The GC/MS Analysis of Impurities in Denatured Ethanol on the SPBä-1 Capillary Column

Applications | 2000 | MerckInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Merck

Summary

Significance of the Topic


Denatured ethyl alcohol (SDA) is a widely applied industrial solvent in chemical, pharmaceutical and QA/QC processes. Trace levels of organic impurities can compromise product quality, safety and regulatory compliance. Accurate profiling of these impurities is therefore essential for process control and assurance of solvent performance.

Objectives and Study Overview


This work describes the development of a gas chromatography–mass spectrometry (GC/MS) method for the identification and quantification of 19 common organic impurities in SDA-3A (ethanol denatured with 5% methanol). A key challenge is preventing overload of the mass spectrometer by the ethanol/methanol solvent front. The strategy employs a nonpolar SPB-1 capillary column to elute solvents ahead of target analytes and implements a solvent delay to protect the detector.

Methodology


Samples were injected on-column (1 µL, split 50:1) into a GC equipped with a 60 m × 0.25 mm ID, 1.0 µm SPB-1 column. Helium was used as carrier gas at 1 mL/min constant flow. The oven was programmed from 45 °C (6 min hold) to 115 °C at 5 °C/min, then to 150 °C at 20 °C/min. Injector temperature was set to 130 °C. The MSD interface was maintained at 220 °C, scanning 30–150 amu. A solvent delay was applied until after ethanol and methanol elution to prevent MS source overload.

Used Instrumentation


  • SPB-1 fused silica capillary column, 60 m × 0.25 mm ID, 1.0 µm film thickness
  • GC injector: split mode, 1 µL injection, 50:1 split ratio
  • Carrier gas: helium, constant flow at 1 mL/min
  • Oven program: 45 °C hold 6 min → 115 °C at 5 °C/min → 150 °C at 20 °C/min
  • MSD interface temperature: 220 °C; scan range: 30–150 amu; applied solvent delay

Main Results and Discussion


The method achieved separation of 19 target impurities including ketones, alcohols, esters and ethers. All analytes were at least 70% resolved; isoamyl alcohol and 1,1-diethoxyethane exhibited partial coelution but were differentiated by unique mass spectra. On-column detection limits ranged from 0.30 to 1 ng. Inertness of the SPB-1 stationary phase minimized adsorption and maintained peak shape. Over a nine-day repeat injection study, no significant retention time shifts were observed, demonstrating excellent column stability for routine analyses.

Benefits and Practical Applications


The described GC/MS approach offers high sensitivity, selectivity and reproducibility for trace impurity profiling in denatured ethanol. The solvent delay protects the MSD source from overload, extending instrument uptime. This method is well suited for quality control, regulatory testing and production monitoring in industrial laboratories.

Future Trends and Potential Applications


Opportunities for further development include adapting the method to other denaturant compositions and complex solvent matrices. Coupling with high-resolution or two-dimensional GC–MS (GC×GC–MS) could enhance separation of coeluting compounds. Integration with automated sampling and digital data handling will support high-throughput environments in pharmaceutical and environmental laboratories.

Conclusion


The 60 m × 0.25 mm ID, 1.0 µm SPB-1 capillary column combined with a tailored GC/MS method provides reliable, robust and sensitive detection of organic impurities in SDA-3A. The solvent delay approach effectively shields the MS detector from solvent overload while ensuring full characterization of analytes, meeting the rigorous demands of industrial quality assurance.

References


No external literature references were specified in the original document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
The Analysis of Alcoholic Beverages on the 30m x 0.25mm ID, 1.0μm SPB-20 Capillary Column
Supelco Park Bellefonte, PA 16823-0048 USA Telephone 800-247-6628 • 814-359-3441 Fax 800-447-3044 • 814-359-3044 email: supelco@s ial.com http://www.sigma-aldrich.com Application Note 164 The Analysis of Alcoholic Beverages on the 30m x 0.25mm ID, 1.0µm SPBä-20 Capillary Column Traditional methods for monitoring…
Key words
alcohol, alcoholisoamyl, isoamylamyl, amylacetate, acetatebutanol, butanolfusel, fuselpropanol, propanolacetaldehyde, acetaldehydearoma, aromadhy, dhymin, miniso, isoacetic, aceticacetone, acetonebutyl
Resolving Industrial Solvent Mixtures on Packed and Capillary GC Columns
Bulletin 824C Resolving Industrial Solvent Mixtures on Packed and Capillary GC Columns Nonpolar SPB-1 and polar SUPELCOWAX 10 capillary columns separate complex mixtures of most commonly used solvents with good resolution. Resolution on 0.53mm ID columns is similar to that…
Key words
acetate, acetatecellosolve, cellosolvelate, latebutyl, butylethyl, ethylisobutyl, isobutylalcohol, alcoholmethyl, methylxylene, xyleneketone, ketonepropyl, propylisopropyl, isopropyltoluene, toluenesolvents, solventsmesityl
Organic Volatile Impurities in Pharmaceutical Products: Selectivity of Capillary GC Columns
Application Note 70 Organic Volatile Impurities in Pharmaceutical Products: Selectivity of Capillary GC Columns Methods for separating and quantifying organic volatile impurities (OVIs) – residual solvents potentially present in pharmaceutical preparations – are described by the US Pharmacopoeia in USP…
Key words
dipole, dipoleethyl, ethylmibk, mibkmek, mekelution, elutionether, etherbutanol, butanolpyridine, pyridinemtbe, mtbeisopropanol, isopropanoldispersive, dispersiveacetate, acetateacetone, acetonetrichloroethylene, trichloroethyleneethanol
Organic Volatile Impurities in Pharmaceutical Products: Selectivity of Capillary GC Columns
TheReporter Reprinted from Volume 14, No. 2, 1995 T295012 © 1999 Sigma-Aldrich Co. For more information, or current prices, contact your nearest Supelco subsidiary listed below. To obtain further contact information, visit our website (www.sigma-aldrich.com), see the Supelco catalog, or…
Key words
dipole, dipoleethyl, ethylmibk, mibkmek, mekether, etherelution, elutiontrichloroethylene, trichloroethyleneacetate, acetatebutanol, butanolheptane, heptanepyridine, pyridinemtbe, mtbetetrahydrofuran, tetrahydrofuranisopropanol, isopropanoldispersive
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike