GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Using Analyte Protectants and Solvent Selection to Maximize the Stability of Organophosphorus Pesticides during GC/MS Analysis

Applications | 2013 | Agilent TechnologiesInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Industries
Environmental, Food & Agriculture
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Gas chromatography–mass spectrometry (GC/MS) of organophosphorus pesticides faces challenges due to analyte instability in common solvents and active-site induced losses during injection. Reliable quantification demands optimized solvent and protectant selection to maintain analyte integrity and accurate measurements.

Objectives and Study Overview


This study evaluates multiple solvents and analyte protectants to determine conditions that maximize stability and recoveries of organophosphorus pesticide residues in GC/MS analysis. Solvent performance was compared against blank extracted matrix, and the effectiveness of various protectants was assessed both in immediate recovery and in long-term storage stability tests.

Methodology and Instrumentation


The experimental design involved preparing 10 ppb standard solutions of 24 organophosphorus pesticides in solvents including hexane, methylene chloride, ethyl acetate, nonane and MTBE. Analyte protectants tested at 1000 ppm were 3-ethoxy-1,2-propanol, L-gulonic acid γ-lactone, olive oil, and d-sorbitol. Stability was monitored over four weeks. GC/MS analysis employed an Agilent 7890A GC with multimode inlet and Agilent 7000B triple quadrupole in EI mode, using two 15 m × 0.25 mm, 0.25 µm HP-5ms columns in backflush configuration, helium carrier gas, MMI injection of 1 µL, and multiple reaction monitoring.

Key Results and Discussion


Hexane consistently delivered the highest recoveries for all pesticides, outperforming methylene chloride and ethyl acetate. Solvents with lower polarity indices yielded improved peak intensities and shapes. Among protectants, d-sorbitol achieved the most significant recovery enhancement for over half of the analytes and substantially improved stability during four-week storage. 3-ethoxy-1,2-propanol also provided marked improvements for early and late eluting compounds, whereas olive oil had limited benefit.

Benefits and Practical Applications


Optimizing solvent and protectant selection enhances the accuracy and reproducibility of pesticide quantitation by reducing analyte losses at active sites and minimizing matrix effects. This approach streamlines method development, eliminates reliance on complex matrices for calibration, and ensures long-term standard stability in routine QA/QC and research laboratories.

Future Trends and Potential Applications


Emerging strategies may involve novel protectant chemistries tailored to specific analyte classes, advanced column surface treatments to further reduce active sites, and integration with automated sample preparation workflows. Expanding these concepts to other labile compounds and miniaturized GC/MS platforms could broaden application scope in environmental and food safety testing.

Conclusion


Selecting hexane as the solvent and d-sorbitol as an analyte protectant significantly improves the stability and recovery of organophosphorus pesticides during GC/MS analysis. These optimized conditions facilitate accurate, reproducible quantitation and robust long-term standard storage.

References


  1. M. Anastassiades, K. Mastovská, S. J. Lehotay. Evaluation of analyte protectants to improve gas chromatographic analysis of pesticides. J. Chromatogr. A. 2003;1000:163–184.
  2. V. Kocourek, J. Hajslova, K. Holadova, J. Poustka. Stability of pesticides in plant extracts used as calibrants in the gas chromatographic analysis of residues. J. Chromatogr. A. 1998;800:297–304.
  3. S.L. Reynolds et al. Intercomparison Study of Two Multi-Residue Methods for the Enforcement of EU MRLs for Pesticides in Fruit, Vegetables and Grain (SMT4-CT95-2030), Phase I, Progress Report. EUR 17870 EN, European Commission; 1997.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Fruit and Vegetable Pesticides by GC/MS/MS Using Agilent Inert Flow Path
Analysis of Fruit and Vegetable Pesticides by GC/MS/MS Using Agilent Inert Flow Path Application Note Food Testing and Agriculture Author Abstract Limian Zhao and Mike Szelewski The Agilent Inert Flow Path (IFP) solution was evaluated and validated thoroughly Agilent Technologies,…
Key words
ifp, ifpreponses, reponsesagilent, agilentcounts, countsstrawberry, strawberrydeactivated, deactivatedpesticide, pesticidesandwiched, sandwichedpesticides, pesticideshos, hosinert, inertmatrix, matrixonion, onionrelative, relativeinjection
Multiresidue Pesticide Analysis with the Agilent Intuvo 9000 GC and Agilent 7000 Series Mass Spectrometer
Multiresidue Pesticide Analysis with the Agilent Intuvo 9000 GC and Agilent 7000 Series Mass Spectrometer Application Note Author Abstract Rebecca Veeneman, PhD and This Application Note shows an evaluation of pesticides in seven different Joan Stevens, PhD matrices for the…
Key words
ifo, ifointuvo, intuvolta, ltaflua, fluaxali, xalilin, lingenogrinder, genogrinderflu, flunid, nidacr, acrlon, lonrecovery, recoverytea, tealdr, ldrwere
Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis
Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis Application Note Author Abstract Jessica Westland This application note demonstrates the use of the Agilent 7693A ALS system’s Agilent Technologies, Inc. Reversed 3-Layer Switch sandwich injection functionality for the analysis of…
Key words
matrix, matrixlayer, layerorp, orpsandwich, sandwichbottom, bottomdrawn, drawnmatched, matchednfl, nflpesticides, pesticidesmire, mireyri, yritop, topiph, iphpim, pimthre
Improved GC/MS Analysis of Tomato Pesticides with Agilent Deactivated Silica Tubing
Improved GC/MS Analysis of Tomato Pesticides with Agilent Deactivated Silica Tubing Agilent Ultimate Plus Deactivated Fused Silica Tubing for Guard Columns Application Note Food Safety Authors Abstract Yun Zou and Andy Zhai Agilent Ultimate Plus deactivated fused silica tubing was…
Key words
deactivated, deactivatedtubing, tubingultimate, ultimateguard, guardplus, plusagilent, agilentcolumn, columnfused, fusedsilica, silicasupplier, supplierpesticides, pesticidestomato, tomatomethyl, methylparathion, parathionddv
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike