Iraqi Basrah Light Crude Oil on Rxi®-1HT (15 m x 0.25 mm x 0.10 μm)

Applications |  | RestekInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies, Restek

Summary

Importance of the Topic

Gas chromatography analysis of light crude oil fractions is critical for petroleum characterization, quality control, and process optimization. Accurate separation and quantification of n-alkanes from C10 to C40 enables detailed hydrocarbon profiling, informing refining strategies and environmental assessments.

Study Objectives and Overview

This study aimed to demonstrate the separation performance of an Rxi-1HT capillary column for Iraqi Basrah light crude oil. Four n-alkane standards (C10, C20, C30, C40) served as markers to evaluate retention behavior under high-temperature conditions.

Methodology


  • Sample Preparation: Iraqi Basrah light crude oil diluted in methylene chloride to 50,000 ng/µL.
  • Injection: 1 µL volume, split ratio 10:1, using a wool-packed liner at 275 °C.
  • Oven Program: Initial 40 °C hold 0.1 min, ramp at 20 °C/min to 400 °C, hold 1.9 min.
  • Carrier Gas: Hydrogen at constant flow 1.75 mL/min.
  • Detector: Flame ionization detector at 420 °C, with nitrogen make-up gas at 50 mL/min.


Instrumentation Used


  • Column: Restek Rxi-1HT, 15 m × 0.25 mm ID × 0.10 µm film thickness.
  • GC System: Agilent/HP 6890 equipped with FID.
  • Additional Components: Premium 4 mm precision liner with wool; split vent flow at 17.5 mL/min; hydrogen flow 40 mL/min; air flow 450 mL/min.


Key Results and Discussion

The Rxi-1HT column achieved clear baseline separation of n-alkanes C10, C20, C30, and C40 with retention times of approximately 119, 456, 678, and 839 seconds, respectively. The linear increase in retention time with carbon number confirms predictable thermal stability and column efficiency. Actual column length was adjusted to 15.7 m based on holdup time calculations.

Benefits and Practical Applications of the Method

  • High-temperature capability allows analysis of heavy hydrocarbon fractions in a single run.
  • Robust separation enhances quantification accuracy for quality control in refineries.
  • Efficient hydrogen carrier gas reduces analysis time compared to traditional helium-based methods.


Future Trends and Potential Applications

Advances may include ultrafast temperature programming, novel stationary phases for improved selectivity, and integration with mass spectrometry for structural elucidation. Sustainable carrier gases and miniaturized GC systems will further expand on-site petroleum analysis.

Conclusion

The application of Rxi-1HT column under optimized high-temperature conditions provides reliable separation of key n-alkanes in light crude oil, offering a powerful tool for petroleum analysis and industrial quality control.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Iraqi Basrah Light Crude Oil on Rxi®-5HT (15 m x 0.25 mm x 0.10 μm)
Iraqi Basrah Light Crude Oil on Rxi®-5HT (15 m x 0.25 mm x 0.10 µm) 1. 2. 3. 4. Peaks C10 C20 C30 C40 Column Rxi®-5HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13905) Sample Iraqi Basrah light crude oil…
Key words
iraqi, iraqibasrah, basrahsec, secholdup, holdupezgc, ezgccalculator, calculatorcrude, crudevent, ventflow, flowconstant, constantoil, oillight, lightsplit, splitmake, maketime
Arabian Medium Crude Oil on Rxi®-1HT (15 m x 0.25 mm x 0.10 μm)
Arabian Medium Crude Oil on Rxi®-1HT (15 m x 0.25 mm x 0.10 µm) 1. 2. 3. Peaks C10 C20 C30 Column Rxi®-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950) Sample Arabian medium crude oil Diluent: Methylene…
Key words
arabian, arabiansec, secholdup, holdupezgc, ezgccalculator, calculatorcrude, crudemedium, mediumvent, ventconstant, constantoil, oilsplit, splitmake, maketime, timeoven, ovenpeaks
South Louisiana Crude Oil on Rxi®-1HT (15 m x 0.25 mm x 0.10 μm)
South Louisiana Crude Oil on Rxi®-1HT (15 m x 0.25 mm x 0.10 µm) Peaks 1. C10 2. C20 3. C30 C10 Column Sample Diluent: Conc.: Injection Inj. Vol.: Liner: Inj. Temp.: Split Vent Flow Rate: Oven Oven Temp.: Carrier…
Key words
louisiana, louisianasec, secholdup, holdupezgc, ezgccalculator, calculatorcrude, crudesouth, southmethylene, methylenevent, ventnotes, noteschloride, chlorideconstant, constantoil, oilcolumn, columnsplit
Florida Beach Tarball 11 on Rxi®-1HT (15 m x 0.25 mm x 0.10 μm)
Florida Beach Tarball 11 on Rxi®-1HT (15 m x 0.25 mm x 0.10 µm) 1. 2. 3. 4. 5. Peaks C10 C20 C30 C40 C50 Column Rxi®-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950) Sample Tarball 11 - Florida…
Key words
sec, secholdup, holdupezgc, ezgccalculator, calculatorvent, ventflow, flowconstant, constantsplit, splitmake, maketime, timeoven, ovenpeaks, peaksinjection, injectiongas, gascolumn
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike