TRACE 1310 Auxiliary Oven: Split Fitting for Auxiliary Gas - Installation Note

Manuals | 2014 | Thermo Fisher ScientificInstrumentation
GC
Industries
Manufacturer
Thermo Fisher Scientific

Summary

Importance of the Topic


Controlling split flow in gas chromatography is crucial for precise sample introduction and reproducible results. Implementing a split fitting using available auxiliary gas channels enhances flexibility and resource utilization in analytical workflows.

Objectives and Overview of the Note


This installation note describes the procedure for integrating the TRACE 1310 Auxiliary Oven split option kit P/N 19071018. The goal is to enable one auxiliary gas channel to function as a split flow controller via a splitter and flow restrictor.

Instrumentation Used


The setup involves:
  • TRACE 1310 Gas Chromatograph with Auxiliary Oven module
  • Auxiliary gas module providing three pressure/flow channels
  • Splitter fitting and 1 m, 0.5 mm i.d. flow restrictor
  • Charcoal trap and associated gas inlet fittings (M5, 1/8″ cap)

Methodology


The installation steps include:
  • Power off and remove oven covers to access auxiliary and main oven interiors
  • Plug the existing gas inlet port with the supplied 1/8″ cap
  • Swap the M5 cap for the M5 gas inlet fitting to connect the splitter
  • Position the splitter, flow restrictor, and charcoal trap inside the auxiliary oven per the pneumatic schematic
  • Ensure the flow restrictor resides in the heated oven to maintain stable flow

Key Results and Discussion


The procedure allows straightforward retrofit of split flow control without modifying the main GC gas lines. Visual examples demonstrate installation views. Emphasis is placed on maintaining temperature around the flow restrictor to guarantee flow stability and reproducibility.

Benefits and Practical Applications


  • Utilizes existing auxiliary gas channels to expand injector capabilities
  • Offers flexible placement of splitter and restrictor within the oven
  • Maintains regulated temperature for consistent split ratios
  • Simplifies installation with minimal hardware changes

Future Trends and Potential Applications


As gas chromatography evolves, leveraging auxiliary modules for multi-flow control and automated switching between injection modes may enhance throughput. Customizable split arrangements and integration with advanced control software represent promising directions.

Conclusion


Integrating the split fitting kit into the TRACE 1310 Auxiliary Oven provides a cost-effective approach to adding split flow control. The clear installation procedure and emphasis on temperature regulation ensure reliable performance and broaden analytical flexibility.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
TRACE 1310 Auxiliary Oven: Methanizer Assembly User Guide
TRACE 1310 Auxiliary Oven: Methanizer Assembly User Guide This note provides instruction for installing, maintaining, and troubleshooting the methanizer assembly for the TRACE 1310 Auxiliary Oven. Note This guide updates the sections related to the Methanizer Assembly described in the…
Key words
methanizer, methanizermixer, mixerdowel, dowelprotective, protectivecatalyst, catalystgrid, gridauxiliary, auxiliaryassembly, assemblyfixing, fixingreplace, replacecarbon, carbonscrews, screwstubing, tubingmonoxide, monoxidemetal
TRACE 1300/1310 GC: In-Series Connection TCD Module Installation Note
TRACE 1300/1310 GC: In-Series Connection TCD Module Installation Note This note provides instruction for installing and connecting the TCD module (P/N 19070045) in-series with a second detector installed on the TRACE 1300/1310 GC, or to the methanizer in case of…
Key words
tcd, tcdvent, ventmetal, metalmethanizer, methanizerline, linetubing, tubingseries, seriesconnection, connectionoven, ovenprotective, protectivemodule, modulemeasuring, measuringmixer, mixerauxiliary, auxiliarycap
4000 GC/MS Hardware Operation Manual
4000 GC/MS Hardware Operation Manual
2009|Agilent Technologies|Manuals
NOTICE: Varian, Inc. was acquired by Agilent Technologies in May 2010. This document is provided as a courtesy but is no longer kept current and thus will contain historical references to Varian. For more information, go to www.agilent.com/chem. Varian, Inc.…
Key words
trap, trapmanifold, manifoldboard, boardexternal, externalion, ionvacuum, vacuumassembly, assemblyline, lineforeline, forelinesource, sourcegas, gasreagent, reagentelectronics, electronicshazard, hazardinternal
Installing the Auxiliary EPC Flow Control Manifold
Installing the Auxiliary EPC Flow Control Manifold The Auxiliary EPC Flow Control Manifold kit, can be used to replace any HP 6890 Series Auxiliary EPC flow control manifold. This kit contains: Copyright© 1998 Hewlett-Packard Company Printed in USA 9/98 HP…
Key words
manifold, manifoldbracket, bracketgang, gangribbon, ribbonboard, boardfitting, fittingcable, cableauxiliary, auxiliaryslot, slotpcb, pcblabel, labelcaution, cautionhewlett, hewlettpackard, packardtag
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike