Routine analysis of purgeable organic compounds in drinking water with ISQ 7000 GC-MS
Applications | 2020 | Thermo Fisher ScientificInstrumentation
The monitoring of purgeable organic compounds POCs in drinking water is critical to ensure regulatory compliance and public health safety. VOCs such as halogenated hydrocarbons and ethers can pose health risks at trace levels. EPA Method 524.4 defines criteria for accurate quantitation of these compounds in environmental laboratories.
This study demonstrates a streamlined routine method that meets EPA Method 524.4 requirements using purge and trap P&T with GC MS detection. Key performance metrics including linearity range 0.2 to 50 ppb, method detection limits MDL, precision, accuracy, minimum reporting level MRL confirmation, and robustness over multiple injections are evaluated.
Sample preparation involved a 25 ppm calibration in methanol and nine point calibration from 0.2 to 50 ppb using internal standards and surrogate standards. P&T parameters were optimized to transfer VOCs while minimizing moisture. GC temperature programming enabled sub-15-minute runs and split injection. MS was operated in full scan mode 35 to 260 amu to achieve sensitivity and compound confirmation.
Excellent chromatography was achieved with minimal water interference and consistent peak shapes at low ppb levels. Linearity R2 ≥ 0.995 was confirmed for all 75 compounds. MDLs below 0.25 ppb met regulatory goals. MRL confirmations fell within the 50 to 150% prediction intervals for most analytes. Precision and accuracy for n = 7 at 5 ppb met EPA criteria with RSD values below 20% and recovery between 96 and 120%. Robustness testing over 120 injections showed RSDs below 9%, demonstrating method stability.
Advances in automation and real-time data analytics will further enhance routine VOC monitoring. Integration with high-resolution mass spectrometry may expand compound coverage and sensitivity. Ongoing development of miniaturized P&T modules and predictive maintenance tools will support accelerated environmental testing and field deployable systems.
The combined Atomx XYZ P&T, Thermo Scientific TRACE 1310 GC and ISQ 7000 MS platform under Chromeleon CDS control delivers a robust, sensitive, and high-throughput solution for routine analysis of purgeable organic compounds in drinking water. It meets or exceeds all EPA Method 524.4 performance criteria, ensuring reliable monitoring for environmental laboratories.
GC/MSD, Purge and Trap, GC/SQ
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific, Teledyne LABS
Summary
Significance of the Topic
The monitoring of purgeable organic compounds POCs in drinking water is critical to ensure regulatory compliance and public health safety. VOCs such as halogenated hydrocarbons and ethers can pose health risks at trace levels. EPA Method 524.4 defines criteria for accurate quantitation of these compounds in environmental laboratories.
Objectives and Study Overview
This study demonstrates a streamlined routine method that meets EPA Method 524.4 requirements using purge and trap P&T with GC MS detection. Key performance metrics including linearity range 0.2 to 50 ppb, method detection limits MDL, precision, accuracy, minimum reporting level MRL confirmation, and robustness over multiple injections are evaluated.
Methodology and Instrumentation
Sample preparation involved a 25 ppm calibration in methanol and nine point calibration from 0.2 to 50 ppb using internal standards and surrogate standards. P&T parameters were optimized to transfer VOCs while minimizing moisture. GC temperature programming enabled sub-15-minute runs and split injection. MS was operated in full scan mode 35 to 260 amu to achieve sensitivity and compound confirmation.
Used Instrumentation
- Teledyne Tekmar Atomx XYZ purge and trap system
- Thermo Scientific TRACE 1310 gas chromatograph
- Thermo Scientific ISQ 7000 single quadrupole mass spectrometer with VPI and ExtractaBrite ion source
- Chromeleon 7.2 chromatography data system for unified instrument control and data processing
- TraceGOLD TG-VMS column 20 m × 0.18 mm × 1 µm film
Main Results and Discussion
Excellent chromatography was achieved with minimal water interference and consistent peak shapes at low ppb levels. Linearity R2 ≥ 0.995 was confirmed for all 75 compounds. MDLs below 0.25 ppb met regulatory goals. MRL confirmations fell within the 50 to 150% prediction intervals for most analytes. Precision and accuracy for n = 7 at 5 ppb met EPA criteria with RSD values below 20% and recovery between 96 and 120%. Robustness testing over 120 injections showed RSDs below 9%, demonstrating method stability.
Benefits and Practical Application
- Single software control simplifies operation and reduces training requirements
- High sample throughput with sub-15-minute GC runs and flexible P&T parameters
- Reliable protection of GC column from moisture increases column lifetime
- Regulatory compliance assured through validated performance metrics
- Flexibility to adapt method parameters to laboratory workload and target analytes
Future Trends and Opportunities
Advances in automation and real-time data analytics will further enhance routine VOC monitoring. Integration with high-resolution mass spectrometry may expand compound coverage and sensitivity. Ongoing development of miniaturized P&T modules and predictive maintenance tools will support accelerated environmental testing and field deployable systems.
Conclusion
The combined Atomx XYZ P&T, Thermo Scientific TRACE 1310 GC and ISQ 7000 MS platform under Chromeleon CDS control delivers a robust, sensitive, and high-throughput solution for routine analysis of purgeable organic compounds in drinking water. It meets or exceeds all EPA Method 524.4 performance criteria, ensuring reliable monitoring for environmental laboratories.
References
- EPA Method 524.4 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography Mass Spectrometry
- Thermo Fisher Scientific AppsLab Library for EPA Method 524.4 application notes
- Thermo Scientific Application Note 65632 Routine analysis of volatile organic compounds in drinking water with ISQ 7000 GC MS
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
US EPA Method 524.4 using the Teledyne Tekmar Atomx XYZ and Thermo Scientific™ ISQ™ 7000 Mass Spectrometry (MS) System Coupled with a Thermo Scientific™ TRACE™ 1310 Gas Chromatograph (GC)
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN2002 P+T VOC ENV US EPA Method 524.4 using the Teledyne Tekmar Atomx XYZ and Thermo Scientific™ ISQ™ 7000 Mass Spectrometry (MS) System Coupled with a Thermo Scientific™ TRACE™ 1310 Gas Chromatograph (GC) Amy Nutter, Applications…
Key words
temp, tempbake, bakepurge, purgerinse, rinseavg, avgsweep, sweeptert, tertatomx, atomxether, etherdesorb, desorblpir, lpirupir, upirppb, ppbtime, timesurr
US EPA Method 524.4 with the Tekmar Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ 7000 MS System with an ExtractaBrite Source
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN2005 P+T VOC ENV US EPA Method 524.4 with the Tekmar Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ 7000 MS System with an ExtractaBrite Source Amy Nutter, Applications Chemist; Teledyne Tekmar P…
Key words
temp, tempbake, bakepurge, purgerinse, rinsetert, tertlpir, lpirupir, upirppb, ppbsweep, sweepatomx, atomxdesorb, desorbether, ethertime, timeprecision, precisiondemonstration
Sustainable uncompromised performance: analysis of volatile organic compounds in drinking water with the ISQ 7610 GC-MS using HeSaver-H2 Safer technology
2023|Thermo Fisher Scientific|Applications
Application note | 001773 Environmental Sustainable uncompromised performance: analysis of volatile organic compounds in drinking water with the ISQ 7610 GC-MS using HeSaver-H2Safer technology Goal Authors Adam Ladak1, Terry Jeffers1, and Demonstration of a routine analytical method that meets the…
Key words
lin, linavgcalfact, avgcalfactbake, bakeether, etherpurge, purgebutylbenzene, butylbenzenehelium, heliummethyl, methyltemperature, temperaturebutyl, butylsweep, sweeprinse, rinseprecision, precisiondesorb, desorbmdl
US EPA Method 524.3 with the Teledyne Tekmar Atomx XYZ and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ 7000 MS with an Advanced Electron Ionization (AEI) Source
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1805 P+T VOC ENV US EPA Method 524.3 with the Teledyne Tekmar Atomx XYZ and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ 7000 MS with an Advanced Electron Ionization (AEI) Source Amy Nutter, Applications Chemist; Teledyne…
Key words
temp, tempbake, bakepurge, purgerinse, rinseaccuracy, accuracylpir, lpirupir, upirtert, tertsweep, sweepmidpoint, midpointether, etherdesorb, desorbatomx, atomxtime, timeprecision