Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns
Applications | 2007 | Agilent TechnologiesInstrumentation
The rapid and reliable analysis of food, flavor, and fragrance samples is critical for quality control, product development, regulatory compliance, and safety assurance. High-efficiency capillary gas chromatography (GC) columns with reduced internal diameter offer the promise of significantly shorter runtimes while preserving or improving chromatographic resolution. Combining these columns with intelligent method‐translation software further streamlines method development, enabling laboratories to respond quickly to evolving analytical challenges.
This application study evaluates the performance gains achieved by replacing conventional 0.25 mm internal diameter GC columns with high-efficiency 0.18 mm internal diameter columns in the analysis of essential oils and fragrance samples. Specific aims:
All experiments were conducted on an Agilent 6890 GC coupled to a 5973 mass selective detector (MSD) with a split/splitless inlet. Two column chemistries were assessed:
A predefined wide-scope temperature program for essential oils and fragrances was translated using Agilent’s GC method translation software, ensuring phase ratio matching and consistent retention factor (k) across column dimensions. Carrier gases tested included helium at optimized linear velocities and hydrogen for maximum speed gains. Samples (spearmint and ylang-ylang oils) were diluted 40:1 in acetone prior to injection.
The switch from 0.25 mm to 0.18 mm columns reduced the overall analysis time by 35% under helium without compromising chromatographic resolution for critical peak pairs. For example, the final spearmint oil component eluted in 27.4 minutes on the standard column versus 17.7 minutes on the high-efficiency column. Employing hydrogen carrier gas on the 0.18 mm columns further accelerated analyses, achieving an overall 61% reduction in runtime compared to the original methods. Resolution values for close-eluting compounds remained essentially constant across all configurations, demonstrating that efficiency gains did not sacrifice separation quality.
Key advantages of high-efficiency GC columns and method translation:
These benefits translate into lower per-sample costs, faster release of quality‐critical products, and enhanced laboratory responsiveness.
Emerging directions and potential applications include:
High-efficiency 0.18 mm GC columns, when paired with method translation software, deliver dramatic reductions in analysis time for food and fragrance samples without loss of resolution. The optional use of hydrogen carrier gas unlocks further speed gains. This combination of hardware and software streamlines method development, boosts laboratory throughput, and supports quality assurance in the food, flavor, and fragrance industries.
GC/MSD, GC/SQ, GC columns, Consumables
IndustriesFood & Agriculture
ManufacturerAgilent Technologies
Summary
Importance of the Topic
The rapid and reliable analysis of food, flavor, and fragrance samples is critical for quality control, product development, regulatory compliance, and safety assurance. High-efficiency capillary gas chromatography (GC) columns with reduced internal diameter offer the promise of significantly shorter runtimes while preserving or improving chromatographic resolution. Combining these columns with intelligent method‐translation software further streamlines method development, enabling laboratories to respond quickly to evolving analytical challenges.
Objectives and Study Overview
This application study evaluates the performance gains achieved by replacing conventional 0.25 mm internal diameter GC columns with high-efficiency 0.18 mm internal diameter columns in the analysis of essential oils and fragrance samples. Specific aims:
- Compare elution times and resolution for standard and high-efficiency columns using helium carrier gas.
- Demonstrate additional speed improvements when substituting hydrogen as the carrier gas.
- Illustrate the utility of GC method translation software in transferring existing temperature programs to new column dimensions.
Methodology and Instrumentation
All experiments were conducted on an Agilent 6890 GC coupled to a 5973 mass selective detector (MSD) with a split/splitless inlet. Two column chemistries were assessed:
- Nonpolar phase: DB-1 columns (30 m × 0.25 mm × 0.25 µm versus 20 m × 0.18 mm × 0.18 µm)
- Polar phase: DB-WAX columns (30 m × 0.25 mm × 0.25 µm versus 20 m × 0.18 mm × 0.18 µm)
A predefined wide-scope temperature program for essential oils and fragrances was translated using Agilent’s GC method translation software, ensuring phase ratio matching and consistent retention factor (k) across column dimensions. Carrier gases tested included helium at optimized linear velocities and hydrogen for maximum speed gains. Samples (spearmint and ylang-ylang oils) were diluted 40:1 in acetone prior to injection.
Main Results and Discussion
The switch from 0.25 mm to 0.18 mm columns reduced the overall analysis time by 35% under helium without compromising chromatographic resolution for critical peak pairs. For example, the final spearmint oil component eluted in 27.4 minutes on the standard column versus 17.7 minutes on the high-efficiency column. Employing hydrogen carrier gas on the 0.18 mm columns further accelerated analyses, achieving an overall 61% reduction in runtime compared to the original methods. Resolution values for close-eluting compounds remained essentially constant across all configurations, demonstrating that efficiency gains did not sacrifice separation quality.
Benefits and Practical Applications
Key advantages of high-efficiency GC columns and method translation:
- Substantial runtime reduction increases sample throughput and instrument availability.
- Minimal method redevelopment required—temperature programs and carrier velocity adjustments are automated.
- Compatibility with existing laboratory helium supply; hydrogen optional for maximum speed.
- Maintained or improved resolution for complex flavor and fragrance matrices.
These benefits translate into lower per-sample costs, faster release of quality‐critical products, and enhanced laboratory responsiveness.
Future Trends and Opportunities
Emerging directions and potential applications include:
- Integration of ultra-fast GC methods with two-dimensional separations for trace‐level profiling.
- Expanding software-assisted method translation to new stationary phase chemistries and microbore columns.
- Automation of carrier gas switching and dynamic flow control for green analytical chemistry.
- Deployment in portable GC systems for on-site food and fragrance authenticity testing.
Conclusion
High-efficiency 0.18 mm GC columns, when paired with method translation software, deliver dramatic reductions in analysis time for food and fragrance samples without loss of resolution. The optional use of hydrogen carrier gas unlocks further speed gains. This combination of hardware and software streamlines method development, boosts laboratory throughput, and supports quality assurance in the food, flavor, and fragrance industries.
Reference
- D. Rood, The Practical Guide to the Care, Maintenance and Troubleshooting of Capillary Gas Chromatographic Systems, Huthig, Heidelberg, 1991.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Practical, Faster GC Applications with High Efficiency GC Columns and Method High-Efficiency GC Columns and Method Translation Software
2010|Agilent Technologies|Presentations
Practical, Faster GC Applications with High Efficiency GC Columns and Method High-Efficiency Translation Software Mark Sinnott Application Engineer July 28, 2010 Page 1 Questions to Ask What information do you need from your analysis? Do you have more baseline than…
Key words
selection, selectioncolumn, columndecrease, decreaselength, lengthdiameter, diameterresolution, resolutionefficiency, efficiencyxlb, xlbopgv, opgvthickness, thicknesselution, elutionstationary, stationaryfilm, filmvelocity, velocitytranslation
Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software
2008|Agilent Technologies|Presentations
Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software Ken Lynam Application Chemist March 4th, 2008 Pittcon 2008, New Orleans, LA Page 1 Presentation outline Background and terminology High Efficiency Capillary GC Columns Utility of Method Translation…
Key words
xlb, xlbtranslation, translationcarrier, carriermethod, methodresolution, resolutionoil, oillength, lengthefficiency, efficiencycolumn, columnthickness, thicknessdiameter, diameterfilm, filmspearmint, spearmintdecrease, decreasefish
Tips and Tricks of Faster GC Analysis
2009|Agilent Technologies|Presentations
Tips and Tricks of Faster GC Analysis Simon Jones Application Engineer E-seminar October 1, 2009 Page 1 Practical Fast GC Applications with Capillary GC Columns May 19, 2009 Questions to Ask What information do you need from your analysis? Do…
Key words
practical, practicalcapillary, capillaryfast, fastapplications, applicationscolumns, columnsdecrease, decreaselength, lengthresolution, resolutiondiameter, diametercolumn, columnthickness, thicknessefficiency, efficiencyfilm, filmxlb, xlbltm
Secrets of GC Column Dimensions
2010|Agilent Technologies|Presentations
Secrets of GC Column Dimensions GC Columns and Consumables Mark Sinnott Application Engineer Folsom California March 12, 2010 The Secrets of GC Column Dimensions Agilent Restricted Page 1 Secrets of GC Column Dimensions •Do I have the right column phase?…
Key words
secrets, secretsrestricted, restricteddimensions, dimensionscolumn, columnagilent, agilentthickness, thicknessfilm, filmmegabore, megaborediameter, diameterlength, lengthincrease, increaseresolution, resolutionefficiency, efficiencycapacity, capacityopgv