A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas
Applications | 2009 | Teledyne LABSInstrumentation
The analysis of volatile organic compounds (VOCs) in environmental samples is essential for regulatory compliance and pollution monitoring. Traditionally, helium is used as the purge gas in Purge & Trap (P&T) systems, but its escalating cost and supply constraints have driven the search for alternatives. Demonstrating that nitrogen can substitute helium without compromising analytical performance offers significant cost savings and operational flexibility.
This study compared VOC responses when using helium versus nitrogen as the purge gas on water and soil matrices. A Teledyne Tekmar Atomx P&T autosampler coupled with an Agilent 7890A GC and 5975C MSD was employed to assess calibration linearity, detection limits, and reproducibility for 95 target analytes following USEPA Method 8260.
The experiment involved preparing water standards (0.5–200 ppb) and soil extracts (1.0–200 ppb) spiked with a 50 ppb working solution. Samples were purged for 11 minutes at 40 mL/min, trapped on a #9 adsorbent, thermally desorbed, and analyzed by GC/MS under identical temperature and flow conditions. Method detection limits (MDLs) were determined from seven low-level replicates, and relative standard deviations (%RSD) were calculated for each matrix and purge gas.
Both water and soil matrices yielded comparable calibration performance with nitrogen and helium. Average %RSD values were 10.1 % (N₂) vs. 8.3 % (He) for water, and 10.3 % (N₂) vs. 8.1 % (He) for soil. MDLs remained within 0.32–0.48 ppb for nitrogen and 0.33–0.53 ppb for helium. Overlay chromatograms of 50 ppb standards showed minimal retention time shifts and similar peak shapes, indicating that nitrogen does not substantially degrade sensitivity or precision for most analytes.
Advancements in electronic flow control and on-site nitrogen generation will enhance method robustness and ease of operation. Future work may explore nitrogen purge for other environmental matrices, optimize trap materials for improved VOC retention, and integrate real-time monitoring of gas purity to ensure consistent analytical performance.
This investigation confirms that nitrogen serves as a reliable and economical alternative to helium for VOC analysis using P&T GC/MS. Comparable precision, detection limits, and chromatographic quality support transfer to routine environmental laboratories, offering significant cost and logistical advantages without sacrificing data integrity.
GC/MSD, Purge and Trap, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, Teledyne LABS
Summary
Význam tématu
The analysis of volatile organic compounds (VOCs) in environmental samples is essential for regulatory compliance and pollution monitoring. Traditionally, helium is used as the purge gas in Purge & Trap (P&T) systems, but its escalating cost and supply constraints have driven the search for alternatives. Demonstrating that nitrogen can substitute helium without compromising analytical performance offers significant cost savings and operational flexibility.
Cíle a přehled studie
This study compared VOC responses when using helium versus nitrogen as the purge gas on water and soil matrices. A Teledyne Tekmar Atomx P&T autosampler coupled with an Agilent 7890A GC and 5975C MSD was employed to assess calibration linearity, detection limits, and reproducibility for 95 target analytes following USEPA Method 8260.
Použitá metodika
The experiment involved preparing water standards (0.5–200 ppb) and soil extracts (1.0–200 ppb) spiked with a 50 ppb working solution. Samples were purged for 11 minutes at 40 mL/min, trapped on a #9 adsorbent, thermally desorbed, and analyzed by GC/MS under identical temperature and flow conditions. Method detection limits (MDLs) were determined from seven low-level replicates, and relative standard deviations (%RSD) were calculated for each matrix and purge gas.
Použitá instrumentace
- Teledyne Tekmar Atomx Purge & Trap Concentrator/Vial Autosampler
- Agilent 7890A Gas Chromatograph with J&W Scientific DB-VRX column (30 m × 0.25 mm, 1.4 µm)
- Agilent 5975C inert XL Mass Selective Detector (scan range m/z 35–300)
- Electronic mass flow controller configurable for helium or nitrogen
Hlavní výsledky a diskuse
Both water and soil matrices yielded comparable calibration performance with nitrogen and helium. Average %RSD values were 10.1 % (N₂) vs. 8.3 % (He) for water, and 10.3 % (N₂) vs. 8.1 % (He) for soil. MDLs remained within 0.32–0.48 ppb for nitrogen and 0.33–0.53 ppb for helium. Overlay chromatograms of 50 ppb standards showed minimal retention time shifts and similar peak shapes, indicating that nitrogen does not substantially degrade sensitivity or precision for most analytes.
Přínosy a praktické využití metody
- Substitution of nitrogen reduces reliance on costly helium cylinders and lowers per‐analysis gas expenses by up to threefold.
- Integrating high-purity nitrogen generators can eliminate gas supply logistics and further improve laboratory safety and efficiency.
- Maintains compliance with USEPA 8260 methodology while offering a cost-effective alternative.
Budoucí trendy a možnosti využití
Advancements in electronic flow control and on-site nitrogen generation will enhance method robustness and ease of operation. Future work may explore nitrogen purge for other environmental matrices, optimize trap materials for improved VOC retention, and integrate real-time monitoring of gas purity to ensure consistent analytical performance.
Závěr
This investigation confirms that nitrogen serves as a reliable and economical alternative to helium for VOC analysis using P&T GC/MS. Comparable precision, detection limits, and chromatographic quality support transfer to routine environmental laboratories, offering significant cost and logistical advantages without sacrificing data integrity.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determining Volatile Organic Compounds from Difficult Soil Matrices Utilizing both Headspace and Purge and Trap Techniques
2009|Agilent Technologies|Applications
Determining Volatile Organic Compounds from Difficult Soil Matrices Utilizing both Headspace and Purge and Trap Techniques Application Note By: Anne Jurek Abstract Purge and Trap concentration (P&T) along with Gas Chromatography/Mass Spectrometry (GC/MS) analysis is the traditionally used method for…
Key words
temp, tempvaried, variedpurge, purgebake, bakepreheat, preheatdesorb, desorbsweep, sweeptime, timerinse, rinseflow, flowsoil, soilcondensate, condensatestandby, standbyblue, bluedry
Optimization Techniques for Performance of USEPA Methods 5030, 5035, and Determinative Methods 524.2 and 8260 utilizing the Atomx Concentrator/Multimatrix Autosampler and an Agilent 7890A GC and 5975 inert XL MSD
2009|Agilent Technologies|Applications
Optimization Techniques for Performance of USEPA Methods 5030, 5035, and Determinative Methods 524.2 and 8260 utilizing the Atomx Concentrator/Multimatrix Autosampler and an Agilent 7890A GC and 5975 inert XL MSD Application Note By: Anne Jurek Abstract Environmental Analysts are frequently…
Key words
temp, tempbake, bakepurge, purgedesorb, desorbrinse, rinsevrx, vrxlens, lenssweep, sweeptime, timeflow, flowapp, appneedle, needledry, drystandby, standbycondenser
Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler.
2009|Agilent Technologies|Applications
Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract This application note discusses the processes used in USEPA Method 5035 for solid waste…
Key words
temp, tempbake, bakerinse, rinsepurge, purgesweep, sweepmethanol, methanoldesorb, desorbcondensate, condensatetime, timeflow, flowprepurge, prepurgeneedle, needlesurr, surratomx, atomxsparge
Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler
2009|Agilent Technologies|Applications
Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler Application Note By: Anne Jurek Abstract In order to determine the amount of Volatile Organic Compounds (VOCs) in water and soil matrices…
Key words
temp, tempbake, bakepurge, purgerinse, rinsedesorb, desorbcondensate, condensatesweep, sweepsoil, soilflow, flowtime, timewater, waterprepurge, prepurgeneedle, needlespcc, spccsparge