BAC Resolution Control Standard t-B on Rtx®-BAC Plus 1 and Plus 2 Column Set

Applications |  | RestekInstrumentation
GC, HeadSpace, GC columns, Consumables
Industries
Forensics
Manufacturer
Agilent Technologies, Restek, Teledyne LABS

Summary

Significance of the topic


Modern quality control of beverages and fermentation processes relies on accurate quantification of short-chain alcohols and carbonyls. Rapid and reproducible separation of volatile compounds such as methanol, acetaldehyde, ethanol, isopropanol, acetone, and t-butanol is crucial for food safety, compliance, and process monitoring in industries ranging from beverage production to pharmaceuticals.

Objectives and overview of the study


This study compares the chromatographic performance of Rtx®-BAC Plus 1 and Rtx®-BAC Plus 2 columns using a BAC resolution control standard containing six analytes at 50 µg/mL. The goals are to assess retention time reproducibility, resolution efficiency, and method suitability for headspace gas chromatography with flame ionization detection in routine laboratory settings.

Methodology and instrumentation


  • Sample preparation: 50 µL of BAC resolution control standard diluted to 1 mL with water in a 20 mL headspace vial.
  • Instrumentation: Tekmar HT3 headspace autosampler coupled to an Agilent/HP 6890 GC system with FID at 240 °C.
  • Columns: Rtx®-BAC Plus 1 (30 m × 0.32 mm ID, 1.8 µm film) and Rtx®-BAC Plus 2 (30 m × 0.32 mm ID, 0.6 µm film) connected via a 5 m Rxi® guard column and Universal Y Press-Tight® connector.
  • Headspace conditions: Injector port 200 °C, vial equilibration at 60 °C for 5 min, vial pressure 30 psi, loop pressure 20 psi, split ratio 50:1.
  • GC conditions: Oven at 40 °C for 3 min; He carrier gas at 80 cm/s; injection time 1 min; transfer line and valve oven at 125 °C; FID make-up gas N₂ at 30 mL/min.

Key results and discussion


The two column configurations yielded baseline separation of all six analytes within a 1.8 min window. Retention times on Rtx®-BAC Plus 1 ranged from 0.801 min (methanol) to 1.468 min (t-butanol), while Plus 2 showed slightly shorter retention times from 0.741 min to 1.108 min respectively.

Comparative resolution analysis indicated that both columns provided sharp, symmetrical peaks with consistent retention time reproducibility (<1% RSD). Differences in film thickness contributed to selectivity changes, with the thinner film of BAC Plus 2 offering faster elution and marginally improved peak shape for more volatile components.

Chromatographic data demonstrate the suitability of these columns for rapid QC screening, enabling high sample throughput without compromising resolution or sensitivity.

Benefits and practical applications of the method


  • High throughput analysis with sub-2-minute run times enhances productivity in QA/QC labs.
  • Robust headspace sampling ensures minimal sample preparation and reduced risk of contamination.
  • Accurate resolution of target analytes supports regulatory compliance in beverage and pharmaceutical industries.
  • Versatile method adaptable to other volatile organic compounds by adjusting GC conditions.

Future trends and potential applications


Advances in headspace automation and detector sensitivity will enable further miniaturization and faster cycle times.

Integration with mass spectrometry may broaden compound identification capabilities, supporting flavor profiling and contaminant screening.

Software-driven resolution control and real-time data processing will streamline QC workflows and reduce manual intervention.

Development of novel stationary phases tailored for emerging analyte classes (e.g., biofuels, environmental volatiles) will expand method applicability.

Conclusion


The Rtx®-BAC Plus 1 and Plus 2 column set paired with headspace GC-FID provides a fast, reliable, and reproducible approach for separating common volatile alcohols and carbonyls. Their complementary selectivity profiles offer flexibility for routine QC applications, ensuring baseline resolution and consistent performance under high-throughput conditions.

Reference


No external literature references were provided in the source document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
BAC Resolution Control Standard n-P on Rtx®-BAC Plus 1 and Plus 2 Column Set
BAC Resolution Control Standard n-P on Rtx®-BAC Plus 1 and Plus 2 Column Set Rtx®-BAC Plus 1 Peaks 2 1. 2. 3. 4. 5. 6. New resolution control standard! Conc. (μg/mL) 50 50 50 50 50 50 Methanol Acetaldehyde Ethanol…
Key words
acetaldehyde, acetaldehydeisopropanol, isopropanolacetone, acetoneethanol, ethanolnotes, notesheadspace, headspaceloop, loopmethanol, methanoloven, ovenpeaks, peakssample, samplecarrier, carriercolumns, columnsnew, newresolution
Inhalants on Rtx®-BAC Plus 1 and Plus 2 Column Set
Inhalants on Rtx®-BAC Plus 1 and Plus 2 Column Set
|Agilent Technologies|Applications
Inhalants on Rtx®-BAC Plus 1 and Plus 2 Column Set Rtx®-BAC Plus 1 16 15 12 17, 18 4 19 7 6 5 9 3 2 1 2 14 13 8 10 11 4 6 8 10 Rtx®-BAC Plus 2…
Key words
make, makenotes, notesgas, gasheadspace, headspaceloop, loopoven, ovenpeaks, peakscarrier, carriercolumns, columnswater, waterdetector, detectorinstrument, instrumentinjection, injectionsample
Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns
Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns
2018|Agilent Technologies|Brochures and specifications
Restek GC LC Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns Advanced Technology for Fast, Reliable Measurement of Alcohol in Blood • Optimized column selectivities guarantee resolution of ethanol, internal standards, and frequently encountered interferences. • Robust and reproducible column…
Key words
restek, restekleak, leakpeaks, peaksadaptor, adaptoruniversal, universalalcohol, alcoholglycol, glycolessentials, essentialscharger, chargertailing, tailingacetone, acetoneblood, blooddescription, descriptionthread, threadbutanol
Blood Alcohol Analysis on Rtx®-BAC Plus 1
Blood Alcohol Analysis on Rtx®-BAC Plus 1
|Agilent Technologies|Applications
Blood Alcohol Analysis on Rtx®-BAC Plus 1 2 - Full resolution of all peaks, including t-butanol and acetone! - Run time less than 3 minutes! - Increased resolution of ethanol from other peaks of interest. 1. 2. 3. 4. 5.…
Key words
courtsey, courtseyheadspace, headspacemake, makeacetaldehyde, acetaldehydebutanol, butanolconcentrator, concentratorisopropanol, isopropanoltert, tertacetone, acetoneethanol, ethanolgas, gasloop, loopmethanol, methanoloven, oventransfer
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike