GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

A Total Solution for the Analysis of Melamine and Cyanuric Acid in Pet Food by GC/MS and Aqueous Normal-Phase LC/MS/MS

Applications | 2007 | Agilent TechnologiesInstrumentation
GC/MSD, GC/SQ, LC/MS, LC/MS/MS, LC/QQQ
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Melamine and cyanuric acid adulteration in pet food has led to severe toxicity incidents in animals, highlighting the need for reliable analytical techniques capable of detecting these compounds at trace levels in complex matrices.

Study Objectives and Overview


This work compares two analytical workflows: a GC/MS method employing trimethylsilyl derivatization for screening and quantitation, and an LC/MS/MS approach using aqueous normal-phase separation for simultaneous confirmation and quantitation with streamlined sample preparation.

Methodology and Sample Preparation


  • GC/MS: Pet food samples are extracted with diethylamine, water, and acetonitrile, followed by derivatization with BSTFA and pyridine at 70 °C. Analysis is performed on an Agilent 7890A GC coupled to a 5975C MSD in SIM/scan mode.
  • LC/MS/MS: Ground samples are extracted with 50:50 acetonitrile:water, centrifuged, and filtered. The extract is injected onto a ZORBAX Rx-Sil column under isocratic conditions (95% acetonitrile with 5 mM ammonium acetate) and detected on an Agilent 1200SL-6410 triple quadrupole in MRM mode, switching from negative to positive polarity for cyanuric acid and melamine, respectively.

Used Instrumentation


  • GC/MS: Agilent 7890A GC, 5975C MSD, DB-5MS column, electron impact ionization, SIM/scan detection.
  • LC/MS/MS: Agilent 1200SL Rapid Resolution LC, 6410 Triple Quad MS, ZORBAX Rx-Sil column, electrospray ionization in both polarities, MRM acquisition.

Main Results and Discussion


The GC/MS SIM protocol achieved a screening detection limit of 10 µg/g and demonstrated improved selectivity by choosing characteristic ions. The LC/MS/MS workflow exhibited linear calibration from 50 pg/mL to 50 ng/mL for melamine and 1 ng/mL to 100 ng/mL for cyanuric acid, with detection limits of 50 pg/mL and 1 ng/mL, respectively. Both methods provided baseline separation in under 5 minutes and effective quantitation in spiked wheat gluten and pet food samples.

Benefits and Practical Applications


  • The GC/MS method serves as a rapid screening tool adaptable for confirmation and quantitation when moderate sensitivity suffices.
  • The LC/MS/MS technique eliminates derivatization, reduces sample preparation time, and extends the dynamic range, making it ideal for trace analysis in biological fluids and higher levels in food matrices.
  • Both approaches support quality control, regulatory compliance, and routine monitoring of pet food safety.

Future Trends and Applications


Advancements in HILIC column technology, microflow electrospray interfaces, and high-resolution mass spectrometry will further improve sensitivity, reduce solvent usage, and enable automated workflows. These enhancements are expected to broaden the application of these techniques to other polar contaminants in food and biological samples.

Conclusion


Agilent GC/MS and LC/MS/MS methods provide complementary analytical solutions for melamine and cyanuric acid in complex food matrices. While GC/MS offers a robust screening platform, LC/MS/MS delivers superior sensitivity and a simplified workflow suited for high-throughput and trace-level analysis.

Reference


  1. FDA GC-MS Screening Method for Melamine and Related Compounds Version 2 May 2007
  2. K I Petrus Hemstrom Hydrophilic interaction chromatography Journal of Separation Science 2006 29(12) 1784-1821

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
GC/MS Analysis of Melamine and Cyanuric Acid Below 1 μg/g in Infant Formula
GC/MS Analysis of Melamine and Cyanuric Acid Below 1 μg/g in Infant Formula Abstract The establishment of a 1 μg/g safety threshold for melamine in infant foods has led to a need for sensitive methods. Here, we established GC/MS conditions…
Key words
melamine, melaminecyanuric, cyanuricammelide, ammelideammeline, ammelinebenzoguanamine, benzoguanamineacid, acidnylon, nyloninfant, infantmin, minderivatization, derivatizationratios, ratiosmatrix, matrixdeaths, deathsformula, formulaillnesses
Trace-Level Analysis of Melamine in Milk Products on Agilent 7890A/5975C GC/MSD Using a New Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges
Trace-Level Analysis of Melamine in Milk Products on Agilent 7890A/5975C GC/MSD Using a New Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application Note Gas Chromatography/Mass Spectrometry/Food Safety Authors Abstract Min Cai and Yun Zou A GC/MS method…
Key words
melamine, melaminemilk, milktms, tmssampliq, sampliqabundance, abundancescx, scxtri, trirecovery, recoverymode, modelevel, levellyzed, lyzedreagents, reagentsresin, resintrace, tracecai
Simultaneous Extraction of Melamine and Cyanuric Acid from Food Products Using Strata® Melamine SPE and Ultra-fast LC/MS/MS Analysis Using Kinetex™ HILIC, Rapid LC/MS/MS Analysis on Luna® HILIC, or Rapid GC/MS Analysis on Zebron™ ZB-XLB HT
TN-0021 APPLICATIONS Simultaneous Extraction of Melamine and Cyanuric Acid from Food Products Using Strata® Melamine SPE and Ultra-fast LC/MS/MS Analysis Using Kinetex™ HILIC, Rapid LC/MS/MS Analysis on Luna® HILIC, or Rapid GC/MS Analysis on Zebron™ ZB-XLB HT Shahana Wahab Huq,…
Key words
melamine, melaminecyanuric, cyanuricacid, acidzebron, zebronbaby, babyxlb, xlbinferno, infernoluna, lunakinetex, kinetexhilic, hilicstrata, strataformula, formulaextraction, extractioncomplete, completeextracts
Rapid Screening of Melamine and Cyanuric Acid in Milk Products Using Agilent J&W HP-5ms GC Column and Agilent 7890A/5975C GC/MSD with Column Backflushing
Rapid Screening of Melamine and Cyanuric Acid in Milk Products Using Agilent J&W HP-5ms GC Column and Agilent 7890A/5975C GC/MSD with Column Backflushing Application Note Food Safety Authors Abstract Min Cai, Chongtian Yu, Yun Zou, A rapid screening method for…