GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Analysis of Nitro Compounds

Applications |  | SCION InstrumentsInstrumentation
GC, GC columns, Consumables
Industries
Manufacturer
SCION Instruments

Summary

Importance of Nitro Compound Analysis



Aromatic nitro compounds play a crucial role in various industrial processes and environmental monitoring due to their toxicity and prevalence as pollutants. Reliable analytical methods are needed to ensure workplace safety, product quality, and environmental compliance.

Study Objectives and Overview



This application note (AN0056) aims to establish a robust gas chromatography–flame ionization detection (GC-FID) protocol for the separation and identification of common aromatic nitro compounds. Using a SCION GC-FID system, the method targets efficient resolution of eight analytes under controlled temperature and flow conditions.

Methodology and Instrumentation



The sample analysis was performed with the following configuration:
  • Injector split ratio: 1:100, injection volume: 1 µL
  • Column: SCION-1701, 30 m × 0.25 mm ID, 0.25 µm film thickness
  • Oven temperature program: 45 °C (3 min hold), ramp to 280 °C at 10 °C/min
  • Carrier gas: Helium at 1 mL/min
  • Detector: Flame ionization detector (FID)

Key Results and Discussion



The optimized method successfully separated eight target compounds with clear baseline resolution. Peak identification was confirmed for benzene, toluene, chlorobenzene, ethylbenzene, m-xylene, p-xylene, o-xylene, and styrene. The temperature program provided sharp peaks and consistent retention time reproducibility.

Benefits and Practical Applications



This GC-FID approach offers:
  • High sensitivity and selectivity for aromatic nitro compounds
  • Short analysis time (~26 minutes total runtime)
  • Easy integration into quality control workflows
  • Applicability in environmental monitoring and industrial QA/QC

Future Trends and Opportunities



Emerging developments may include coupling GC with mass spectrometry for enhanced specificity, rapid temperature programming for faster throughput, and miniaturized GC systems for field-deployable analysis.

Conclusion



The described GC-FID method delivers reliable, reproducible separation of key aromatic nitro analytes. Its straightforward setup and robust performance make it a valuable tool for laboratories engaged in environmental and industrial monitoring.

References



No specific literature references were provided in the original document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Aromatics using SCION-WAXMS GC Column
Analysis of Aromatics using SCION-WAXMS GC Column Application Note AN0039 SCION GC-FID Conditions S/SL 250°C, Split 75mL/min Sample 0.1µL (0.1% cyclohexane) Column SCION-WaxMS 60m x 0.25mm x 0.25µm Oven Programme 55°C to 150°C at 4°C/min Carrier Gas Helium 2.0 bar…
Key words
scion, scionxylene, xylenefid, fidwaxms, waxmsprogramme, programmeinstruments, instrumentsaromatics, aromaticsethylbenzene, ethylbenzenenetherlands, netherlandsbenzene, benzenetoluene, toluenecolumn, columnoven, ovencarrier, carrieridentification
Headspace Analysis of Solvents in Printed Paper
Headspace Analysis of Solvents in Printed Paper Application Note AN0053 SCION GC-HS-FID GC-HS-FID conditions Injector HS Column SCION-5MS 30m x 0.32mm x 1.0µm Oven 45°C (12 min) to 100°C at 10°C/min (8 mins) Carrier Gas Helium, 2mL/min FID 250°C Peak…
Key words
scion, scionpropylacetate, propylacetatepropanol, propanolfid, fidiso, isoxylene, xyleneethylacetate, ethylacetateinstruments, instrumentsethylbenzene, ethylbenzenenetherlands, netherlandspaper, paperethanol, ethanoltoluene, tolueneheadspace, headspaceunknown
Determination of Halogenated Hydrocarbons in Water
Determination of Halogenated Hydrocarbons in Water Application Note AN0045 SCION GC-HS-MS GC-HS-MS conditions Peak Identification HS 80°C, 30min Injector 170°C, 0.6:1 Sample 14mL +7g Na2SO4 in 20mL vial Column SCION-624MS 60m x 0.32mm x 1.8µm Oven Carrier Gas 45°C (10mins)…
Key words
scion, sciontrichloronitromethane, trichloronitromethanexylene, xylenetetrachloromethane, tetrachloromethaneinstruments, instrumentstrichlorofluoromethane, trichlorofluoromethanetetrachloroethene, tetrachloroetheneisopropylbenzene, isopropylbenzenetrichloroethene, trichloroethenehalogenated, halogenateddibromochloromethane, dibromochloromethanepropylbenzene, propylbenzenebromodichloromethane, bromodichloromethanebromoform, bromoformdichloromethane
Analysis of Beer using SCION-WAXMS GC Column
Analysis of Beer using SCION-WAXMS GC Column
|SCION Instruments|Applications
Analysis of Beer using SCION-WAXMS GC Column Application Note AN0040 SCION GC-HS-FID Headspace Conditions Oven 60°C Sample Loop 60°C Transfer Line 100°C Vial Equib Time 20 minutes Fill Pressure 15psi at 50mL/min Inject Time 0.5 minutes GC-FID conditions S/SL Sample…
Key words
scion, scionisoamyl, isoamylfid, fidethyl, ethylacetate, acetatesulphide, sulphidewaxms, waxmsprogramme, programmeisobutanol, isobutanolinstruments, instrumentsoven, ovenacetaldehyde, acetaldehydebeer, beerformate, formatenetherlands
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike