Chlorinated Disinfection Byproducts, Chlorinated Solvents, and Halogenated Pesticides US EPA Method 551.1 - Rtx®-1301
Applications | | RestekInstrumentation
Chlorinated disinfection by-products, solvents, and halogenated pesticides are common environmental contaminants with significant health and regulatory implications. Accurate trace-level analysis is essential to assess exposure risks, enforce water quality standards, and guide remediation efforts.
The primary aim of this work is to present a GC-ECD method based on US EPA Method 551.1 for simultaneous determination of 35 chlorinated by-products, solvents, and pesticides. The study demonstrates chromatographic conditions optimized for sensitivity, selectivity, and reproducibility to support routine environmental monitoring.
Standard solutions containing each analyte at 5–10 µg/mL were prepared from certified pesticide/herbicide and disinfection by-products mixes. A 1.0 µL splitless injection (0.5 min solvent delay) was performed using a 4 mm split liner packed with fused silica wool. The temperature program was:
Baseline separation was achieved for all 35 target compounds, with sharp peaks and consistent retention times. ECD provided sub-ppb detection limits for halogenated species. The temperature program enabled efficient elution of low- and high-boiling analytes in a single run.
Emerging column chemistries and advanced detectors (e.g., micro-ECD, MS) could further lower detection limits and improve selectivity. Automated sample preparation and portable GC units may enable on-site monitoring in remote locations.
The described GC-ECD method based on EPA 551.1 using an Rtx-1301 column offers a reliable, sensitive, and high-throughput solution for monitoring chlorinated disinfection by-products, solvents, and halogenated pesticides in environmental matrices.
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerRestek
Summary
Importance of the Topic
Chlorinated disinfection by-products, solvents, and halogenated pesticides are common environmental contaminants with significant health and regulatory implications. Accurate trace-level analysis is essential to assess exposure risks, enforce water quality standards, and guide remediation efforts.
Objectives and Study Overview
The primary aim of this work is to present a GC-ECD method based on US EPA Method 551.1 for simultaneous determination of 35 chlorinated by-products, solvents, and pesticides. The study demonstrates chromatographic conditions optimized for sensitivity, selectivity, and reproducibility to support routine environmental monitoring.
Methodology
Standard solutions containing each analyte at 5–10 µg/mL were prepared from certified pesticide/herbicide and disinfection by-products mixes. A 1.0 µL splitless injection (0.5 min solvent delay) was performed using a 4 mm split liner packed with fused silica wool. The temperature program was:
- 35 °C (hold 22 min)
- Ramp to 145 °C at 10 °C/min (hold 2 min)
- Ramp to 225 °C at 20 °C/min (hold 15 min)
- Ramp to 260 °C at 10 °C/min (hold 30 min)
Used Instrumentation
- Gas chromatograph with electron capture detector (ECD) at 290 °C
- Capillary column: Rtx-1301, 30 m × 0.25 mm ID, 1.0 µm film
- Carrier gas: Helium, constant pressure for 30 cm/s linear velocity at 35 °C
- Injection liner: 4 mm split liner with fused silica wool
Main Results and Discussion
Baseline separation was achieved for all 35 target compounds, with sharp peaks and consistent retention times. ECD provided sub-ppb detection limits for halogenated species. The temperature program enabled efficient elution of low- and high-boiling analytes in a single run.
Benefits and Practical Applications
- Comprehensive analysis of multiple halogenated groups in one injection
- High sensitivity suitable for drinking water and environmental samples
- Robust method with minimal matrix interference
- Support for regulatory compliance and QA/QC laboratories
Future Trends and Potential Applications
Emerging column chemistries and advanced detectors (e.g., micro-ECD, MS) could further lower detection limits and improve selectivity. Automated sample preparation and portable GC units may enable on-site monitoring in remote locations.
Conclusion
The described GC-ECD method based on EPA 551.1 using an Rtx-1301 column offers a reliable, sensitive, and high-throughput solution for monitoring chlorinated disinfection by-products, solvents, and halogenated pesticides in environmental matrices.
References
- US EPA Method 551.1: Determination of Chlorinated Disinfection Byproducts, Chlorinated Solvents, and Halogenated Pesticides.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Chlorinated Disinfection Byproducts US EPA Method 551.1 Rtx®-5
|Restek|Applications
Chlorinated Disinfection Byproducts US EPA Method 551.1 Rtx®-5 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. chloroform 1,1,1-trichloroethane carbon tetrachloride trichloroacetonitrile trichloroethylene bromodichloromethane chloral hydrate dichloroacetonitrile 1,1-dichloro-2-propanone…
Key words
bromochloroacetonitrile, bromochloroacetonitriledibromoacetonitrile, dibromoacetonitrilechloropicrin, chloropicrintrichloroacetonitrile, trichloroacetonitriledichloroacetonitrile, dichloroacetonitriledbcp, dbcpchloral, chloralhydrate, hydratedisinfection, disinfectiontetrachloroethylene, tetrachloroethylenebyproducts, byproductstrichloroethylene, trichloroethylenedibromochloromethane, dibromochloromethanebromodichloromethane, bromodichloromethanebromoform
Analysis of Halogenated Environmental Pollutants Using Electron Capture Detection
2014|Agilent Technologies|Guides
TECHNICAL GUIDE Analysis of Halogenated Environmental Pollutants Using Electron Capture Detection • Sample preparation techniques for liquid, solid, and biota samples. • Chromatographic methods and best practices for halogenated pollutants. • Includes analysis of chlorinated pesticides, PCBs, chlorinated herbicides, haloacetic…
Key words
clpesticides, clpesticidesrtx, rtxpress, presschlorinated, chlorinatedcolumns, columnsextraction, extractiontight, tightconfirmational, confirmationalbhc, bhckit, kitinjection, injectionspe, speendrin, endrinresprep, resprepconnectors
A Guide to Preparing and Analyzing Chlorinated Pesticides
1999|Restek|Guides
Technical Guide A Guide to Preparing and Analyzing Chlorinated Pesticides Inside: Extraction Methods for Liquid, Solid, and Biota Samples Sample Cleanup Methods Analysis of Chlorinated Pesticides Summary 2 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○…
Key words
pack, packbhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordanehch, hchheptachlor, heptachlorclpesticides, clpesticidesdata, dataeach, eachrtx, rtxester, esterextraction, extractionpress, presstight
New Application Specific Columns
2013|Agilent Technologies|Presentations
New Application Specific Columns Simon Jones GC columns Application Engineer What Columns are New? DB-CLP1 DB-CLP2 Environmental DB-UI 8270D DB-Select 624 UI 467 US Pharmacopeia What are the Benefits? Address specific issues with certain methods Response Tailing factor Resolution Named…
Key words
min, minbhc, bhcendrin, endrinheptachlor, heptachlorchlordane, chlordaneendosulfan, endosulfanpermethrin, permethrinpesticides, pesticidesepoxide, epoxidemethyl, methylaldrin, aldrinnonachlor, nonachlordecachlorobiphenyl, decachlorobiphenylbenzo, benzotrans