Separation of Nitro and Halogenated Aromatic Compounds

Applications |  | SCION InstrumentsInstrumentation
GC, GC columns, Consumables
Industries
Manufacturer
SCION Instruments

Summary

Significance of the Topic


Accurate separation and detection of nitro and halogenated aromatic compounds are critical in environmental monitoring, quality control and industrial process surveillance due to their toxicity, persistence and regulatory importance.

Objectives and Study Overview


The aim of this application note (AN0052) is to demonstrate a reliable GC-FID method for the simultaneous separation of fourteen nitro and halogenated aromatic analytes, including dichlorobenzenes, trichlorobenzenes, dinitrotoluenes and related compounds.

Methodology and Instrumentation


Sample introduction was carried out via a split injection (1:100) of 1 µL onto a SCION-1701 capillary column (30 m × 0.25 mm × 0.25 µm). The oven was programmed from 45 °C (held for 3 min) to 280 °C at 10 °C/min. Helium was used as carrier gas at 1 mL/min. Detection employed a flame ionization detector (FID).

Main Results and Discussion


The method achieved clear baseline resolution of all fourteen target compounds in a single run. Early eluting analytes are dichlorobenzene isomers, followed by hexachloroethane and nitrobenzene. Mid-range retention includes isophorone and trichlorobenzene, while late elution covers dinitrotoluene isomers, azobenzene and hexachlorinated species. The SCION-1701 column provided sufficient selectivity to distinguish positional isomers and highly chlorinated analytes, demonstrating good peak symmetry and reproducibility.

Benefits and Practical Applications


  • Comprehensive separation of environmental pollutants and industrial analytes in a single analysis
  • High reproducibility and robust peak resolution for QA/QC workflows
  • Straightforward implementation on standard GC-FID systems

Future Trends and Applications


Adaptation of this method to tandem mass spectrometry could enhance compound identification. Faster temperature ramps and smaller diameter columns may reduce analysis time. Integration with automated sample handling will streamline high-throughput monitoring.

Conclusion


This GC-FID approach effectively separates a broad range of nitro and halogenated aromatics with excellent resolution, making it suitable for environmental, regulatory and industrial applications.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Beer using SCION-WAXMS GC Column
Analysis of Beer using SCION-WAXMS GC Column
|SCION Instruments|Applications
Analysis of Beer using SCION-WAXMS GC Column Application Note AN0040 SCION GC-HS-FID Headspace Conditions Oven 60°C Sample Loop 60°C Transfer Line 100°C Vial Equib Time 20 minutes Fill Pressure 15psi at 50mL/min Inject Time 0.5 minutes GC-FID conditions S/SL Sample…
Key words
scion, scionisoamyl, isoamylfid, fidethyl, ethylsulphide, sulphideacetate, acetatewaxms, waxmsprogramme, programmeisobutanol, isobutanoloven, oveninstruments, instrumentsacetaldehyde, acetaldehydebeer, beerformate, formatenetherlands
Analysis of low ppm levels of ethylene, propylene and methanol in water
Analysis of low ppm levels of ethylene, propylene and methanol in water AN0041 Application Note SCION GC-FID GC-FID conditions Direct Injector Sample Column Oven Programme Carrier Gas FID 100°C 0.3µL SCION-PLOT Q 25m x 0.53mm x 20µm 35°C (5mins) to…
Key words
scion, scionfid, fidprogramme, programmeinstruments, instrumentsconc, concnetherlands, netherlandsppm, ppminjector, injectordirect, directoven, ovencarrier, carrieridentification, identificationnote, notepeak, peakgas
Headspace Analysis of Solvents in Printed Paper
Headspace Analysis of Solvents in Printed Paper Application Note AN0053 SCION GC-HS-FID GC-HS-FID conditions Injector HS Column SCION-5MS 30m x 0.32mm x 1.0µm Oven 45°C (12 min) to 100°C at 10°C/min (8 mins) Carrier Gas Helium, 2mL/min FID 250°C Peak…
Key words
scion, scionpropylacetate, propylacetatepropanol, propanolfid, fidiso, isoxylene, xyleneethylacetate, ethylacetateinstruments, instrumentsethylbenzene, ethylbenzenenetherlands, netherlandspaper, paperethanol, ethanoltoluene, tolueneheadspace, headspaceunknown
CFC separation using Porous Polymer Material
CFC separation using Porous Polymer Material
|SCION Instruments|Applications
CFC separation using Porous Polymer Material Application Note AN0044 SCION GC-FID GC-FID conditions Injector 250°C, Split Sample 1µL Column SCION-BOND Q 25m x 0.53mm x 10µm Oven Programme 100°C (2mins) to 250°C at 10°C/min Carrier Gas Helium 40kPa FID 250°C…
Key words
scion, scionfid, fidcfc, cfcprogramme, programmeinstruments, instrumentsbutane, butanepentane, pentanenetherlands, netherlandsporous, poroushexane, hexanepolymer, polymerinjector, injectoroven, ovenmaterial, materialcarrier