Determination of C2-C12 aldehydes in water by SPME on-fiber derivatization and GC/MS
Applications | 2015 | CTC AnalyticsInstrumentation
Aldehydes are common environmental and biological compounds that influence water safety, oxidative stress markers, and food flavor quality. Monitoring C2–C12 aldehydes in drinking water is crucial for assessing disinfection by-products, ensuring compliance with health guidelines, and detecting off-flavor compounds in beverages packaged in PET.
The study aimed to develop and validate a sensitive, solvent-free method for simultaneous quantification of C2–C12 aldehydes in aqueous samples. On-fiber derivatization using O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine (PFBHA) was combined with PAL SPME Arrow extraction and GC/MS detection to achieve low detection limits and robust performance.
Derivatization and extraction were conducted in a 20 mL headspace vial at 60 °C for 30 min. Key steps included:
The method achieved:
This approach provides rapid, solvent-free sample preparation and robust quantitation suitable for routine monitoring of drinking water quality, beverage flavor profiling, and oxidative stress studies. The increased sorptive volume of the SPME Arrow enhances sensitivity and throughput in environmental and food analysis laboratories.
Future developments may include the use of selected ion monitoring (SIM) to lower detection limits further, adaptation to broader panels of carbonyl compounds, integration into automated workflows, and real-time monitoring in industrial and clinical settings.
The combination of on-fiber PFBHA derivatization with PAL SPME Arrow extraction and GC/MS detection offers a sensitive, precise, and high-throughput method for C2–C12 aldehyde analysis in water. It represents a significant advancement over conventional SPME techniques.
GC/MSD, SPME, GC/IT
IndustriesEnvironmental, Food & Agriculture
ManufacturerCTC Analytics
Summary
Importance of the Topic
Aldehydes are common environmental and biological compounds that influence water safety, oxidative stress markers, and food flavor quality. Monitoring C2–C12 aldehydes in drinking water is crucial for assessing disinfection by-products, ensuring compliance with health guidelines, and detecting off-flavor compounds in beverages packaged in PET.
Objectives and Study Overview
The study aimed to develop and validate a sensitive, solvent-free method for simultaneous quantification of C2–C12 aldehydes in aqueous samples. On-fiber derivatization using O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine (PFBHA) was combined with PAL SPME Arrow extraction and GC/MS detection to achieve low detection limits and robust performance.
Methodology and Instrumentation
Derivatization and extraction were conducted in a 20 mL headspace vial at 60 °C for 30 min. Key steps included:
- Reagent preparation: 100 mg PFBHA in 0.05 M H₂SO₄, purified at 90 °C.
- Internal standard: 2,3,5,6-tetrafluorobenzaldehyde in methanol.
- SPME device: PAL SPME Arrow (PDMS phase, 1.15 mm diameter, 15.3 µL sorptive volume).
- GC/MS setup: Varian 3400 GC with Saturn Ion Trap, BGB-5 column (30 m × 0.25 mm, 0.25 µm), hydrogen carrier gas, 70 °C to 280 °C ramp.
Key Results and Discussion
The method achieved:
- Limits of quantitation around 50 ng/L for most aldehydes, outperforming standard SPME fibers.
- Precision with RSD values below 9% (except C11 affected by co-elution).
- Excellent linearity up to 20 µg/L (R² > 0.99 for all compounds).
- Stable fiber performance over hundreds of extractions without degradation.
- Elevated background levels of acetaldehyde (C2) in blanks, limiting quantitation below 50 ng/L.
Benefits and Practical Applications
This approach provides rapid, solvent-free sample preparation and robust quantitation suitable for routine monitoring of drinking water quality, beverage flavor profiling, and oxidative stress studies. The increased sorptive volume of the SPME Arrow enhances sensitivity and throughput in environmental and food analysis laboratories.
Future Trends and Potential Applications
Future developments may include the use of selected ion monitoring (SIM) to lower detection limits further, adaptation to broader panels of carbonyl compounds, integration into automated workflows, and real-time monitoring in industrial and clinical settings.
Conclusion
The combination of on-fiber PFBHA derivatization with PAL SPME Arrow extraction and GC/MS detection offers a sensitive, precise, and high-throughput method for C2–C12 aldehyde analysis in water. It represents a significant advancement over conventional SPME techniques.
Reference
- P. Martos, J. Pawliszyn; Anal. Chem. 1998, 70, 2311–2320.
- F. Ventura, B. Cancho, M. T. Galceran; J. Chromatogr. A. 2002, 943, 1–13.
- S. W. Tsai, C. M. Chang; J. Chromatogr. A. 2003, 1015, 143.
- J. Beranek, A. Kubatova; J. Chromatogr. A. 2008, 1209, 44–54.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
GC-MS determination of stale aldehydes in beer by SPME on-fibre derivatization
2016|Agilent Technologies|Applications
GC/MS Application Note GC-MS determination of stale aldehydes in beer by SPME on-fibre derivatization www.palsystem.com GC-MS determination of stale aldehydes in beer by SPME on-fibre derivatization Qianjun Wu1, Hualei Chen2, Zhaoxiao Yang2 1 National Beer Quality Supervision & Inspection Center,…
Key words
beer, beerstale, stalealdehydes, aldehydesctc, ctcanalytics, analyticsspme, spmecaproaldehyde, caproaldehydepfhba, pfhbafuraldehyde, furaldehydeaging, agingderivatization, derivatizationphenylacetaldehyde, phenylacetaldehydevaleraldehyde, valeraldehydefibre, fibrepal
Quantification of nitrosamines in water by automated PAL SPME-Arrow and GC/MS
2016|CTC Analytics|Applications
GC Application Note Quantification of nitrosamines in water by automated PAL SPME-Arrow and GC/MS www.palsystem.com Quantification of nitrosamines in water by automated PAL SPME-Arrow and GC/MS Peter Egli, Beat Schilling, BGB Analytik AG, Adliswil, Switzerland Guenter Boehm, CTC Analytics AG,…
Key words
nitroso, nitrosospme, spmearrow, arrowverlaid, verlaidatogram, atograminutes, inuteshrom, hromctc, ctcanalytics, analyticsounts, ountsnitrosamines, nitrosaminespal, pallots, lotsfibers, fibersconditioning
Determination of iodoform in drinking water by PAL SPME Arrow and GC/MS
2015|CTC Analytics|Applications
GC Application Note Determination of iodoform in drinking water by PAL SPME Arrow and GC/MS www.palsystem.com Determination of iodoform in drinking water by PAL SPME Arrow and GC/MS Peter Egli, Beat Schilling, BGB Analytik AG, Adliswil, Switzerland Guenter Boehm, CTC…
Key words
pal, palspme, spmearrow, arrowanalytics, analyticsctc, ctcsorption, sorptioniodoform, iodoformarea, areaswitzerland, switzerlanddrinking, drinkingsaturn, saturncoincidental, coincidentaltime, timeimprint, imprintacknowledges
Determination of Odor Compounds in Water by SPME Arrow Gas Chromatography/Mass Spectrometry
2018|Agilent Technologies|Applications
GC/MS Application Note www.palsystem.com Determination of Odor Compounds in Water by SPME Arrow Gas Chromatography/Mass Spectrometry Determination of Odor Compounds in Water by SPME Arrow Gas Chromatography/Mass Spectrometry Jianxia Lv, CTC Analytics AG, Beijing, China Abstract In this application the…
Key words
arrow, arrowspme, spmectc, ctcanalytics, analyticsodor, odorgeosmin, geosminwater, waterpal, palpatented, patentedoff, offdetermination, determinationmethod, methodswitzerland, switzerlandcoincidental, coincidentaldrinking