Evaluating USEPA Method 524.3 Utilizing Newly Permissible Method Modifications to Purge and Trap Techniques

Applications | 2009 | Teledyne LABSInstrumentation
GC/MSD, Purge and Trap, GC/SQ
Industries
Environmental
Manufacturer
Agilent Technologies, Teledyne LABS

Summary

Significance of the Topic


Volatile organic compounds (VOCs) in drinking water pose health and regulatory challenges. Precise, sensitive detection of trace VOCs ensures safe public water supplies and compliance with environmental standards. The revised USEPA Method 524.3 expands flexibility in purge and trap parameters, offering potential for faster analysis and higher throughput.

Study Objectives and Overview


This work evaluates the updated USEPA Method 524.3 using modified purge and trap settings on a Stratum PTC and Aquatek 70 system coupled to GC/MS. An initial demonstration of capability (IDC) was performed under recommended conditions. Subsequently, purge extraction volumes from 250 mL to 630 mL were compared to identify optimal recovery for 76 target analytes. A final IDC confirmed system performance with the selected parameters.

Methodology and Instrumentation


The analytical setup comprised a Stratum PTC purge and trap concentrator and Aquatek 70 autosampler with refrigerated bath, an Agilent 7890A GC and 5975C Inert XL MS. A Restek inlet liner and Teledyne Tekmar #9 trap were used. Separation was achieved on an Agilent J&W DB-VRX 30 m × 0.25 mm × 1.4 µm column. Purge volumes, times, flows, desorb conditions, and temperature programming were systematically varied. Calibration standards (0.5–40 ppb) were prepared in preserved water and analyzed in 40 mL headspace vials. Quadratic regression quantified analytes; MRLs, precision and accuracy were determined per method criteria.

Main Results and Discussion


Purge volume screening revealed 350 mL at 70 mL/min purge flow, 5 min purge time, 0.5 min desorb produced the best compromise of recovery across all analytes. Calibration curves for both 440 mL and 350 mL purges exhibited R² ≥ 0.997. Precision (%RSD) for 10 ppb standards was < 8% and recoveries within ± 20%. MRL experiments (n=7 replicates at low spike) passed 50–150% recovery and half-range prediction interval criteria for all compounds. Tap water tests confirmed reliable quantification of trace VOCs (e.g., chloroform, bromodichloromethane) and surrogate recoveries near 100%.

Practical Benefits and Applications


Optimized purge and trap conditions shorten cycle time by over six minutes per sample, significantly boosting laboratory productivity without compromising sensitivity or regulatory compliance. The approach supports routine monitoring of drinking water, QA/QC in treatment facilities, and environmental assessments.

Future Trends and Opportunities


Advances may include further miniaturization of trap technology, integration with automated sample handling, and real-time data analytics using machine learning for VOC pattern recognition. Expanding the method to complex matrices (groundwater, wastewater) and coupling to high-resolution mass spectrometry could enhance selectivity and lower detection limits.

Conclusion


The study demonstrates that modified purge and trap parameters under USEPA Method 524.3 maintain analytical performance while reducing analysis time. The selected 350 mL purge at 70 mL/min and brief desorb delivers consistent recoveries, meets all QC requirements, and enhances throughput for drinking water VOC analysis.

Reference


  • USEPA Method 524.3, “Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry,” Version 1.0, June 2009.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Using Nitrogen Purge Gas for US EPA Drinking Water Methods
Using Nitrogen Purge Gas for US EPA Drinking Water Methods Application Note Abstract The USEPA developed Method 524.4, “Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry (Using Nitrogen Purge Gas)”, for identifying and measuring purgeable…
Key words
pass, passpurge, purgetemp, tempbake, bakedrinking, drinkingdesorb, desorbepa, epanitrogen, nitrogenwater, watermethods, methodscondenser, condenserppb, ppbtime, timegas, gasnote
Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler
Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is equally important to prove that the changes in technology…
Key words
tert, terttemp, temppurge, purgebake, bakeether, etherbutyl, butylbutylbenzene, butylbenzeneptc, ptcethyl, ethylstratum, stratumcondenser, condenserdesorb, desorbmethyl, methyltetrahydrofuran, tetrahydrofurantime
US EPA Method 524.2 - Atomx XYZ and Agilent 7890B GC/5977A MS
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1801 P+T VOC ENV US EPA Method 524.2 - Atomx XYZ and Agilent 7890B GC/5977A MS Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 524.2 was used to determine…
Key words
temp, tempbake, bakeppb, ppbpurge, purgerinse, rinseaverage, averagebutylbenzene, butylbenzeneatomx, atomxsweep, sweepmdl, mdlxyz, xyzdesorb, desorbchloride, chloridemethacrylate, methacrylatewater
US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1803 P+T VOC ENV US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA…
Key words
temp, temppurge, purgebake, bakedesorb, desorbacetate, acetaterinse, rinsetert, tertppb, ppbsweep, sweepether, ethertime, timecondensate, condensateaverage, averagemethyl, methylbutyl
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike