The Triple-Axis Detector: Attributes andOperating Advice

Technical notes | 2008 | Agilent TechnologiesInstrumentation
GC/MSD
Industries
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


The development of advanced detector geometries in gas chromatography–mass spectrometry is critical for improving sensitivity, reducing noise and lowering detection limits. The Triple-Axis Detector (TAD) represents a significant step forward, offering enhanced performance for trace analysis and quantitative applications.

Objectives and Overview


The study aimed to introduce the new TAD design, highlight its key operating features and provide practical advice for optimal use. Key objectives included:
  • Characterizing signal intensity and noise reduction relative to the standard detector
  • Assessing continuous performance in gain-normalized tuning methods
  • Outlining procedures to maximize detector longevity

Methodology and Instrumentation


The evaluation employed gain-normalized tuning using PFTBA calibrant gas and comparisons of signal profiles for chlorobiphenyl standards. Instrument response and noise levels were measured under identical acquisition conditions to quantify improvements.

Used Instrumentation


  • Triple-Axis Detector with off-axis geometry
  • Electron multiplier (EM) with optimized gain configuration
  • PFTBA calibrant gas for tuning
  • EM Saver function in selected-ion monitoring (SIM) methods

Main Results and Discussion


The TAD demonstrated approximately two-fold higher absolute ion counts at constant gain and up to seven-fold greater signal-to-noise ratio due to reduced baseline noise. Calibration curves for trichlorobiphenyl exhibited doubled slopes, indicating improved sensitivity and extended detection limits into the femtogram range. Lower operating voltages and mass bias “tilt” toward lighter fragments were managed through gain-normalized tuning.

Benefits and Practical Applications


  • Enhanced detection limits for trace compounds in environmental and food analyses
  • Improved quantitation reproducibility at low concentration levels
  • Extended EM lifetime when operated with EM Saver and proper storage
  • Optimized performance in SIM workflows for extended acquisitions

Future Trends and Applications


Ongoing detector innovations may integrate real-time gain adjustments, further reduce mass bias and implement advanced digital detection schemes. The TAD platform lays the groundwork for next-generation MS analyses in QA/QC, metabolomics and trace contaminants monitoring.

Conclusion


The Triple-Axis Detector introduces significant gains in sensitivity and noise reduction for GC-MS applications. Proper use of gain-normalized tuning and EM Saver safeguards performance and device longevity, making the TAD a valuable tool for laboratories demanding high sensitivity and reliability.

References


  1. Prest H, Foote J, Kernan J, Peterson D. Enhancements to Gain Normalized Instrument Tuning. Agilent Technologies publication 5989-7654EN.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Enhancements to Gain Normalized Instrument Tuning: Understanding the Benefits and Features
Enhancements to Gain Normalized Instrument Tuning: Understanding the Benefits and Features Technical Overview Harry Prest, James D. Foote, Jeffrey T. Kernan, and Dave Peterson Introduction The automated optimization of ion detection in a mass spectrometer (MS), commonly referred to as…
Key words
gain, gainfactor, factornormalized, normalizedemv, emvaged, agedreic, reicatune, atunevoltage, voltagecalibrant, calibrantsaged, sagedsnew, snewnew, newtuning, tuningsignal, signalrtic
The 5975C Series MSDs: Normalized Instrument Tuning
The 5975C Series MSDs: Normalized Instrument Tuning
2007|Agilent Technologies|Technical notes
The 5975C Series MSDs: Normalized Instrument Tuning Technical Overview Jeffrey T. Kernan and Harry Prest Introduction The automated optimization of ion detection in a mass spectrometer, commonly referred to as “autotuning,” has been a tremendous simplification in MS operation. This…
Key words
gain, gaintuning, tuningvoltage, voltageemv, emvmenu, menuservicing, servicingadjust, adjustsame, samenormalization, normalizationgains, gainsnor, normultiplier, multipliersettings, settingsemvs, emvshighenergy
The 5975C Series MSDs: Method Optimization and Trace Ion Detection
The 5975C Series MSDs: Method Optimization and Trace Ion Detection Technical Overview Randy Roushall and Harry Prest Introduction Trace Ion Detection The new software control offered by the ChemStation Software (G1701EA) and hardware improvements of the 5975C MSD give the…
Key words
tid, tidpeak, peakion, iondetection, detectionbetter, bettercompound, compoundtrace, tracefour, fourshapes, shapesquantitation, quantitationheights, heightspeaks, peaksoff, offover, overeluting
Agilent 5975C TAD Series GC/MSD System - Data Sheet
Agilent 5975C TAD Series GC/MSD System - Data Sheet
2012|Agilent Technologies|Brochures and specifications
Agilent 5975C TAD Series GC/MSD System Data Sheet GC/MSD Overview The Agilent 5975C TAD Series Gas Chromatograph/Mass Selective Detector (GC/MSD) with the Triple-Axis High Energy Diode (HED) Electron Multiplier (EM) Detector provides the flexibility, capabilities, and performance demanded by modern…
Key words
sim, simoptional, optionalmsd, msdscan, scanrtl, rtlmass, massaxis, axissource, sourceionization, ionizationelectronic, electronicnoncondensing, noncondensingmethane, methanecontrol, controlsetup, setupmaintenance
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike