Analysis of Sulfonic acid ester
Applications | | GL SciencesInstrumentation
Sulfonic acid esters, such as methyl methanesulfonate and its higher homologs, are potent alkylating agents and may occur as trace impurities in chemical and pharmaceutical products. Their reliable detection at low levels is critical for product safety, regulatory compliance and risk assessment.
The seven sulfonic acid esters were baseline-separated within a 7.5–21.5 min window. Peak shapes were sharp and symmetrical, indicating minimal analyte adsorption on the inert column surface. SIM detection provided high selectivity, with each ester yielding a dominant ion signal at its target m/z, reducing background noise and enhancing signal-to-noise ratios.
The presented GC-MS method using an InertCap 5MS/NP column delivers reliable separation and sensitive detection of seven sulfonic acid esters. Its simplicity, robustness and high selectivity make it well suited for routine monitoring of potential genotoxic impurities in pharmaceutical and industrial applications.
GC/MSD, GC columns, Consumables
IndustriesManufacturerGL Sciences
Summary
Importance of the Topic
Sulfonic acid esters, such as methyl methanesulfonate and its higher homologs, are potent alkylating agents and may occur as trace impurities in chemical and pharmaceutical products. Their reliable detection at low levels is critical for product safety, regulatory compliance and risk assessment.
Objectives and Study Overview
- Develop a robust GC-MS method for simultaneous analysis of seven sulfonic acid esters.
- Assess chromatographic performance of the InertCap 5MS/NP column under standard temperature programming.
- Demonstrate selective detection using Selected Ion Monitoring (SIM) for improved sensitivity.
Used Instrumentation
- Gas chromatograph coupled to a mass spectrometer operating in SIM mode.
- InertCap 5MS/NP capillary column (0.25 mm × 30 m, 0.25 µm film thickness).
- Helium carrier gas at a linear velocity of 36.1 cm/s.
- Splitless injector at 250 °C with a 1 min hold; MS interface at 280 °C.
Methodology
- Column temperature program: hold at 40 °C for 5 min, ramp at 10 °C/min to 280 °C.
- Injection volume: 1 µL of 10 ng/mL standard solution in ethyl acetate.
- Detection by MS-SIM monitoring characteristic fragment ions (m/z 80, 95, 109, 97, 141, 172, 186, 155, 200) assigned to each ester.
Main Results and Discussion
The seven sulfonic acid esters were baseline-separated within a 7.5–21.5 min window. Peak shapes were sharp and symmetrical, indicating minimal analyte adsorption on the inert column surface. SIM detection provided high selectivity, with each ester yielding a dominant ion signal at its target m/z, reducing background noise and enhancing signal-to-noise ratios.
Benefits and Practical Applications
- High sensitivity enables quantification at trace levels (down to low ng/mL).
- Minimal sample preparation with direct injection of dilute solutions.
- Inert column chemistry minimizes analyte degradation and adsorption.
- Applicable to quality control of intermediates, raw materials and final products in pharmaceutical and industrial laboratories.
Future Trends and Opportunities
- Adoption of ultra-fast GC temperature programs and shorter columns for increased throughput.
- Integration with high-resolution MS for confirmatory analysis and non-target screening.
- Automation of sample preparation and data processing for routine quality assurance.
- Development of novel stationary phases tailored for highly reactive analytes.
Conclusion
The presented GC-MS method using an InertCap 5MS/NP column delivers reliable separation and sensitive detection of seven sulfonic acid esters. Its simplicity, robustness and high selectivity make it well suited for routine monitoring of potential genotoxic impurities in pharmaceutical and industrial applications.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
4-t-Octylphenol
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications 4-t-Octylphenol Data No. GA 255-0644 4.0 (x10,000) TIC 3.5 1 3.0 2.5 I.S. 2.0 1.5 1.0 0.5 0.0 16.0 17.0 18.0 19.0 20.0 21.0 22.0 Time (min) Conditions System Column GA234 drinking water GC…
Key words
フェノール, フェノールクロロフェノール, クロロフェノールアセナフテン, アセナフテンchlorophenol, chlorophenolwater, waterinertsearch, inertsearchinertcap, inertcapphenol, phenold10, d10シリル, シリルディー, ディーdrinking, drinkingsupply, supplybstfa, bstfaガスクロマトグラフィー
Organochlorine Pesticide
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Organochlorine Pesticide Data No. GA130-0644 10.0 (x100,000) 9.5 9.0 8.5 7+8 8.0 7.5 7.0 6.5 6 6.0 10 5.5 5.0 4 4.5 4.0 12 11 5 3 9 3.5 2 3.0 1 2.5 2.0…
Key words
hch, hchbhc, bhcアルドリン, アルドリンヘプタクロル, ヘプタクロルベンゼンヘキサクロライド, ベンゼンヘキサクロライドミトタン, ミトタンエンドリン, エンドリンリンデン, リンデンmitotane, mitotanehold, holdhexachloride, hexachlorideガスクロマトグラフィー, ガスクロマトグラフィーイナートサーチ, イナートサーチキャピラリーカラム, キャピラリーカラムイナートキャップ
Organochlorine Pesticide
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Organochlorine Pesticide Data No. GA129-0644 8.0 (x100,000) 7.5 7.0 6.5 6.0 5.5 5.0 4.5 6 4.0 3.5 4 3 3.0 2.5 10 8 11 12 7 5 9 2 1 2.0 1.5 1.0 0.5…
Key words
hch, hchbhc, bhcアルドリン, アルドリンヘプタクロル, ヘプタクロルベンゼンヘキサクロライド, ベンゼンヘキサクロライドミトタン, ミトタンエンドリン, エンドリンリンデン, リンデンmitotane, mitotanehold, holdhexachloride, hexachlorideガスクロマトグラフィー, ガスクロマトグラフィーイナートサーチ, イナートサーチキャピラリーカラム, キャピラリーカラムイナートキャップ
Determination of Haloacetic Acids in Drinking Water According to EPA Method 552.3 using Hydrogen Carrier Gas
2020|Shimadzu|Applications
No. SSI-GC-2002 Gas Chromatography No. GC-2002 Determination of Haloacetic Acids in Drinking Water According to EPA Method 552.3 using Hydrogen Carrier Gas ■ Abstract Haloacetic acids (HAAs) are known carcinogens that may occur as disinfection byproducts in drinking water. Currently…
Key words
dalapon, dalaponratio, ratiocdbaa, cdbaabdcaa, bdcaaarea, areabcaa, bcaambaa, mbaatbaa, tbaadbaa, dbaatcaa, tcaamcaa, mcaadcaa, dcaahaas, haashaloacetic, haloaceticcarrier