EPA 8270E with Pulsed Split Injection and Retention Time Locking on an 8890 GC with a 5977 Series MSD

Applications | 2020 | Agilent TechnologiesInstrumentation
GC/MSD, GC/SQ
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Significance of the topic


The reliable detection and quantitation of semivolatile organic compounds (SVOCs) in environmental samples are essential for regulatory compliance and public health. Method 8270E by US EPA sets stringent requirements for sensitivity, dynamic range, and system inertness when analyzing over 200 SVOCs. Modern GC/MS platforms, when optimized, can meet these demands while reducing sample carryover, maintenance frequency, and data processing time.

Study objectives and overview


This work demonstrates the optimization, calibration, and performance evaluation of EPA Method 8270E on an Agilent 8890 gas chromatograph coupled to a 5977 single-quadrupole mass spectrometer. Key features include pulsed split injection for rapid sample transfer, a high-temperature extractor lens, and Agilent’s retention time locking (RTL) capability to maintain stable retention behavior after column maintenance.

Methodology and instrumentation


System configuration and consumables were selected to maximize inertness and throughput. Critical methodological steps included:
  • Pulsed split injection at 280 °C with a 4:1 split and 30 psi injection pulse pressure for 0.6 minutes to compress vapor and reduce analyte residence time in the inlet.
  • A 9 mm extractor lens maintained at 300 °C to minimize surface adsorption of high-boiling analytes, such as phthalates and PAHs.
  • Autotune with decafluorotriphenylphosphine (DFTPP) to meet updated ion abundance criteria specified in EPA 8270E.
  • Retention time locking using acenaphthene-d10 to re-establish original retention times within 0.17 seconds after trimming 0.5 m of column.

Instrumentation


  • Agilent 8890 GC with split/splitless inlet and DB-UI 8270D column (30 m × 0.25 mm, 0.25 µm).
  • Agilent 5977A MSD with extractor EI source and 9 mm aperture lens.
  • Agilent 7693 automatic liquid sampler with 10 µL PTFE-tipped syringe.
  • MassHunter Workstation software for acquisition, RTL calibration, and quantitative analysis.

Main results and discussion


An initial 13-point calibration covering 0.2–160 µg/mL achieved R2 values ≥ 0.99 for over 90% of compounds. Relative response factor (RRF) precision averaged 6.4% RSD over the range, with method repeatability of ten replicate injections at 0.8 µg/mL yielding a mean RSD of 3.5%. Key observations included:
  • Excellent linear dynamic range for challenging analytes such as di-n-octyl phthalate and benzo[a]pyrene (100.0% and 96.9% average accuracy, respectively).
  • Tune evaluation confirmed all DFTPP ion abundance criteria and low tailing of probe compounds pentachlorophenol (TF = 1.0) and benzidine (TF = 0.7).
  • RTL maintained retention time consistency to within ±0.003 minutes across 75 target analytes after column trimming.

Benefits and practical applications


This optimized method configuration delivers:
  • Broad dynamic range with minimized detector saturation, reducing the need for multiple injections or re-analysis.
  • Enhanced system inertness for thermally labile and high-molecular-weight SVOCs, lowering maintenance frequency.
  • Retention time stability across maintenance events, streamlining data processing and compound identification.

Future trends and potential applications


Advances in GC/MS detection, such as tandem mass spectrometry and high-resolution accurate mass instruments, will further improve SVOC specificity and sensitivity. Automated method adjustment, digital retention libraries, and AI-driven data review are expected to enhance lab productivity and robustness in environmental monitoring and industrial QA/QC.

Conclusion


The combination of pulsed split injection, high-temperature extractor design, optimized detector gain, and retention time locking on the Agilent 8890/5977 platform meets or exceeds the quality control requirements of EPA Method 8270E. This approach offers a robust, high-throughput solution for reliable SVOC analysis.

References


  1. US EPA. Semivolatile Organic Compounds by GC/MS; Method 8270E. Revision 4, 2018.
  2. Agilent Technologies. EPA Method 8270 for SVOC Analysis on the 5977A Series GC/MSD. 2013.
  3. Agilent Technologies. EPA 8270 Re-optimized for Widest Calibration Range on the 5977 Inert Plus GC/MSD. 2018.
  4. Agilent Technologies. A Quick Start to Optimizing Detector Gain for GC/MSD. 2013.
  5. Agilent Technologies. Optimized GC/MS Analysis for PAHs in Challenging Matrices. 2019.
  6. Agilent Technologies. Retention Time Locking with the MSD Productivity ChemStation. 2008.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Semivolatile Organic Compounds with US EPA 8270E Using the Agilent 7000E Triple Quadrupole GC/MS
Application Note Analysis of Semivolatile Organic Compounds with US EPA 8270E Using the Agilent 7000E Triple Quadrupole GC/MS Authors Abstract Eric Fausett, Rachael Ciotti, and Dale Walker Agilent Technologies, Inc. This application note illustrates a sensitive method used to analyze…
Key words
avg, avgphthalate, phthalatefluoranthene, fluoranthenebenzo, benzotune, tunedelta, deltatime, timeanthracene, anthracenecalibration, calibrationbenzyl, benzylsvocs, svocschecktune, checktunepyrene, pyrenecounts, countsrelative
Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry
Application Note Environmental Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis of…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenepass, passnitrobenzene, nitrobenzeneanthracene, anthracenelinear, lineararamite, aramitemethanesulfonate, methanesulfonatediallate, diallatepyrene, pyreneazobenzene, azobenzenebenzyl, benzylnitroso, nitrosohydroinert
Analysis of Semivolatile Organic Compounds with Hydrogen Carrier Gas and HydroInert Source by Gas Chromatography/Triple Quadrupole Mass Spectrometry (GC/MS/MS)
Application Note Environmental Analysis of Semivolatile Organic Compounds with Hydrogen Carrier Gas and HydroInert Source by Gas Chromatography/Triple Quadrupole Mass Spectrometry (GC/MS/MS) Author Abstract Angela Smith Henry, PhD Agilent Technologies, Inc. Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenequadratic, quadraticpass, passaramite, aramitemethanesulfonate, methanesulfonatenitrosodi, nitrosodianthracene, anthracenepyrene, pyrenebenzyl, benzylbutyl, butylparathion, parathionnitroso, nitrosohydroinert
Optimized analysis of semi-volatile organic compounds (SVOC) in environmental samples in compliance with U.S. EPA Method 8270 utilizing GC-MS
Application note | 000761 Mass spectrometry Optimized analysis of semi-volatile organic compounds (SVOC) in environmental samples in compliance with U.S. EPA Method 8270 utilizing GC-MS Introduction Authors Semi-volatile organic compounds (SVOCs) are ubiquitous in our environment and raise Amit Gujar…
Key words
avgcalfact, avgcalfactphthalate, phthalatebenzo, benzofluoranthene, fluoranthenepass, passfactor, factorbenzyl, benzylresponse, responseanthracene, anthracenepyrene, pyreneminimum, minimumname, namefit, fitdiethylphthalate, diethylphthalatedioctyl
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike