Specifications for PAL-xt Systems

Brochures and specifications | 2013 | CTC AnalyticsInstrumentation
GC, Sample Preparation, HPLC
Industries
Manufacturer
CTC Analytics

Summary

Significance of the topic


The PAL-xt system is a versatile automated platform for precise sample preparation and injection in analytical laboratories. Reliable performance in liquid handling and chromatographic injection is critical for quality control, method validation, and high‐throughput analyses across pharmaceutical, environmental, and industrial applications.

Objectives and overview of the document


This specification sheet presents performance characteristics of the PAL-xt system in various test modes: gravimetric volume transfer, gas chromatographic injections (split, high‐split, headspace) and HPLC loop filling techniques (standard, DLW, DLW-2). It defines repeatability, linearity and carry‐over benchmarks under controlled conditions.

Methodology and instrumentation


Gravimetric tests measured transferred water volumes using a precision balance. Chromatographic tests employed iso‐octane and alkane mixtures for GC, and caffeine or benzophenone in aqueous/methanol matrices for HPLC.
  • Gas chromatography: Agilent 6890 with split/splitless injector, FID detector, helium carrier, retention gap or HP-5 columns.
  • Headspace GC: Agilent 6890 with capillary glass microcaps sampling.
  • HPLC standard injections: Agilent 1100 UV‐VIS system, Zorbax XDB-C8 column.
  • HPLC DLW / DLW-2: Dionex Summit with integrated degasser and variable wavelength detector, Zorbax SB-C18 and Halo C18 columns.

PAL-xt parameters such as syringe volume, fill speed, injection delays and wash cycles were optimized for each mode.

Main results and discussion


Gravimetric volume transfers achieved relative standard deviations (RSDs) of 0.10–0.20% across 10–2500 µL ranges and linearity correlations ≥ 0.9999. GC injection repeatability ranged 0.30–0.80% RSD, with carry‐over below 0.08% for split mode and ≤ 0.05% for headspace. HPLC loop overfill and partial‐loop filling methods exhibited RSDs ≤ 0.20% for standard injections and ≤ 0.50% for DLW techniques, with carry‐over below 40 ppm.

Benefits and practical applications


The demonstrated precision and low carry‐over of the PAL-xt platform support routine QA/QC tasks, calibration series, and trace‐level analyses. Automated control of sample volume and timing enhances throughput and reproducibility in environmental monitoring, food safety, clinical testing, and pharmaceutical laboratories.

Future trends and possibilities


Ongoing developments may include integration of miniaturized valves, expanded solvent compatibility, and adaptive software for real‐time injection optimization. Coupling with mass spectrometry and advanced analytics could further broaden applications in metabolomics and high‐throughput screening.

Conclusion


The PAL-xt system meets stringent specifications for liquid handling and chromatographic injections, providing reliable repeatability, excellent linearity, and minimal carry‐over. Its flexibility and performance make it a valuable tool for diverse analytical workflows.

Reference


No external literature references provided in the source.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Specifications for PAL Systems COMBI | GC | LC HTC | HTS | LHX | HTX
Specifications for PAL Systems COMBI | GC | LC HTC | HTS | LHX | HTX
2013|CTC Analytics|Brochures and specifications
SYSTEM Prep and Load Platform Specifications for PAL Systems COMBI | GC | LC HTC | HTS | LHX | HTX Not Valid for PAL-xt Systems Revised March 2013 | V5 PAL is a registered trademark of CTC Analytics AG…
Key words
del, delinj, injdeviation, deviationlhx, lhxfill, fillinject, injectpullup, pulluppst, pstrepeatability, repeatabilitycombi, combispeed, speedsyringe, syringecln, clnrelative, relativeagi
Specifications PAL RTC / RSI including Series II (LC & GC)
Specifications PAL RTC / RSI including Series II (LC & GC)
2019|CTC Analytics|Brochures and specifications
Specifications PAL RTC / RSI including Series II (LC & GC) PAL is a registered trademark of CTC Analytics AG | Switzerland Imprint Specifications PAL RTC and PAL RSI (LC & GC) Rev 9 Date of print: 04.2019 CTC Analytics…
Key words
rsi, rsidelay, delayrtc, rtcpal, paldispense, dispensepullup, pullupvolume, volumeinject, injectfill, fillspecifications, specificationspenetration, penetrationflow, flowspeed, speedseries, seriesincluding
PAL3 Series 2 Autosampler
PAL3 Series 2 Autosampler
2024|Agilent Technologies|Brochures and specifications
Data Sheet PAL3 Series 2 Autosampler Overview for gas chromatography samples General specifications Dimensions 85 cm X-Axis 120 cm X-Axis 160 cm X-Axis Width 915 mm (36.6 in) 1,270 mm (50.8 in) 1768 mm (70.7 in.) Depth 795 mm (31.8…
Key words
inject, injectdelay, delayfill, fillflow, flowsplit, splitpullup, pulluppenetration, penetrationinjector, injectordimensions, dimensionsdepth, depthspeed, speedvolume, volumetemperature, temperaturemodule, moduleweight
Clarity Control Module CTC PAL
Clarity Control Module CTC PAL
2024|DataApex|Manuals
CTC PAL Clarity Control Module ENG Code/Rev.: M063/90B Date: 2024-11-13 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex® and…
Key words
pal, palctc, ctcpenetration, penetrationagitator, agitatorclarity, claritydialog, dialogincubation, incubationagitated, agitatedtime, timepre, preinjection, injectionspme, spmevial, vialfiber, fiberneedle