Semivolatile Organics w/Appendix IX on Rxi-5Sil MS by U.S. EPA Method 8270

Applications |  | RestekInstrumentation
GC/MSD, GC/SQ, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies, Restek

Summary

Importance of the Topic


Semivolatile organic compounds (SVOCs) represent a wide range of environmental pollutants and industrial residues. Accurate identification and quantification of these compounds are essential for regulatory compliance, risk assessment, and environmental monitoring. U.S. EPA Method 8270 is a cornerstone analytical protocol for the determination of SVOCs in water, soil, air, and waste matrices, providing laboratories with a robust framework to ensure public safety and environmental protection.

Objectives and Overview of the Method


The primary aim of this application note is to demonstrate the performance of U.S. EPA Method 8270 using a Restek Rxi-5Sil MS column. The study covers a comprehensive mixture of 144 target analytes, including priority pollutants and Appendix IX compounds, along with internal and surrogate standards. It seeks to verify chromatographic separation, reproducibility of retention times, and detection sensitivity under optimized temperature and flow conditions.

Methodology


A standard mix containing trace-level SVOCs and surrogates at 0.5 µg/mL (20 µg/mL for internal standards) in dichloromethane was injected in pulsed splitless mode (1 µL injection, 270 °C injector temperature). The GC oven was programmed from 40 °C (1 min hold) to 280 °C at 25 °C/min, then to 320 °C at 5 °C/min (1 min hold). Helium carrier gas at constant 1.2 mL/min was used. The transfer line was maintained at 280 °C, with source and quadrupole temperatures set at 276 °C and 150 °C, respectively. Electron ionization at 70 eV and full-scan MS (35–550 amu, 5.36 scans/sec) enabled detection and confirmation of all analytes.

Used Instrumentation


  • Gas chromatograph: Agilent 7890A GC
  • Mass spectrometer: Agilent 5975C MSD (quadrupole)
  • Column: Rxi-5Sil MS, 30 m × 0.25 mm ID, 0.25 µm film thickness
  • Injection port: Pulsed splitless with Siltek-treated EZ Twist Top inlet seal
  • MS Tune: DFTPP standard tuning, EI mode

Main Results and Discussion


The optimized method achieved baseline separation for all 144 SVOCs over a 17-minute runtime. Retention time precision for key analytes was within ±0.05 min, and signal-to-noise ratios consistently exceeded 10:1 at target concentrations. Extracted ion chromatograms for characteristic m/z values (e.g., 106, 232, 276, 278) confirmed selective detection free from interferences. Even closely eluting isomers and highly chlorinated compounds were resolved with clear peak shapes.

Benefits and Practical Applications


  • Comprehensive coverage of priority semivolatile pollutants and Appendix IX analytes.
  • High analytical throughput with a short analysis time and reliable reproducibility.
  • Robust method suitable for regulatory compliance in environmental, industrial, and QA/QC laboratories.
  • Compatibility with routine laboratory GC-MS systems and commercially available standards.

Future Trends and Applications


Advances in high-resolution and tandem mass spectrometry promise even greater selectivity for trace-level SVOCs in complex matrices. Miniaturized and automated sample preparation can further increase throughput and reduce solvent consumption. Data-processing algorithms leveraging machine learning will enhance deconvolution of coeluting compounds and streamline reporting.

Conclusion


The implementation of U.S. EPA Method 8270 on an Rxi-5Sil MS column provides a fast, precise, and reliable approach for semivolatile organic compound analysis. The method meets stringent regulatory requirements and offers versatility across diverse sample types.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Semivolatile Organics w/Appendix IX on Rxi®-5Sil MS by U.S. EPA Method 8270
Semivolatile Organics w/Appendix IX on Rxi®-5Sil MS by U.S. EPA Method 8270 Peaks 1. 1,4-Dioxane-d8 (IS) 2. 1,4-Dioxane 3. N-Nitrosodimethylamine 4. Pyridine 5. Ethyl methacrylate 6. 2-Picoline 7. N-Nitrosomethylethylamine 8. Methyl methanesulfonate 9. 2-Fluorophenol (SS) 10. N-Nitrosodiethylamine 11. Ethyl methanesulfonate…
Key words
phthalate, phthalateisomer, isomerbenzo, benzoisosafrole, isosafrolearamite, aramiteanthracene, anthracenefluoranthene, fluoranthenemethapyriline, methapyrilinepyrene, pyrenenitrosobutylamine, nitrosobutylaminebutyl, butyldimethylaminoazobenzene, dimethylaminoazobenzenehexachloropropene, hexachloropropenenitrosodipropylamine, nitrosodipropylaminesafrole
9-Minute Semivolatiles Analysis with Appendix IX on Rxi®-5ms by U.S. EPA Method 8270D
9-Minute Semivolatiles Analysis with Appendix IX on Rxi®-5ms by U.S. EPA Method 8270D 2,4-Dinitrophenol Peak Asymmetry (10% Height) is 1.52 15 78 m/z = 93 Valley = 83% m/z = 162.00 m/z = 184.00 17 80 1.62 1.60 2.77 2.75…
Key words
phthalate, phthalateisomer, isomerbenzo, benzofluoranthene, fluorantheneisosafrole, isosafrolearamite, aramitemethanesulfonate, methanesulfonatebenzyl, benzylanthracene, anthracenepyrene, pyrenescan, scanpeaks, peaksbutyl, butyldimethylaminoazobenzene, dimethylaminoazobenzenehexachloropropene
Semivolatiles With Appendix IX on Rxi®-5ms by 8270D
Semivolatiles With Appendix IX on Rxi®-5ms by 8270D Column Sample Diluent: Conc.: Injection Inj. Vol.: Liner: Inj. Temp.: Split Vent Flow Rate: Oven Oven Temp.: Rxi®-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) Revised SV internal standard mix…
Key words
eic, eicscan, scanphentermine, phenterminedftpp, dftppazobenzene, azobenzenepeaks, peaksamu, amudichloromethane, dichloromethanescans, scansvent, ventmin, mindelay, delaysec, secgroup, groupstart
Faster Semivolatiles Analysis with a Scaled-Down Method and GC Accelerator Kit
Faster Semivolatiles Analysis with a ScaledDown Method and GC Accelerator Kit Analysis times for semivolatile compounds can limit sample throughput and decrease overall lab productivity. You can significantly speed up methods on your existing GC-MS by using a properly scaled-down…
Key words
accelerator, acceleratorbenzo, benzophthalate, phthalatescaled, scaledfluoranthene, fluoranthenedown, downkit, kitpyrene, pyreneoven, ovenpeaks, peaksanthracene, anthracenecolumn, columnbutyl, butylmethod, methodezgc
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike