Installation Instructions for 5mm Uniliner® Inlet Liner with Adaptor in 1/4-inch Injection Port
Manuals | 2008 | RestekInstrumentation
Gas chromatography inlets require precise alignment and sealing when converting from packed to capillary columns. Specialized inlet liners and adaptors enable reproducible injections, optimal peak shapes, and leak-free performance.
This application note presents a detailed protocol for installing a 5 mm Uniliner inlet liner with a 1/4-inch adaptor. It covers both direct and on-column injection approaches to achieve superior chromatographic results via a secure, leak-tight interface and a controlled expansion chamber.
Installation comprises these key steps:
When properly installed, the system delivers:
Expected developments include:
Adherence to the outlined installation protocol for Uniliner inlet liners and adaptors ensures consistent, high-quality GC performance. Precise column cutting, correct adaptor assembly, and thorough leak testing are critical for reliable analytical outcomes.
GC columns
IndustriesManufacturerRestek
Summary
Significance of the Topic
Gas chromatography inlets require precise alignment and sealing when converting from packed to capillary columns. Specialized inlet liners and adaptors enable reproducible injections, optimal peak shapes, and leak-free performance.
Objectives and Study Overview
This application note presents a detailed protocol for installing a 5 mm Uniliner inlet liner with a 1/4-inch adaptor. It covers both direct and on-column injection approaches to achieve superior chromatographic results via a secure, leak-tight interface and a controlled expansion chamber.
Methodology and Used Instrumentation
Installation comprises these key steps:
- Select the injection mode (direct or on-column) and slide the Uniliner liner into the adaptor.
- Attach the adaptor to the injector body using a 1/4-inch stainless steel nut; tighten 1/4-turn past finger-tight.
- Trim each column end (~10 cm) with a ceramic wafer to ensure square cuts; mount 0.8 mm graphite ferrules and 1/16-inch nuts.
- Insert the column into the adaptor, maintain a 1 inch gap to the septum, then seat and tighten the ferrule to secure the column.
- Adjust the liner position by 1/32-inch and retighten the ferrule until the column is immobile under moderate force.
- Verify carrier gas flow (5–10 cc/min) and leak integrity using an electronic leak detector; avoid liquid leak sprays.
- 5 mm Uniliner inlet liner
- 1/4-inch Uniliner inlet liner adaptor
- Septum nut, injector body, and solvent expansion chamber
- Graphite ferrules and stainless steel nuts
- Fused silica capillary columns (0.32 mm or 0.53 mm ID)
- Electronic leak detector (Restek Leak Detector, cat.# 22839)
Main Results and Discussion
When properly installed, the system delivers:
- Leak-free connections robust to mechanical stress.
- Sharpened peak shapes with minimal tailing.
- Recommended head pressures for helium or hydrogen: 3–30 psig for 0.32 mm ID columns and 2–14 psig for 0.53 mm ID columns, varying by column length.
Benefits and Practical Applications
- Seamless switching between direct and on-column injections.
- Enhanced inertness using Siltek-treated liners to reduce analyte adsorption and background signals.
- Improved reproducibility and sensitivity for trace analysis in environmental, food safety, and petrochemical laboratories.
Future Trends and Potential Uses
Expected developments include:
- Next-generation deactivation coatings for ultra-low bleed and higher thermal stability.
- Smart inlet fittings with integrated sensors for real-time diagnostics.
- Miniaturized injection systems for microvolume analyses and high throughput workflows.
Conclusion
Adherence to the outlined installation protocol for Uniliner inlet liners and adaptors ensures consistent, high-quality GC performance. Precise column cutting, correct adaptor assembly, and thorough leak testing are critical for reliable analytical outcomes.
References
- Restek Online Column Installation Guide, Restek Inc.
- U.S. Patent 6,444,326 (Siltek Deactivation Process).
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Installing a Uniliner® Liner without an Adaptor in a Standard Packed Column Injection Port
2008|Restek|Manuals
Installing a Uniliner® Liner without an Adaptor in a Standard Packed Column Injection Port Overview These instructions are to be used ONLY if an adaptor cannot be used. A stainless steel Reducing Kit (cat.# 20312) or a reducing ferrule should…
Key words
uniliner, unilinercolumn, columninside, insideterminates, terminatesadaptor, adaptorliner, linerexpansion, expansionferrule, ferrulerestek, restekseptum, septumvisions, visionschamber, chamberinjection, injectionseating, seatingtouches
A Guide to Direct and On-column Flash Vaporization Injection
2002|Restek|Guides
Technical Guide A Guide to Direct and On-column Flash Vaporization Injection Inside: Overviews of direct and on-column flash vaporization injection techniques Advantages of direct injection over splitless injection Practical guidelines to maximize analysis accuracy and reproducibility Recommendations for converting a…
Key words
uniliner, unilinerinjection, injectiondirect, directliner, linersilcosteel, silcosteelsplitless, splitlessweldment, weldmentferrules, ferrulesgraphite, graphiteliners, linerscolumn, columninlet, inletferrule, ferruledescription, descriptioninstalls
Operating Hints for Using Split/Splitless Injectors
2002|Restek|Guides
Technical Guide 1 Operating Hints for Using Split/Splitless Injectors Inside: Overviews of split and splitless injection techniques Backpressure-regulated injection systems Headpressure-regulated injection systems Operating in the split injection mode Inlet liners for split injections Operating in the splitless injection Mode…
Key words
siltek, siltekliners, linersuniliner, unilinersplit, splitdeactivation, deactivationliner, linersplitless, splitlessinjection, injectionsplitter, splitterwool, woolinlet, inletgcs, gcsfeaturing, featuringport, porttight
GC Inlet Maintenance: Restek’s Quick-Reference Guide
2017|Agilent Technologies|Guides
Restek GC Accessories GC Inlet Maintenance: Restek’s Quick-Reference Guide • Help ensure quality data by selecting the right GC inlet supplies. • Prevent unwanted downtime with a preventative GC inlet maintenance program. • Easily order the GC inlet supplies you…
Key words
gcs, gcsinlet, inletdescription, descriptionliners, linersliner, linerkit, kitlength, lengthpuller, pullerwool, woolmle, mleplated, platedseptum, septumtool, toolsiltek, siltekmerlin