Determination of Bath Salts (Pyrovalerone Analogs) in Biological Samples
Applications | 2012 | Agilent TechnologiesInstrumentation
Bath salts containing pyrovalerone analogs have emerged as potent designer stimulants with significant abuse potential. Reliable detection in biological specimens is crucial for forensic casework, toxicology screening, and public health monitoring.
This work presents a targeted GC–MS/MS protocol on an Agilent 220 Quadrupole Ion Trap platform for quantitative and qualitative analysis of pyrovalerone analogs (PVP, MDPV, Pyrovalerone, Naphyrone) in serum, whole blood, urine, vitreous fluid, and tissue homogenates over a working range of 50–1 000 ng/mL.
Sample Preparation
Linearity for all analogs was excellent (R² ≥ 0.96) across 50–1 000 ng/mL. Limits of detection and quantitation were 20 ng/mL and 50 ng/mL, respectively. No significant interferences observed. Retention times were reproducible within ±2%. Identification relied on characteristic m/z transitions (precursor 126 → quantifier 84) and ion ratio confirmation within 20% of calibration values.
This GC–MS/MS approach offers high sensitivity and selectivity, minimizing matrix effects and reducing false positives/negatives. The method is well suited for forensic toxicology laboratories requiring robust screening and confirmation of emerging psychoactive substances.
Prospective enhancements include expanding the analyte panel to novel analogs, coupling with high-resolution mass spectrometry for untargeted screening, automating sample preparation workflows, and miniaturizing extraction protocols to reduce solvent consumption.
The described GC Quadrupole Ion Trap MS/MS method provides a fast, reliable, and confident solution for the determination of pyrovalerone analogs in diverse biological matrices, supporting forensic and clinical toxicology needs.
GC/MSD, GC/IT
IndustriesForensics
ManufacturerAgilent Technologies
Summary
Importance of Topic
Bath salts containing pyrovalerone analogs have emerged as potent designer stimulants with significant abuse potential. Reliable detection in biological specimens is crucial for forensic casework, toxicology screening, and public health monitoring.
Objectives and Study Overview
This work presents a targeted GC–MS/MS protocol on an Agilent 220 Quadrupole Ion Trap platform for quantitative and qualitative analysis of pyrovalerone analogs (PVP, MDPV, Pyrovalerone, Naphyrone) in serum, whole blood, urine, vitreous fluid, and tissue homogenates over a working range of 50–1 000 ng/mL.
Methodology and Instrumentation
Sample Preparation
- Alkalinize 3 mL of specimen with carbonate/bicarbonate buffer (pH 9.8) and ammonia.
- Perform liquid–liquid extraction using n-butyl chloride.
- Dry organic layer under nitrogen, reconstitute in ethyl acetate with ropivacaine as internal standard.
- Column: DB-5MS equivalent, 25 m × 0.20 mm, 0.33 µm.
- Injection: 0.5 µL splitless, inlet 250 °C, He at 1.3 mL/min.
- Oven: 70 °C (1 min), ramp 25 °C/min to 310 °C (4.4 min).
- MS: EI-MS/MS 50–200 Da, trap 210 °C, transfer line 310 °C, solvent delay 7 min.
Main Results and Discussion
Linearity for all analogs was excellent (R² ≥ 0.96) across 50–1 000 ng/mL. Limits of detection and quantitation were 20 ng/mL and 50 ng/mL, respectively. No significant interferences observed. Retention times were reproducible within ±2%. Identification relied on characteristic m/z transitions (precursor 126 → quantifier 84) and ion ratio confirmation within 20% of calibration values.
Benefits and Practical Applications
This GC–MS/MS approach offers high sensitivity and selectivity, minimizing matrix effects and reducing false positives/negatives. The method is well suited for forensic toxicology laboratories requiring robust screening and confirmation of emerging psychoactive substances.
Future Trends and Uses
Prospective enhancements include expanding the analyte panel to novel analogs, coupling with high-resolution mass spectrometry for untargeted screening, automating sample preparation workflows, and miniaturizing extraction protocols to reduce solvent consumption.
Conclusion
The described GC Quadrupole Ion Trap MS/MS method provides a fast, reliable, and confident solution for the determination of pyrovalerone analogs in diverse biological matrices, supporting forensic and clinical toxicology needs.
References
- Baselt RC, Cravey RH. Disposition of Toxic Drugs and Chemicals in MAN. 7th ed. Biomedical Publications; 2004.
- Moffat AC. Clarke’s Isolation and Identification of Drugs. 3rd ed. Vol 2. The Pharmaceutical Press; 2004.
- Yohannan JC, Bonsenzo JS. The Characterization of 3,4-Methylenedioxypyrovalerone (MDPV). Microgram Journal. 2010;7(1).
- Brandt SD, Sumnall HR, Measham F, Cole J. Analyses of Second-Generation “Legal Highs” in the UK: Initial Findings. Drug Testing and Analysis. 2010;2:377–382.
- Saint Louis University Forensic Toxicology Laboratory. Standard Operation Procedures: Blood Drug Screen by GCNPD.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determination of Cannabinoids (THC) in Biological Samples
2012|Agilent Technologies|Applications
Determination of Cannabinoids (THC) in Biological Samples Application Note Forensic Toxicology Authors Abstract Joe Crifasi A method has been developed on the Agilent 220 Quadrupole Ion Trap using Saint Louis University EI-MS/MS for the identification and quantification of Delta-9-THC, 11-Hydroxy-THC,…
Key words
thc, thcrelative, relativeacquisition, acquisitioncooh, coohmin, mincounts, countsabundance, abundancetime, timeforensic, forensiccontrols, controlslouis, louissaint, saintcalibration, calibrationtoxicology, toxicologydrug
Ion Trap GC-MS/MS Analysis of “Bath Salts” in Biological Samples
2011|Agilent Technologies|Posters
Ion Trap GC-MS/MS Analysis of “Bath Salts” in Biological Samples Joseph Crifasi1, Christopher Long1, Ronald Honnold*2, Anthony Macherone2 1St. Louis University Forensic Toxicology Lab, St Louis, MO, USA, 2Agilent Technologies, Santa Clara, CA, USA Introduction Experimental Results Results and and…
Key words
mdpv, mdpvbrai, braiood, oodnaphyrone, naphyronever, verion, iondev, devprecursor, precursorbath, bathintra, intrastd, stdinter, intersalts, saltsamd, amdlouis
Determination of Pentobarbital in Biological Samples
2012|Agilent Technologies|Applications
Determination of Pentobarbital in Biological Samples Application Note Forensic Toxicology Authors Abstract Joe Crifasi An analytical method has been developed on the Agilent 220 Quadrupole Ion Trap Saint Louis University using CI-MS for the identification and quantification of Pentobarbital in…
Key words
pentobarbital, pentobarbitalcontrols, controlslouis, louissaint, sainttoxicology, toxicologyforensic, forensicacquisition, acquisitionnegatives, negativescalibration, calibrationrelative, relativeresµlts, resµltstrap, trapmin, mincrifasi, crifasiusa
Analysis of Designer Drugs Bath Salts by Quadrupole Ion Trap GC/MS
2012|Agilent Technologies|Applications
Analysis of Designer Drugs Bath Salts by Quadrupole Ion Trap GC/MS Application Note Forensics Authors Abstract Ron Honnold, Ph.D. and Rafael Acosta With more sophisticated production methods and ease of obtaining designer drugs Agilent Technologies, Inc. of abuse, forensic laboratories…
Key words
designer, designerbath, bathdrugs, drugsbenzodioxolylbutanamine, benzodioxolylbutanaminepryovalerone, pryovaleronemethylbenzodioxolylbutanamine, methylbenzodioxolylbutanaminemethylenedioxypyrovalerone, methylenedioxypyrovaleronemethylenedioxymethamphetamine, methylenedioxymethamphetaminebutylone, butyloneethylone, ethylonemethcathinone, methcathinonetrap, trapsalts, saltssold, soldmdma