GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

REAL-TIME ANALYSIS OF UNITED STATES EPA METHOD TO-14A COMPOUNDS USING SIFT-MS

Posters |  | Syft TechnologiesInstrumentation
SIFT-MS
Industries
Environmental
Manufacturer
Syft Technologies

Summary

Importance of the Topic


Volatile organic compounds regulated under US EPA TO-14A are key indicators of air quality, with implications for human health and environmental protection. Traditional canister collection followed by GC/MS analysis offers high sensitivity but suffers from delayed results and complex sample preparation. Real-time quantitation of TO-14A VOCs using SIFT-MS addresses these challenges by providing rapid, ultra-trace detection without preconcentration, enabling timely decision making in ambient and industrial monitoring.

Objectives and Study Overview


This study aimed to develop and validate a rapid SIFT-MS method for the comprehensive analysis of TO-14A target compounds. A Syft Technologies Voice200ultra instrument was calibrated against certified TO-14A standard mixtures. Performance metrics such as limits of detection, linear dynamic range, and selectivity were assessed. Field deployment at three sites near Auckland, New Zealand, demonstrated continuous monitoring capabilities.

Instrumentation Used


  • Syft Technologies Voice200ultra SIFT-MS system
  • Reagent ions: H3O+, NO+ and O2+
  • LabSyft Method Editor software and Syft compound library
  • Flow-past sampling via Teflon tubing and GAST pump
  • Certified TO-14A standard mixtures in six-litre passivated SUMMA canisters

Methodology


SIFT-MS employs controlled soft ionization within a flow tube, switching reagent ions in under 10 ms to react selectively with analytes. The resulting product ions are filtered by quadrupole mass analyzers and quantified without any preconcentration step. Calibration was achieved using certified gas standards at nominal concentrations of 1.5, 2.5 and 5 ppbv, establishing detection limits in the low part-per-trillion range and demonstrating excellent linearity over multiple orders of magnitude.

Key Results and Discussion


Most TO-14A compounds were detected with low- to mid-pptv quantitation limits and strong linearity. Limited selectivity was observed for certain small halogenated species due to overlapping product ions (e.g., CCl2+ and CCl2F+). Continuous two-week monitoring at three Auckland sites revealed temporal patterns in biogenic emissions and anthropogenic aromatics. Data correlation with meteorological parameters and co-pollutants such as NOx and ozone is pending validation.

Benefits and Practical Applications


  • Enables rapid preliminary screening prior to regulatory GC/MS analysis, reducing turnaround time and costs
  • Supports continuous ambient air quality assessment and fence-line monitoring at industrial sites
  • Operates without sample preparation or preconcentration, simplifying workflow
  • Offers broad dynamic range and ease of method adaptation for various VOC profiles

Future Trends and Opportunities


Ongoing development of negatively charged reagent ion chemistry is expected to enhance isomeric resolution for challenging halogenated species. Integration with meteorological sensors and networked data platforms could facilitate real-time mapping of VOC emissions. Advances in instrument miniaturization may lead to portable SIFT-MS units for on-site screening and distributed sensor deployments.

Conclusion


The SIFT-MS approach delivers real-time, ultra-trace quantitation of US EPA TO-14A VOCs with sensitivity comparable to established GC/MS methods, while eliminating sample preparation and enabling continuous monitoring. Further reagent ion innovations and extended field validation will broaden its utility for environmental surveillance and regulatory applications.

References


  1. Prince, B.J., Milligan, D.B. & McEwan, M.J. Rapid Commun. Mass Spectrom., 2010, 24, 1763–1769.
  2. Langford, V.S., Graves, I. & McEwan, M.J. Rapid Commun. Mass Spectrom., 2014, 28, 10–18.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Real-Time Ambient Air Monitoring Campaigns Using Syft SIFT-MS
Syft Technologies | Application Note | Environmental Analysis Real-Time Ambient Air Monitoring Campaigns Using Syft SIFT-MS Key Benefits Analytical results in seconds Detection limits in the pptv range Direct air analysis without preconcentration or drying Readily deployed to site and…
Key words
syft, syftsift, siftdeployed, deployedpptv, pptvvocs, vocsreal, realair, airppbv, ppbvbrochure, brochuremonitoring, monitoringanalysis, analysispreconcentration, preconcentrationtechnologies, technologiesgusty, gustyper
EXTENDING SIFT-MS TO DETECTION OF SEMI-VOLATILES ON SURFACES
EXTENDING SIFT-MS TO DETECTION OF SEMI-VOLATILES ON SURFACES Daniel B. Milligan, Yan Li, John D.C. Gray, Barry J. Prince, Vaughan S. Langford, Murray J. McEwan Syft Technologies Limited, Christchurch, New Zealand Email: [email protected] Introduction Results and Discussion Selected Ion Flow…
Key words
swab, swabppbv, ppbvquantity, quantityonto, ontospiked, spikeddetected, detectedsift, siftconcentration, concentrationcwa, cwananogram, nanogramdesorber, desorberion, ionsvocs, svocsreagent, reagentpseuodoephedrine
SOLUTIONS THAT PROTECT PEOPLE - SYFT TECHNOLOGIES
SOLUTIONS THAT PROTECT PEOPLE - SYFT TECHNOLOGIES
2017|Syft Technologies|Brochures and specifications
SOLUTIONS THAT PROTECT PEOPLE SYFT TECHNOLOGIES SIFT-MS SOLUTIONS THAT PROTECT PEOPLE Many volatile organic compounds (VOCs) and semi-VOCs (SVOCs) pose risks to human health and the general well-being of modern society. Some have severe, acute toxicity, such as the chemical…
Key words
sift, siftsyft, syftvocs, vocsfumigants, fumigantsreagent, reagenttoxic, toxicindoor, indoorair, airwarfare, warfaredetection, detectionrapidly, rapidlytechnologies, technologiesanalysis, analysisemissions, emissionsorganosulfur
AUTOMATION APPLICATIONS - SYFT TECHNOLOGIES
AUTOMATION APPLICATIONS - SYFT TECHNOLOGIES
2017|Syft Technologies|Brochures and specifications
AUTOMATION APPLICATIONS SYFT TECHNOLOGIES AUTOMATION APPLICATIONS OF SIFT-MS Automation has been widely utilized by traditional chromatographic techniques for many years, enabling instrumentation to be productive 24/7 – a critical consideration because sample throughput is typically only 20 to 60 samples…
Key words
sift, siftsyft, syftheadspace, headspaceanalysis, analysisvocs, vocsautomation, automationmhe, mhesha, shareagent, reagenttrillion, trillionvery, veryapplications, applicationsgases, gasesresidual, residualrapid
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike