Purge and Trap - The Gases

Technical notes |  | CDS AnalyticalInstrumentation
GC/MSD, Purge and Trap
Industries
Manufacturer
Agilent Technologies, CDS Analytical

Summary

Importance of the Topic


This study addresses the critical need to detect trace levels of volatile organic compounds in water samples, a key requirement in environmental monitoring and regulatory compliance.

Study Objectives and Overview


The application note evaluates the performance of the CDS Model 7000 Purge & Trap system coupled to an Agilent 6890 gas chromatograph with a 5975B mass spectrometer for the analysis of six light gases at sub-ppb concentrations. It focuses on chromatographic resolution, reproducibility, and quantitative linearity.

Methodology


Water samples spiked with target analytes were purged using helium at 40 ml/min for 11 minutes. A multi-bed sorbent trap captured the volatile compounds, which were rapidly desorbed via a timed trap preheat step followed by thermal transfer to the GC/MS system. No cryogenic cooling was required before chromatography.

Used Instrumentation


  • CDS Model 7000 Purge & Trap with trap preheat and bake functions
  • Agilent 6890 Gas Chromatograph coupled to 5975B Mass Spectrometer
  • Varian CP 624 CB capillary column (30 m × 0.25 mm, 1.4 µm film)
  • Helium carrier gas, 1.3 ml/min flow, split ratio 40:1
  • Oven temperature program: 40 °C (5 min hold) to 180 °C at 10 °C/min

Main Results and Discussion


Chromatograms of the six gases at 0.5 ppb demonstrated sharp, well-separated peaks with seven overlaid runs showing average reproducibility of approximately 6 % RSD. Linearity tests from 0.5 to 50 ppb yielded correlation coefficients (R²) ≥ 0.999 for all targets. Method detection limits ranged from 0.03 to 0.25 ppb, confirming high sensitivity.

Benefits and Practical Applications


  • High sensitivity for sub-ppb level detection without cryogenics
  • Rapid trap preheat ensuring efficient collection and release of early eluting compounds
  • Robust reproducibility supporting routine environmental and quality control analyses

Future Trends and Applications


  • Integration with automated sampling systems for higher throughput
  • Development of advanced sorbent materials to enhance selectivity and capacity
  • Miniaturization and field-portable devices for on-site monitoring
  • Advanced data processing and coupling with tandem mass spectrometry for increased specificity

Conclusion


The CDS Model 7000 Purge & Trap combined with GC/MS delivers reliable, sensitive, and reproducible analysis of volatile gases in water at sub-ppb levels. Its excellent linearity and lack of cryogenics streamline routine laboratory operations, supporting diverse environmental and industrial applications.

References


  • CDS Analytical, Inc. Application Note #117: Purge and Trap – The Gases

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Shimadzu Guide to US EPA Methods 524.3 and 524.4 for Analysis of Volatile Organic Compounds in Drinking Water
Gas Chromatograph Mass Spectrometer Shimadzu Guide to US EPA Methods 524.3 and 524.4 for Analysis of Volatile Organic Compounds in Drinking Water No. GCMS-1502 Anne Jurek, EST Analytical ■ Abstract Due to advances in analytical instrumentation the United States Environmental…
Key words
pass, passether, etherppb, ppbpir, pirpurge, purgebutylbenzene, butylbenzenemethyl, methylchloride, chlorideethyl, ethylamyl, amylbutyl, butyltrap, traprun, runencon, enconcenturion
US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and PerkinElmer® Clarus® 600 GC/Clarus® SQ8T MS
US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and PerkinElmer® Clarus® 600 GC/Clarus® SQ8T MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 524.2 is used primarily by…
Key words
temp, temppurge, purgebake, bakeether, etherppb, ppbtert, tertdesorb, desorbvariable, variabletable, tablersd, rsdbutylbenzene, butylbenzenetime, timelumin, lumindrain, draintrap
Routine analysis of volatile organic compounds in drinking water with ISQ 7000 GC-MS
APPLICATION NOTE 65632 Routine analysis of volatile organic compounds in drinking water with ISQ 7000 GC-MS Authors: Adam Ladak1, David Lee2, and Amy Nutter3 1 Thermo Fisher Scientific, Groton, MA, USA 2 Thermo Fisher Scientific, Runcorn, UK 3 Teledyne Tekmar,…
Key words
temp, tempbake, bakepurge, purgerinse, rinseprecision, precisionsurr, surrmethanol, methanolwater, waterdesorb, desorbsweep, sweeproutine, routinechloride, chloridebutylbenzene, butylbenzeneether, ethermethacrylate
US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1803 P+T VOC ENV US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA…
Key words
temp, temppurge, purgebake, bakeacetate, acetatedesorb, desorbrinse, rinseppb, ppbtert, tertsweep, sweepether, ethertime, timecondensate, condensateaverage, averagemethyl, methylbutyl
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike