GC/μECD Analysis and Confirmation of PCBs in Fish Tissue with Agilent J&W DB-35ms and DB-XLB GC Columns

Applications | 2012 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


The presence of persistent and bioaccumulative PCBs in fish tissue poses a significant risk to human health and environmental quality. Regular monitoring of PCB levels in seafood is critical to ensure food safety and compliance with regulatory guidelines.

Study Objectives and Overview


This study aimed to develop a reliable, cost-effective analytical workflow for quantifying 19 PCB congeners in fish tissue. It combined a QuEChERS-based sample cleanup with dispersive solid phase extraction, and a dual column gas chromatography method featuring simultaneous primary and confirmatory determination using micro electron capture detection.

Methodology and Instrumentation Used


The procedure employed Agilent Bond Elut QuEChERS AOAC extraction salts for initial acetonitrile extraction and partitioning, followed by dSPE cleanup for fatty matrices. Analysis was performed on an Agilent 7890A GC with dual μECD detectors and an unpurged 2-way splitter. Heated separation was achieved using a primary J&W DB-35ms column (30 m×0.25 mm×0.25 μm) and a confirmatory J&W DB-XLB column (30 m×0.25 mm×0.50 μm) under a programmed temperature ramp, hydrogen carrier gas, and splitless injection.

Main Results and Discussion


Chromatographic separation of 19 PCBs was completed in under 12 minutes on each column. Calibration from 10 to 400 ng/mL exhibited excellent linearity (r2>0.9939). The method limit of quantitation at 10 ppb is well below regulatory MRLs. Spiked recoveries ranged from 72 to 116% at mid to high concentrations, with acceptable repeatability. Lower recoveries at trace levels were attributed to co-elution and detector sensitivity for low-chlorinated congeners.

Benefits and Practical Applications


The single-injection dual column μECD approach streamlines confirmation and identification in one run, reducing instrument time and maintenance. The QuEChERS-dSPE cleanup provides efficient removal of lipids, enabling routine food safety laboratories to perform high-throughput PCB screening without the need for mass spectrometry.

Future Trends and Possibilities


  • Integration of μECD with tandem mass spectrometry for enhanced selectivity.
  • Automation of QuEChERS workflows for higher sample throughput.
  • Extension of the method to other persistent organic pollutants in complex matrices.
  • Development of miniaturized GC-ECD platforms for field-deployable monitoring.

Conclusion


The combined QuEChERS dSPE and dual column μECD method provides a robust, sensitive, and compliant solution for PCB monitoring in fish tissue. It meets regulatory requirements and offers an efficient alternative to GC/MS for routine analysis.

References


  1. S. Foran, E. Lewandrowski, J. Flood et al. Pathology and Laboratory Medicine, 2005, 129, 74-77.
  2. IPCS Environmental Health Criteria 140, WHO, 1993.
  3. F. Cordle, R. Locke, J. Springer, Environ Health Perspect. 1982, 45, 171-182.
  4. EPA fish advisory guidelines document.
  5. AOAC Method 2007.01.
  6. ATSDR maximum residue guidelines.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
GC-μECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil
GC-µECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil Application Note Food Safety and Environmental Markets Authors Abstract Jennifer McCulley and Ken Lynam An olive oil sample obtained from a local grocery store is analyzed for 20…
Key words
bhc, bhcendosulfan, endosulfanendrin, endrinolive, oliveoil, oilchlordane, chlordaneheptachlor, heptachlorsurrogate, surrogatetetrachloro, tetrachloroµecd, µecdsulfate, sulfatemethoxychlor, methoxychlorxlb, xlbxylene, xylenequechers
Sub μg/L Level Analysis of Chlorinated Pesticide and Herbicide Analysis in Water by GC/μECD using Agilent J&W DB-35ms Ultra Inert and DB-XLB columns
Sub µg/L Level Analysis of Chlorinated Pesticide and Herbicide Analysis in Water by GC/µECD using Agilent J&W DB-35ms Ultra Inert and DB-XLB columns Application Note Environmental Authors Abstract Doris Smith and Ken Lynam Chlorinated pesticides and herbicides in water samples…
Key words
bhc, bhcendosulfan, endosulfanpermethrin, permethrinchlordane, chlordaneheptachlor, heptachlorendrin, endrinsurrogate, surrogatexlb, xlbdisk, disketradiazole, etradiazoletcmx, tcmxpesticides, pesticideschlorinated, chlorinateddcpa, dcpabutachlor
Determination of Haloacetic Acids in Water by GC/μECD Using Agilent J&W DB-35ms Ultra Inert and DB-XLB Columns
Determination of Haloacetic Acids in Water by GC/µECD Using Agilent J&W DB-35ms Ultra Inert and DB-XLB Columns Application Note Environmental, Water Quality Authors Abstract Doris Smith and Ken Lynam Haloacetic acids (HAAs) in water samples are effectively extracted and Agilent…
Key words
methyl, methylhaas, haashaloacetic, haloaceticdibromochloroacetate, dibromochloroacetatexlb, xlbbromochloroacetate, bromochloroacetatebromodichloroacetate, bromodichloroacetatebromoacetate, bromoacetatedibromoacetate, dibromoacetatechloroacetate, chloroacetatetribromoacetate, tribromoacetatetrichloroacetate, trichloroacetatedalapon, dalapondichloroacetate, dichloroacetateester
Agilent J&W GC Columns For Environmental Applications
Agilent J&W GC Columns For Environmental Applications
2012|Agilent Technologies|Brochures and specifications
Agilent J&W GC Columns For Environmental Applications MEASURE TRACE-LEVEL CONTAMINANTS AND MEET REGULATORY REQUIREMENTS WITH CONFIDENCE Reliably and efficiently analyze increasingly small quantities of active solutes Every day, you are on the front lines in the battle to safeguard our…
Key words
identification, identificationbenzo, benzobhc, bhcpeak, peakendrin, endrinagilent, agilentpesticides, pesticidescolumns, columnsendosulfan, endosulfanmin, minheptachlor, heptachlorxylene, xyleneinert, inertcolumn, columnultra
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike