Importance of the Topic


Infant formula fat composition directly influences neonatal growth, neural development and overall health. Accurate profiling of fatty acids ensures formula safety and nutritional equivalence to human milk, while monitoring trans-fat content safeguards against potential adverse effects.

Objectives and Study Overview


This work aimed to develop a rapid, cost-effective retention time locking gas chromatography–flame ionization detection method for quantitative analysis of 37 fatty acid methyl esters (FAMEs) in infant formulas. Complementary GC–MS confirmation was applied to minimize identification errors and to comply with Chinese regulation GB 5413.27-2010.

Methodology and Instrumentation


Sample Preparation
  • Approximately 500 mg of formula was extracted in toluene and derivatized with 10 % acetyl chloride in methanol at 80 °C for 2 hours.
  • After neutralization with sodium carbonate and centrifugation, the clear supernatant was used for GC analysis.

Instrumental Conditions
  • GC-FID: Agilent 7890A with 7683 autosampler; HP-88 column (100 m × 0.25 mm × 0.20 µm); nitrogen carrier at 1 mL/min; inlet at 260 °C, split 30:1; oven 140 °C (5 min) to 240 °C at 4 °C/min; FID at 280 °C.
  • GC-MS: Agilent 7890A–5975C MSD; helium carrier at 20 cm/s; same column and temperature program; scan 40–400 amu after 8.2 min solvent delay.
  • Retention time locking was set on palmitic methyl ester (C16:0) at 18.600 min to ensure reproducible retention across instruments.

Main Results and Discussion


Baseline separation was achieved for the majority of the 37 FAMEs, including cis/trans isomers, except partial overlap of C20:3n3, C22:1n9 and C20:4n6. Resolution of arachidonic methyl ester from nervonic methyl ester improved with higher split ratios, meeting regulatory requirements. GC-MS total ion chromatograms confirmed correct identification and matched GC-FID retention order. Real infant formula and milk powder samples showed varying fatty acid compositions (0.01–5 %), highlighting the need for matrix-matched calibration.

Benefits and Practical Applications


This RTL-GC/FID approach offers
  • High throughput and cost-efficiency for routine quality control.
  • Excellent resolution of polyunsaturated and trans isomers critical for nutritional labeling.
  • Compatibility with regulatory methods and easy retention time transfer between labs.

Future Trends and Opportunities


Emerging directions include
  • Automation of derivatization to increase sample throughput.
  • Integration of high-resolution mass spectrometry for enhanced specificity.
  • Application of shorter columns or temperature-ramped dual columns for faster runs.
  • Green chemistry approaches to reduce solvent use and waste.

Conclusion


The described method delivers reliable separation and quantitation of a broad range of FAMEs in infant formulas. Retention time locking ensures reproducibility across instruments and compliance with international standards, making it suitable for routine nutritional and safety testing.

Reference


1. Stubbs CD, Smith AD. The modification of mammalian membrane polyunsaturated fatty acid composition in relation to membrane fluidity and function. Biochim Biophys Acta. 1984 Jan 27;779(1):89-137.
2. Eder K. Gas chromatographic analysis of fatty acid methyl esters. J Chromatogr B Biomed Appl. 1995 Sep 15;671(1-2):113-131.
3. David F, Sandra P, Vickers AK. Column selection for the analysis of fatty acid methyl esters. Agilent Technologies publication 5989-3760EN; 2011.
4. China National Standard GB 5413.27-2010. Determination of fatty acids in foods for infants and young children; 2010.
5. Giarocco V, Quimby B, Klee M. Retention Time Locking: Concepts and Applications. Agilent Technologies publication 5966-2469EN; 2008.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Improving the Analysis of 37 Fatty Acid Methyl Esters
Improving the Analysis of 37 Fatty Acid Methyl Esters
2018|Agilent Technologies|Applications
Application Note Food Testing Improving the Analysis of 37 Fatty Acid Methyl Esters Using three types of capillary GC columns Authors Yun Zou Agilent Technologies (Shanghai) Co.Ltd, Shanghai 200131 P.R.China Hua Wu Agilent Technologies (China) Co.Ltd, Beijing 100102 P.R.China Abstract…
Key words
acid, acidcolumn, columnfatty, fattyfames, famesinlet, inletfid, fidcarrier, carrierparameter, parameterfastfame, fastfameesters, estersvalue, valueoven, ovenfame, famefatwax, fatwaxinjection
Fatty Acid Methyl Ester Analysis Using the Agilent 8850 GC System
Application Note Food and Beverage Testing Fatty Acid Methyl Ester Analysis Using the Agilent 8850 GC System Author Abstract Jie Zhang Agilent Technologies (Shanghai) Co. Ltd. This application note presents the analysis of fatty acid methyl esters (FAMEs) using an…
Key words
fatty, fattyfastfame, fastfamefames, famescyanopropyl, cyanopropylresponse, responsetrans, transcis, cisprogram, programretention, retentionacids, acidshead, headoven, ovenmin, mincolumns, columnspressure
Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph
Application Note Food Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph Author Youjuan Zhang Agilent Technologies (Shanghai) Co. Ltd., Shanghai 200131 P. R. China Abstract An Agilent 8860 GC equipped with a split/splitless inlet and flame ionization…
Key words
fatty, fattyfastfame, fastfamedha, dhafames, famesacids, acidsisomers, isomersintuvo, intuvonote, notecolumn, columnmin, minagilent, agilentplay, playtime, timewere, wereamount
Fatty Acid Methyl Esters (FAMEs) Analysis on an Agilent 8890 GC and Its Application to Real Samples
Application Note Food Testing Fatty Acid Methyl Esters (FAMEs) Analysis on an Agilent 8890 GC and Its Application to Real Samples Author Jie Zhang Agilent Technologies, Inc. Shanghai, China Abstract This Application Note applied methods GB5009.168-20161 and 5009.257-20162 to FAMEs…
Key words
fatty, fattytrans, transacids, acidsresponse, responsefames, famesrsd, rsdretention, retentionoctadecatrienoic, octadecatrienoicarea, areamin, mintime, timeanpel, anpelesters, estersmethyl, methylcis
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike