Halogenated aromatic hydrocarbons - Separation of fluorotoluene isomers
Applications | 2011 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Importance of the topic
GC analysis of fluorotoluene isomers is critical because these compounds serve as key intermediates in the synthesis of plant protection agents. Due to their similar physical and chemical properties, conventional polar GC phases fail to resolve the ortho-, meta- and para-isomers, creating challenges for accurate purity assessment and quality control of raw materials.Objectives and Study Overview
The primary objective was to achieve baseline separation of 2-, 3- and 4-fluorotoluene using a chiral-selective stationary phase under isothermal conditions. The study evaluates the robustness of this approach for low-level (<0.1%) sample analysis relevant to routine environmental and industrial laboratories.Methodology
- Technique: Capillary gas chromatography in split injection mode
- Column temperature: constant 30 °C
- Carrier gas: nitrogen at 100 kPa (1 bar)
- Injection: split ratio ~1:20, injector temperature 150 °C, 0.2 µL sample volume
- Detection: flame ionization detector at 150 °C
- Sample preparation: fluorotoluene isomers dissolved in methanol at approximately 0.1% (w/v)
Used Instrumentation
- Gas chromatograph equipped with capillary inlet and FID
- Agilent CP-Chirasil-Dex CB column (0.25 mm × 25 m, 0.25 µm film thickness)
- Nitrogen gas supply regulated at 100 kPa
Main Results and Discussion
Baseline separation of the three fluorotoluene isomers was successfully achieved under isothermal conditions. The elution order was 2-fluorotoluene first, followed by 3-fluorotoluene and then 4-fluorotoluene. Well-shaped, symmetrical peaks and clear resolution demonstrated the superiority of the chiral stationary phase over standard polar phases, which could not resolve these isomers.Benefits and Practical Applications
- Ensures precise purity evaluation of agrochemical intermediates
- Enables trace-level isomer discrimination for quality control
- Applicable to reaction monitoring and environmental analysis of halogenated aromatics
Future Trends and Opportunities
The successful separation suggests wider application of chiral GC phases for other halogenated aromatic isomer mixtures. Future work may include temperature programming to reduce analysis time, coupling with mass spectrometry for enhanced identification, and two-dimensional GC approaches to tackle more complex mixtures.Conclusion
The optimized GC method employing the CP-Chirasil-Dex CB column delivers reliable baseline separation of fluorotoluene isomers, meeting the stringent requirements of analytical and quality control laboratories in the agrochemical and environmental sectors.Reference
- Agilent Technologies, Application Note A01445, October 2011
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Progress in Capillary GC Columns and Inlet Inertness Performance
2011|Agilent Technologies|Posters
Progress in Capillary GC Columns and Inlet Inertness Performance Kenneth G. Lynam, Doris Smith, and Limian Zhao Agilent Technologies, Inc. 2850 Centerville Rd. Wilmington, DE USA 19808 Abstract Experimental ¾ Highly sensitive gas phase analyses demand better inertness performance from…
Key words
inert, inertultra, ultraliners, linerswool, woolorganophosphorus, organophosphorusinertness, inertnesspesticides, pesticidesdeactivated, deactivatedoxycodone, oxycodoneliner, linerapple, appleagilent, agilentfish, fishoxymorphone, oxymorphoneisomers
Aromatic hydrocarbons, C6-C8 - High resolution separation of xylene isomers
2011|Agilent Technologies|Applications
Aromatic hydrocarbons, C6-C8 High resolution separation of xylene isomers Application Note Environmental Authors Introduction Agilent Technologies, Inc. The cyclodextrin modification of the Agilent CP-Chirasil DEX CB stationary phase makes high resolution separation of xylene isomers possible. The separation of para-…
Key words
xylene, xylenechirasil, chirasilbenzene, benzenedex, dexcyclodextrin, cyclodextrinmeta, metacumene, cumeneortho, orthopara, paracourtesy, courtesymodification, modificationstyrene, styreneisomers, isomerstoluene, toluenestationary
Separation of cis-trans FAME isomers
2011|Agilent Technologies|Applications
Separation of cis-trans FAME isomers Application Note Materials Testing & Research Authors Introduction Jan Peene Agilent CP-Select CB for FAME is a highly polar capillary column with comparable selectivity for FAME as the Agilent CP-Sil 88. The bonded phase can…
Key words
fame, famepeene, peenejan, jantrans, transcis, cisisomers, isomerscomparable, comparablepolar, polarselect, selectprinted, printedcolumn, columninjector, injectorresearch, researchusa, usatesting
PCBs - Analysis of regulation-relevant PCB in fats
2011|Agilent Technologies|Applications
PCBs Analysis of regulation-relevant PCB in fats Application Note Food Testing & Agriculture Authors Introduction Agilent Technologies, Inc. For the separation of PCB the Agilent CP-Sil 8 CB phase provides highest selectivity. This column is commonly used as the industry…
Key words
fats, fatspcb, pcbpcbs, pcbscourtesy, courtesyregulation, regulationagriculture, agriculturenetherlands, netherlandsrelevant, relevantprinted, printedinjector, injectorusa, usatesting, testingauthors, authorsfood, foodsize