Carpet Volatile Emissions
Applications | | CDS AnalyticalInstrumentation
Carpet and other interior materials can emit VOCs that degrade air quality and pose health hazards. Monitoring these emissions supports compliance with indoor air standards and informs product design.
This application note describes a method for quantifying VOCs emitted from new carpet samples using dynamic headspace sampling and thermal desorption GC/MS. The target analytes include styrene, 4-phenylcyclohexene (4-PCH) and dodecanol.
A 7.5 cm × 7.5 cm carpet sample was placed in an 800 mL sealed dynamic headspace vessel. Helium was used as purge gas at 50 mL/min for one hour. Volatiles were captured on a Tenax sorbent tube via an external vent port. The tube was analyzed with a Dynatherm TDA autosampler interfaced to a Dynatherm 9300 thermal desorption unit and GC/MS. GC conditions included an HP-5MS column (30 m × 0.25 mm × 0.25 µm), helium carrier (50:1 split), injector at 300 °C, and a temperature program from 40 °C to 210 °C. Desorption parameters comprised preheat, trap focusing and transfer line heating.
The results confirm reliable quantification of carpet VOCs via dynamic headspace thermal desorption GC/MS.
Advancements may include automated continuous monitoring, coupling with real-time mass spectrometry, and expansion to a broader range of indoor pollutants. Integration of data analytics could enhance risk assessment and material certification.
The presented method offers a robust tool for assessing carpet emissions, combining dynamic headspace sampling with thermal desorption GC/MS for accurate, reproducible analysis of key VOCs. This supports improved indoor air quality management and product development.
GC/MSD, HeadSpace
IndustriesMaterials Testing
ManufacturerCDS Analytical
Summary
Importance of the Topic
Carpet and other interior materials can emit VOCs that degrade air quality and pose health hazards. Monitoring these emissions supports compliance with indoor air standards and informs product design.
Objectives and Overview of the Study
This application note describes a method for quantifying VOCs emitted from new carpet samples using dynamic headspace sampling and thermal desorption GC/MS. The target analytes include styrene, 4-phenylcyclohexene (4-PCH) and dodecanol.
Methodology and Instrumentation
A 7.5 cm × 7.5 cm carpet sample was placed in an 800 mL sealed dynamic headspace vessel. Helium was used as purge gas at 50 mL/min for one hour. Volatiles were captured on a Tenax sorbent tube via an external vent port. The tube was analyzed with a Dynatherm TDA autosampler interfaced to a Dynatherm 9300 thermal desorption unit and GC/MS. GC conditions included an HP-5MS column (30 m × 0.25 mm × 0.25 µm), helium carrier (50:1 split), injector at 300 °C, and a temperature program from 40 °C to 210 °C. Desorption parameters comprised preheat, trap focusing and transfer line heating.
Main Results and Discussion
- Chromatograms after one-hour purge showed distinct peaks for styrene, 4-PCH and dodecanol.
- Calibration curves exhibited high linearity with R2 values above 0.99 for all analytes over 0–1200 ng.
- Reproducibility tests of 200 ng 4-PCH spikes yielded a relative standard deviation of 1.87% over 15 runs.
The results confirm reliable quantification of carpet VOCs via dynamic headspace thermal desorption GC/MS.
Benefits and Practical Applications
- Non-invasive sampling of full materials under realistic conditions.
- High sensitivity and reproducibility for regulatory compliance.
- Applicability to various building products to assess indoor air quality.
Future Trends and Potential Applications
Advancements may include automated continuous monitoring, coupling with real-time mass spectrometry, and expansion to a broader range of indoor pollutants. Integration of data analytics could enhance risk assessment and material certification.
Conclusion
The presented method offers a robust tool for assessing carpet emissions, combining dynamic headspace sampling with thermal desorption GC/MS for accurate, reproducible analysis of key VOCs. This supports improved indoor air quality management and product development.
Instrumental Setup
- Dynamic headspace vessel (800 mL) with Helium purge at 50 mL/min.
- Tenax sorbent sampling tubes.
- Dynatherm TDA autosampler and Dynatherm 9300 thermal desorption unit.
- GC/MS with HP-5MS column (30 m × 0.25 mm × 0.25 µm) and helium carrier gas.
- Injector and transfer lines heated to 300 °C and 260 °C, respectively.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Tile and Molding Volatiles
|CDS Analytical|Applications
Tile and Molding Volatiles Application Note Environment Many of the manufactured floor and base molding products that are produced today come with an adhesive backing. This would include items such as resilient/ rubber backed flooring, linoleum tile, as well as…
Key words
tile, tilemolding, moldingvolatiles, volatilesfloor, floortrapping, trappinginvolves, involvesdesorption, desorptionadhesive, adhesivetenax, tenaxthermal, thermalabundance, abundancedestransfer, destransfermoldhat, moldhatorbedtube, orbedtubeventtechnology
VDA 278 Analysis Using Thermal Desorption
|CDS Analytical|Applications
#97a VDA 278 Analysis Using Thermal Desorption Application Note Automotive Author: C. Zawodny The German Automobile Association method VDA 278 is a thermal desorption analysis for organic emissions from interior components of a vehicle. This analysis deals with a VOC…
Key words
voc, vocfog, fogtra, tration, tionabundance, abundancetub, tubcarpetfibers, carpetfibersfogstd, fogstdiniti, initivalv, valvfibers, fibersfigure, figuretda, tdalong, longcarpet
Detection of Accelerants in Arson Residues using Thermal Desorption-GC/MS
|CDS Analytical|Applications
#99a Detection of Accelerants in Arson Residues using Thermal Desorption-GC/MS Application Note Forensics Author: C. Zawodny When arson is suspected with a fire of unknown origin, the remaining residues are examined for traces of flammable accelerants. Typical accelerants include the…
Key words
desorption, desorptionthermal, thermalaccelerant, accelerantaccelerants, accelerantsarson, arsondebris, debrisgasoline, gasolineconsisting, consistingsoftwood, softwoodprofiles, profilesextinguished, extinguishedleft, leftdynatherm, dynathermcarpet, carpetseasoned
Optimising Analytical Performance and Extending the Application Range of Thermal Desorption for Monitoring Air Indoors and Inside Vehicle Cabins
2013|Markes|Applications
Thermal Desorption Technical Support Note 28: Optimising Analytical Performance and Extending the Application Range of Thermal Desorption for Monitoring Air Indoors and Inside Vehicle Cabins Key Words: Pollution, ventilation, tracer gases, exposure, humidity, artifacts, sorbents, Fick’s Law, real-time technology Introduction…
Key words
desorption, desorptionthermal, thermalsorbent, sorbentsampling, samplingtube, tubeair, airsorbents, sorbentstubes, tubesindoor, indoorvolatile, volatilediffusive, diffusivesmoker, smokerbreath, breathstrength, strengthemissions