Evaluation of Tandem Mass Spectrometry for the Analysis of Environmental Samples

Applications | 2010 | BrukerInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Industries
Environmental
Manufacturer
Bruker

Summary

Importance of the topic


The analysis of trace organic contaminants in drinking water sources is critical for ensuring public health and regulatory compliance. EPA Method 525.2 covers over 100 analytes including pesticides, herbicides, plasticizers, polycyclic aromatic hydrocarbons, polychlorinated biphenyls, and other industrial chemicals. Complex matrices from surface and groundwater can generate elevated baselines and interferences in conventional GC-MS analyses, potentially leading to false positives or missed detections at regulatory levels.

Study objectives and overview


  • Compare tandem MS/MS acquisition with selected ion monitoring (SIM) for calibration linearity, precision, accuracy, and quantitation of over 100 target compounds spiked into a challenging groundwater extract according to EPA Method 525.2.
  • Evaluate performance in both pure solvent and real-world water matrix at low concentrations (0.01–1.0 ng/µL).

Methodology


Calibration solutions in ethyl acetate and matrix spikes at 50 ng/mL were prepared. A capillary GC method with a VF-5ms column (30 m × 0.25 mm × 0.25 µm) and a temperature program from 40 °C to 320 °C over 36 min was used. Both SIM and MS/MS acquisition modes on a triple quadrupole platform were assessed.

Instrumentation used


  • Bruker 300-MS triple quadrupole mass spectrometer
  • Bruker 450-GC gas chromatograph
  • Bruker 1177 S/SL injector with Siltek frit insert
  • CTC Combi PAL™ autosampler

Main results and discussion


  • Excellent calibration linearity (average r² > 0.998) and low detection limits (down to 0.01 ng/µL) were achieved for most analytes in both modes.
  • SIM acquisition suffered from significant matrix interferences, resulting in biased recoveries and elevated baselines for compounds such as BHC isomers, trifluralin, and others.
  • MS/MS acquisition effectively filtered co-extracted matrix components, delivering accurate and consistent quantitation in both solvent and groundwater extract with average recoveries of 118–126% in matrix and RSD < 8%.
  • False positives and threshold triggering observed in SIM were eliminated by the additional selectivity of tandem MS/MS.

Benefits and practical applications


  • Enhanced selectivity and reduced matrix effects improve confidence in compliance monitoring of public water supplies.
  • Lower quantitation limits enable early detection of emerging contaminants at sub-ppb levels.
  • Robust performance and wide analyte coverage streamline routine screening workflows in environmental and drinking water laboratories.

Future trends and opportunities


  • Adoption of high-resolution tandem MS and data-independent acquisition could further boost sensitivity and identification of unknowns.
  • Expansion of spectral libraries and automated processing will accelerate quantitative analysis of emerging pollutants.
  • Portable and miniaturized tandem MS/MS instruments may allow on-site water quality screening and real-time decision support.

Conclusion


Tandem MS/MS on a triple quadrupole mass spectrometer offers superior accuracy, precision, and selectivity compared to SIM for low-level analysis of organic contaminants in complex water matrices. This approach addresses matrix interferences and false positives, supporting reliable environmental monitoring and regulatory compliance under EPA Method 525.2.

Reference


  • Munch J.W. EPA Method 525.2: Determination of Organic Compounds in Drinking Water by Liquid–Solid Extraction and Capillary Column Gas Chromatography/Mass Spectrometry, Revision 2, 1995.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Determination of chemical contaminants in marine fish by GCMS/ MS using QuEChERS as an extraction method
Application Note GCMS-TQ8040 Determination of chemical contaminants in marine fish by GCMS/MS using QuEChERS as an extraction method No. GC-13-ADI-013 ‰ Introduction As Persistent Organic Pollutants (POPs) are environmental contaminants with a strong potential for bioaccumulation, these chemicals are expected…
Key words
loq, loqbhc, bhcendosulfan, endosulfanfluoranthene, fluoranthenebenzo, benzomrm, mrmchlordane, chlordanefish, fishheptachlor, heptachlormin, minanthracene, anthracenepyrene, pyrenebeta, betaalpha, alphatime
Extraction of Semi-volatile Organic Compounds in Drinking Water with NEW Atlantic® ReadyDisk C18 Solid Phase Extraction Disks
Application Note AN917 Extraction of Semi-volatile Organic Compounds in Drinking Water Using NEW Atlantic ReadyDisk C18 | Page 1 ® Extraction of Semi-volatile Organic Compounds in Drinking Water with NEW Atlantic ReadyDisk C18 Solid Phase Extraction Disks ® Authors: Andrew…
Key words
phthalate, phthalateoff, offbhc, bhcbenzo, benzomdl, mdldryvap, dryvapwere, wereheptachlor, heptachlorendrin, endrinendosulfan, endosulfananthracene, anthracenefluoranthene, fluoranthenesoak, soaktime, timeextraction
Automated Sample Preparation Using the PAL3 RTC System for EPA 8270E Semivolatile Organic Analysis by GC/TQ
Application Note Environmental Automated Sample Preparation Using the PAL3 RTC System for EPA 8270E Semivolatile Organic Analysis by GC/TQ Authors Abstract Gwen Lim Sin Yee, CTC Analytics AG, Zwingen, Switzerland Growing demand for SVOC analysis at laboratories reveals the analysis…
Key words
vortex, vortexlloq, lloqspiked, spikedrecovery, recoveryphthalate, phthalatematrix, matrixpreparing, preparinglle, llebhc, bhcautomated, automatedpreparation, preparationrsd, rsdsvocs, svocscalibration, calibrationanalyzing
Determination of chemical contaminants in marine fish by GCMS/MS using QuEChERS as an extraction method
PO-CON1550E Determination of chemical contaminants in marine fish by GCMS/MS using QuEChERS as an extraction method ASMS 2015 TP 338 Ankush Bhone, Durvesh Sawant, Dheeraj Handique, Prashant Hase, Sanket Chiplunkar, Ajit Datar, Jitendra Kelkar and Pratap Rasam Shimadzu Analytical (India)…
Key words
marine, marinefish, fishgcms, gcmsquechers, quecherscontaminants, contaminantsmrm, mrmextraction, extractionsmart, smartchemical, chemicalbhc, bhcmethod, methoddetermination, determinationpops, popsendosulfan, endosulfanfluoranthene
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike