Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Applications | 2016 | Teledyne LABSInstrumentation
GC/MSD, Purge and Trap, GC/SQ
Industries
Environmental
Manufacturer
Agilent Technologies, Teledyne LABS

Summary

Significance of the topic


Purge and Trap (P&T) is a cornerstone technique for measuring volatile organic compounds (VOCs) in water at trace levels. Its ability to concentrate analytes down to low parts per billion is vital for environmental monitoring, regulatory compliance and quality control in water analysis.

Goals and overview of the study


This study evaluates the performance of the Teledyne Tekmar Lumin Moisture Control System (MCS) under US EPA Method 524 and compares it with an earlier condensate trap design. Key objectives include minimizing water vapor transfer, reducing dry purge and desorption times, and assessing the impact on analytical throughput and method detection limits.

Methodology and instrumentation


The investigation comprised replicate analyses of deionized water samples using varying purge volumes, purge times, flow rates and trap desorption times. The following instrumentation was employed:
  • Teledyne Tekmar Lumin Purge and Trap Concentrator with Moisture Control System
  • Vocarb 3000 (K) sorbent trap
  • Agilent 7890A gas chromatograph coupled to a 5975C mass spectrometer

Samples were purged with helium, followed by thermal desorption and GC/MS analysis under controlled oven and MS scanning conditions.

Main results and discussion


The Lumin MCS removed up to 55% more water vapor compared to the previous Stratum trap without any dry purge step. Elimination of dry purge yielded at least one minute reduction in total cycle time. Desorption time optimization showed that reducing trap desorb from four to 0.5 minutes cut water transfer by over 80%, while preserving peak areas for the majority of typical EPA 524 compounds. When evaluating low, standard and high purge parameter sets, response factor variability and method detection limits remained comparable across all conditions, even as total P&T cycle time was halved under optimized settings.

Benefits and practical applications of the method


  • Significant reduction in water carryover into GC/MS, enhancing instrument uptime
  • Shortened P&T cycle times enable higher sample throughput
  • Maintained analytical performance and detection limits for regulatory VOC analysis

Future trends and potential applications


Advances may include further refinement of moisture control for hydrophilic sorbents, automated scheduling of optimized parameter sets, adaptation to diverse sample matrices such as soils or sludges, and integration with high-throughput workflows in environmental and industrial laboratories.

Conclusion


The Lumin MCS demonstrates clear advantages for EPA Method 524 sampling by substantially reducing water vapor transfer and analysis time without compromising sensitivity or precision, offering a robust solution for high-volume VOC monitoring.

Used Instrumentation


  • Teledyne Tekmar Lumin Purge and Trap Concentrator with Moisture Control System
  • Vocarb 3000 (K) trap
  • Agilent 7890A GC/5975C MS


References


1. Munch JW. US EPA Method 524.2 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry. EPA National Exposure Research Laboratory; 1995.
2. Prakash B, Zaffiro AD, Zimmerman M, Munch DJ, Pepich BV. US EPA Method 524.3 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry. EPA Technical Support Center; 2009.
3. Munch DJ, Wendelken SC. US EPA Method 524.4 Measurement of Purgeable Organic Compounds in Water by Gas Chromatography/Mass Spectrometry Using Nitrogen Purge Gas. EPA Office of Water; 2013.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
US EPA Method 524.2 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 and GC/ISQ™ MS system
US EPA Method 524.2 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 and GC/ISQ™ MS system Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 524.2 is…
Key words
temp, tempbake, bakepurge, purgeppb, ppbether, etherlumin, lumindesorb, desorbvariable, variablemcs, mcstable, tablersd, rsdbutylbenzene, butylbenzenetime, timedrain, drainwater
US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS
US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC / 5977A MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 524.2 is used primarily by…
Key words
temp, tempbake, bakepurge, purgeppb, ppblumin, luminether, etherdesorb, desorbvariable, variablemcs, mcsrsd, rsdtable, tablebutylbenzene, butylbenzenetime, timedrain, drainmdl
US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and PerkinElmer® Clarus® 600 GC/Clarus® SQ8T MS
US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and PerkinElmer® Clarus® 600 GC/Clarus® SQ8T MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 524.2 is used primarily by…
Key words
temp, temppurge, purgebake, bakeether, etherppb, ppbtert, tertdesorb, desorbvariable, variablersd, rsdtable, tablebutylbenzene, butylbenzenetime, timedrain, drainlumin, lumintrap
US EPA Method 524.2 with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1806 P+T VOC ENV US EPA Method 524.2 with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method…