joseph, Michigan USA Key Words: GCxGC-TOFMS, Petroleum, Classifications 2. Experimental Conditions Gasoline GC: Agilent 6890 Gas Chromatograph equipped with a LECO GCxGC Thermal Modulator Injection: o Split…
Key words
paraffin, paraffingcxgc, gcxgcvolatility, volatilityleco, lecopetroleum, petroleumdiesel, dieselpeak, peakmodulator, modulatorautomatically, automaticallylocates, locatesthermal, thermalpeaks, peaksclasses, classesdimension, dimensionmodulated
An Introduction to LECO's Comprehensive Two-Dimensional Gas Chromatography (GCxGC) with ChromaTOF Software
2009|LECO|Technical notes
® An Introduction to LECO's Comprehensive Two-Dimensional Gas Chromatography (GCxGC) with ChromaTOF® Software LECO Corporation; Saint Joseph, Michigan USA NOTE: This document is intended to provide an introduction to LECO's GCxGC systems. Suggestions and recommendations are intended to help new…
Key words
gcxgc, gcxgcsection, sectiondimension, dimensionbrief, briefdelivering, deliveringpeak, peakright, rightwidth, widthlibrary, librarytechnical, technicalmethod, methodwhere, wherefirst, firstparameters, parametersspectra