GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Seminar GC Troubleshooting

Presentations | 2014 | MerckInstrumentation
GC
Industries
Manufacturer
Merck

Summary

Importance of Topic


Gas chromatography (GC) remains a cornerstone technique in analytical chemistry for separation and quantitation of volatile and semi-volatile compounds. Effective troubleshooting ensures reliable results, reduces downtime, extends column and instrument life, and secures data quality in research, industrial QA/QC, environmental monitoring, and food safety laboratories.

Objectives and Overview


This summary captures a structured approach to GC troubleshooting, focusing on three main system components—injection port, column, and detector—and illustrates the strategy with practical case studies. It aims to guide users through logical fault isolation, common error sources, corrective actions, and preventive maintenance.

Methods and Instrumentation


General troubleshooting workflow:
  • Review prior records and instrument logs before disassembly.
  • Verify method parameters and instrument condition; compare against a reference standard.
  • Isolate the fault by substituting a test column or swapping detector/inlet.
  • Change one variable at a time to identify the true cause.

Key methodological considerations:
  • Carrier gas purity and use of inline purifiers to reduce baseline noise and extend column life.
  • Injection technique: syringe size (10–50% fill), injection speed, and reproducibility via solvent flush or adapters.
  • Liner selection: split vs. splitless liners, internal packing (glass/quartz wool), deactivation, and internal volume suitable for solvent vapor clouds.
  • Splitless vent timing (0.5 – 2 inlet volumes; 30 s – 20 min splitless time) to balance analyte retention and solvent purge.
  • Pressure pulsed injections (temporary headpressure increase) to contain large vapor clouds and boost transfer efficiency.
  • Column parameters: inner diameter, film thickness, length, and installation height to optimize efficiency, selectivity, and peak shape.
  • Detector upkeep: clean FID collectors and MSD ion source, replace septa and seals to prevent noise, bleed, and active sites.

Main Results and Discussion


Carrier gas experiments showed elevated baseline noise when using tank hydrogen vs. generator gas, confirming contamination as a bleed source. Inline purifiers and indicating traps effectively restored a stable baseline.

Injection port studies demonstrated that slow manual injection produced double peaks, while rapid, smooth delivery yielded sharp single peaks. Splitless experiments revealed that extended splitless times and higher inlet temperatures enhanced response of heavy analytes but required liner volume and solvent focusing adjustments. Pressure pulsing further improved signal but risked solvent-front distortion without liner modifications.

Column trials highlighted that reducing column ID from 0.53 mm to 0.10 mm increased plate number and resolution, allowing shorter columns. Thicker stationary phase films improved solvent-front focusing for early eluters but reduced separation of late eluting compounds. Incorrect column installation height led to peak tailing and lower response for heavy analytes. Clipping contaminated inlet ends restored analyte signals.

Detector case studies showed that a dirty FID produced spiky noise, while a fouled MS source reduced pesticide response after high-matrix injections. Routine cleaning and part replacement returned sensitivity to expected levels.

Benefits and Practical Applications


  • Systematic troubleshooting reduces guesswork and preserves instrument integrity.
  • Optimized injection and column parameters enhance sensitivity, resolution, and reproducibility.
  • Effective gas purification and detector maintenance sustain baseline stability and accurate quantitation.
  • Case study approaches translate to diverse matrices in food safety, environmental, and pharmaceutical analyses.

Future Trends and Potential Uses


Emerging directions include automated diagnostic software for fault prediction, advanced interface deactivation chemistries, real-time instrument health monitoring using IoT sensors, and machine learning algorithms to accelerate problem identification and optimize GC–MS methods.

Conclusion


A disciplined, stepwise approach to GC troubleshooting—prioritizing parameter checks, reference standards, and isolated component swaps—prevents unnecessary changes and accelerates resolution. Focus on injection port integrity, column health, and detector cleanliness is critical. Maintaining detailed logs and routine preventive maintenance ensures long-term performance and data reliability.

Použitá instrumentace


  • Gas chromatograph with electronic pressure control (EPC)
  • Carrier gas supply: hydrogen (tank and generator), purified with inline traps
  • Injection port: split and splitless liners (glass, quartz wool, tapered designs), pressure pulsing capability
  • Columns: various IDs (0.10 – 0.53 mm), film thicknesses (0.18 – 1 µm), phases (5% phenylmethyl siloxane)
  • Detectors: flame ionization detector (FID), electron capture detector (ECD), mass spectrometer (MS) with full-scan capability
  • Consumables: low-bleed septa (pre-drilled options), gold-plated inlet seals, purifier cartridges

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Better Connections - Agilent GC inlet resource guide
Better Connections, Better Results with Your Agilent GC System GC inlet resource guide Here’s What’s Inside Introduction Septa Troubleshooting Ordering Information Ferrules Troubleshooting Ordering Information 2 3 4 4 7 8 9 13 Column Nuts Ordering Information Liners 14 15…
Key words
liner, linerproblem, probleminlet, inletferrules, ferrulescolumn, columnnormal, normalgraphite, graphitenut, nutliners, linersnuts, nutssepta, septaferrule, ferruleagilent, agilentwool, woolreplace
Basic Capillary GC Theory and Practical Troubleshooting
Basic Capillary GC Theory and Practical Troubleshooting Part 4, Troubleshooting Katherine K. Stenerson Supelco, 595 North Harrison Road, Bellefonte, PA 16823 T404077 HCE GC Troubleshooting: Topics • Basic Troubleshooting Strategy • Preventing Problems - Gas Purification Injection Technique Liner Selection…
Key words
bleed, bleedliner, linercolumn, columninjection, injectionpoor, poortroubleshooting, troubleshootingsupelco, supelcogas, gaspeaks, peaksmissing, missingstrategy, strategytechnique, techniquenonlinearity, nonlinearitycarrier, carrierpurification
Capillary GC Troubleshooting: a Practical Approach
SUPELCO Capillary GC Troubleshooting: a Practical Approach Katherine Stenerson Gas Separations R&D Supelco, Supelco Park, Bellefonte, PA 16823 ©1999 Sigma-Aldrich Co. Outline ✔ Basic Troubleshooting Strategy ✔ Preventing Problems ✔ Identifying Common Problems ✔ Recommended Reading ✔ Discussion ©1999 Sigma-Aldrich…
Key words
supelco, supelcobleed, bleedcolumn, columnliner, linergas, gascarrier, carrierproblem, problempurifiers, purifiersinjection, injectionpoor, poorvisible, visibleguide, guideimproper, improperabundance, abundanceoverload
GC TROUBLESHOOTING GUIDE
GC TROUBLESHOOTING GUIDE
2017|Phenomenex|Guides
GC TROUBLESHOOTING GUIDE www.phenomenex.com/GC Access GC Resources 24/7 Explore Online Tips & Tools You asked for a better online experience. One filled with technical resources, easy-to-find products, and useful tools to make your GC work easier. Good news… it’s here!…
Key words
phenomenex, phenomenexsymptom, symptomremedy, remedycolumn, columnproblems, problemssuggested, suggestedcause, causepolar, polarpeak, peaktemperature, temperaturepossible, possiblebleed, bleedseptum, septumpeaks, peaksshape
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike