Improved Inertness and QC Testing Standards for EPA Method 8270D with Zebron™ ZB-SemiVolatiles GC Columns

Applications | 2013 | PhenomenexInstrumentation
GC, GC columns, Consumables
Industries
Environmental
Manufacturer
Phenomenex

Summary

Importance of the Topic


Environmental analysts using EPA 8270D for semi-volatile compounds face challenges due to active sites on general-purpose GC columns. Acidic phenols, basic amines, and reactive pesticides can cause peak tailing, reduced sensitivity and calibration failures, leading to time-consuming column maintenance and trim steps. Inertness and consistent performance are therefore critical to reliable quantitation, especially at low analyte levels.

Objectives and Study Overview


This study evaluates the inertness and QC testing of the newly developed Zebron ZB-SemiVolatiles GC column compared to common alternatives. Key goals include assessing column activity toward acids, bases and pesticides, defining minimum performance criteria for pyridine and pentachlorophenol, and validating improved calibration success under stringent test conditions.

Methodology and Instrumentation


The study applied a split injection (100:1) of a composite QC Test Mix at 20 ppm containing pyridine, pentachlorophenol, DFTPP internal standard, benzidine, and DDT. GC conditions featured:
  • 30 m x 0.25 mm x 0.25 μm columns operated with hydrogen carrier gas
  • Oven ramp from 40 °C to 300 °C at 15 °C/min
  • FID detection at 325 °C
Each column underwent comparative testing on the same instrument to generate response factors (RF) based on peak height ratios to DFTPP, following EPA 8270D conventions.

Results and Discussion


  • Zebron ZB-SemiVolatiles achieved RFs of 1.06 for pyridine and 0.37 for pentachlorophenol, exceeding newly established minimums of 0.6 and 0.3, respectively.
  • Competing columns (Restek Rxi-5ms, Rxi-5Sil MS, Agilent HP-5ms Ultra Inert, DB-5ms Ultra Inert) failed to meet one or both criteria, exhibiting lower RFs or DDT breakdown.
  • Pyridine peak shapes showed pronounced tailing on non-inert columns, indicating basic site interactions that can impair quantitation.
  • ZB-SemiVolatiles demonstrated minimal DDT degradation and stable responses at low-level injections, supporting robust calibration at trace levels.

Benefits and Practical Applications


The optimized Zebron ZB-SemiVolatiles column offers:
  • Balanced inertness for acids, bases and pesticides, reducing peak tailing and signal loss
  • Enhanced sensitivity at low analyte levels, improving calibration success without high-concentration workarounds
  • Extended column lifetime and reduced maintenance by minimizing contaminant buildup
  • Predictable performance through defined QC criteria aligned to EPA 8270D demands

Future Trends and Opportunities


Advancements may include:
  • Further refinement of stationary phase deactivation chemistries for universal inertness
  • Automation of column QC testing at the manufacturer level for guaranteed batch performance
  • Integration of real-time diagnostic tools to monitor column activity in-line
  • Expanded use of highly inert columns in emerging fields such as forensic and metabolomics applications

Conclusion


By establishing stringent QC assays and adopting Enviro-Inert technology, the Zebron ZB-SemiVolatiles column delivers reliable inertness and reproducible responses for challenging semi-volatile analyses. Its proven performance under harsh GC conditions ensures accurate quantitation, streamlined workflows and reduced system downtime.

Reference


No literature references were provided in the source document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
ZB-SemiVolatiles with Enviro-Inert™ Technology
ZB-SemiVolatiles with Enviro-Inert™ Technology
2014|Phenomenex|Brochures and specifications
™ ZB-SemiVolatiles with Enviro-Inert™ Technology This column has REDUCED TestAmerica’s DOWNTIME and INCREASED our PRODUCTIVITY “ “ — TestAmerica Laboratories, Inc. Buffalo www.phenomenex.com/GC Introducing the Zebron™ Experience Get More Than Just A Column When you choose Zebron, you get more…
Key words
semivolatiles, semivolatilesphenomenex, phenomenexzebron, zebronpcp, pcppyridine, pyridineenviro, enviromin, mininert, inertbhc, bhcprice, priceour, ourbenzo, benzoresponse, responsepolyimide, polyimideendosulfan
3 Ways To Choose All GC 5 Phases Are Not The Same!
3 Ways To Choose All GC 5 Phases Are Not The Same!
2016|Phenomenex|Brochures and specifications
3 Ways To Choose All GC 5 Phases Are Not The Same! • ZB-5 • ZB-5MSi • ZB-5HT Inferno™ • ZB-5ms • ZB-5MSPLUS™ • ZB-SemiVolatiles www.phenomenex.com/5Phase Choosing Your 5 Phase Is Easy Use the NEW GC Column Finder web app…
Key words
phenomenex, phenomenexphase, phasechemistry, chemistrysemivolatiles, semivolatilescolumn, columnprofile, profilecertified, certifiedsupelco, supelcozebron, zebronapplications, applicationsrestek, restekinertness, inertnesspolarity, polaritypesticides, pesticideshydrocarbons
Improved Analysis of EPA Method 8270D on a Zebron™ ZB-SemiVolatiles Column
TN-2049 APPLICATIONS Improved Analysis of EPA Method 8270D on a Zebron™ ZB-SemiVolatiles Column Kory Kelly Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA An improved method for EPA Method 8270D has been exhibited on a Zebron ZB-SemiVolatiles GC column.…
Key words
semivolatiles, semivolatileszebron, zebronfluoranthene, fluoranthenebenzo, benzoshape, shapecompounds, compoundstemplate, templatepeak, peakcolumn, columnshared, shareddeactivated, deactivatedcolumns, columnsions, ionsmin, minseparation
GC Product Guide 2025/26
GC Product Guide 2025/26
2025|Phenomenex|Brochures and specifications
GC Product Guide 2025/26 Welcome Table of Contents 2 07 Improved Performance Columns 34 High Temperature Columns 44 Standard Columns 52 GC ACCESSORIES 66 Inlet Base Seals 67 Installation Nuts 69 Ferrules 71 O-rings 73 Septa/Septum 73 Inlet Liners 75…
Key words
appendices, appendiceszebron, zebronaccessories, accessoriescolumns, columnscont’d, cont’dcontents, contentsphenomenex, phenomenextable, tablecolumn, columnordering, orderingpolarity, polarityguard, guardphase, phasesyringes, syringessemivolatiles
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike