Rugged GC/MS/MS Pesticide Residue Analysis Fulfilling the USDA Pesticide Data Program (PDP) Requirements

Applications | 2012 | Agilent TechnologiesInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Monitoring pesticide residues in fruits and vegetables is critical for food safety and public health. The USDA’s Pesticide Data Program (PDP) establishes rigorous quality control standards for residue analysis in highly consumed produce. Developing robust, rapid, and sensitive analytical methods that meet PDP requirements enhances regulatory compliance and consumer protection.

Objectives and Study Overview


This study describes the development, optimization, and routine implementation of a gas chromatograph coupled with tandem mass spectrometry (GC/MS/MS) method for analyzing over 70 pesticides in fruits and vegetables. The aim was to achieve a 20-minute cycle time, high throughput, and compliance with PDP validation and quality control criteria.

Methodology and Instrumentation


The method employs QuEChERS sample preparation, followed by GC/MS/MS analysis on an Agilent 7890A GC with an Agilent 7000B Triple Quadrupole MS. A Multi-Mode Inlet (MMI) in programmable temperature vaporization (PTV) solvent vent mode allows injection of 2 µL extracts. Concurrent column backflushing via a purged ultimate union removes late-eluting matrix components, protecting the MS ion source and reducing maintenance. Key instrument components include:
  • MMI PTV inlet with air cooling and solvent vent
  • Configured dual HP-5MS UI columns with electronic pneumatics control
  • Agilent 7693A automatic liquid sampler
  • Multiple reaction monitoring (MRM) with optimized collision energies

Analyte protectants (L-gulonic acid γ-lactone and D-sorbitol) and internal standards (including 13C12-p,p’-DDT and triphenyl phosphate) were added to improve precision, recoveries, and ruggedness.

Main Results and Discussion


The method delivered excellent calibration fits (correlation coefficients R ≥ 0.997), calibration integrity (%D ≤ 20%), and mean recoveries of 95–102% with RSDs of 1.5–8.3% across all pesticides and calibration levels (1× to 10× LOQ). Concurrent backflushing effectively eliminated nonvolatile residues, reducing cycle times and source contamination. Use of compound-specific internal standards significantly improved precision for analytes prone to degradation in the inlet.

Benefits and Practical Applications


  • High throughput: 20-minute run time for >70 pesticides
  • Enhanced source protection: reduced contamination and maintenance
  • Improved ruggedness and precision: analyte protectants and ISTDs
  • Regulatory compliance: fully meets USDA-PDP validation and QC criteria

Future Trends and Possibilities


Advances in compound-specific labeled standards and improved MRM databases will further streamline method development. Automation of sample preparation and integration with high-throughput MS platforms may enable expanded analyte panels and faster turnaround. Emerging backflushing technologies and inert flow path materials will enhance instrument uptime.

Conclusion


This GC/MS/MS method combining PTV solvent vent injection, concurrent backflushing, analyte protectants, and targeted internal standards provides a powerful, rugged, and efficient tool for routine pesticide residue monitoring in compliance with USDA-PDP requirements.

References


1. PDP website, www.ams.usda.gov/science/pdp
2. Wylie PL and Meng C-K. A Method for Trace Analysis of 175 Pesticides Using Agilent Triple Quadrupole GC/MS/MS. Agilent Technol. 5990-3578EN.
3. Mastovska K et al. J Chromatogr A. 1054 (2004) 335–349.
4. Szelewski MJ and Quimby B. New Tools for Rapid Pesticide Analysis. Agilent Technol. 5989-1716EN.
5. Meng C-K. Improving Productivity with Backflush. Agilent Technol. 5989-6018EN.
6. Anastassiades M et al. J AOAC Int. 86 (2003) 412–431.
7. Agilent G1472A Backflushing Kit Manual. G1472-90001, 2010.
8. Meng C-K. Pesticides and Environmental Pollutants MRM Database. Agilent Technol. 5990-9453EN.
9. Mastovska K and Wylie PL. J Chromatogr A. DOI:10.1016/j.chroma.2012.09.094.
10. Anastassiades M et al. J Chromatogr A. 1015 (2003) 163–184.
11. Mastovska K et al. Anal Chem. 77 (2005) 8129–8137.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
GC/MS/MS Pesticide Residue Analysis - A reference guide
GC/MS/MS Pesticide Residue Analysis A reference guide Table of Contents 1. Introduction4 2. Pesticide multiresidue analysis 4 3. Sample preparation using the QuEChERS approach 7 3.1 QuEChERS history and major modifications 3.2 Cleanup options in the QuEChERS method 3.3 EMR…
Key words
counts, countsmatrix, matrixpesticide, pesticideemr, emrannex, annexacquisition, acquisitiontime, timemrm, mrmprotectants, protectantsbackflushing, backflushingpesticides, pesticidespigmented, pigmentedinlet, inletcaptiva, captivamin
Lowering Detection Limits for Routine Analysis of Pesticides Residues in Foods Using the Agilent 7000C Triple Quadrupole GC/MS
Lowering Detection Limits for Routine Analysis of Pesticides Residues in Foods Using the Agilent 7000C Triple Quadrupole GC/MS Application Note Food Safety Author Abstract Melissa Churley An analytical method that is well established for the Agilent 7000 Series Triple Agilent…
Key words
squash, squashplum, plumpesticide, pesticidewinter, wintermrl, mrlloq, loqcounts, countsmrls, mrlsacquisition, acquisitionpesticides, pesticidessum, sumrelative, relativebackflushing, backflushingloqs, loqsresidue
Analysis of Multipesticide Residues in Tobacco
Analysis of Multipesticide Residues in Tobacco
2015|Agilent Technologies|Applications
Analysis of Multipesticide Residues in Tobacco The Agilent Triple Quadrupole GC/MS/MS Analyzer for Pesticides in Tobacco Application Note Food Testing & Agriculture Authors Abstract Vivian Xianyu Chen and Andy Zhai A multiple reaction monitoring (MRM) method was developed and evaluated…
Key words
tobacco, tobaccosulfone, sulfonedemeton, demetonhch, hchmethyl, methyldisulfoton, disulfotonpesticides, pesticidessulfoxide, sulfoxidersd, rsdcounts, countsrecovery, recoverymatrix, matrixpesticide, pesticiderelative, relativefenthion
Brewing Excellence: Quantitating Over 200 Pesticides in Black Tea with Steady Performance and Maximized Uptime by GC/MS/MS
Application Note Food Brewing Excellence: Quantitating Over 200 Pesticides in Black Tea with Steady Performance and Maximized Uptime by GC/MS/MS Authors Abstract Anastasia A. Andrianova and Limian Zhao Agilent Technologies, Inc. This application note presents results for the sensitive and…
Key words
quadratic, quadraticlinear, linearmethyl, methylcounts, countsresponses, responsesacquisition, acquisitioncleanup, cleanuptea, teamode, modepesticide, pesticidetime, timebromophos, bromophosethyl, ethylmidcolumn, midcolumnblack
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike