The On-line VOC Analysis of Industrial Effluents using Ambient Headspace

Applications |  | GL SciencesInstrumentation
GC, HeadSpace
Industries
Environmental
Manufacturer
Agilent Technologies, GL Sciences

Summary

Significance of the Topic


Ambient headspace on-line analysis provides a streamlined approach for monitoring volatile organic compounds in industrial effluents with minimal manual intervention. This method addresses regulatory and safety requirements by enabling rapid detection of concentration trends and potential releases.

Objectives and Overview of the Study


  • Establish an automated hourly sampling and analysis system capable of detecting sudden increases in VOC levels with detection limits around 0.5 ppm and acceptable RSDs for trend monitoring
  • Demonstrate the feasibility of room temperature ambient headspace equilibration for representative VOCs in effluent matrices
  • Validate method performance over extended operation periods with minimal maintenance

Methods and Instrumentation Used


A fully automated system was configured to extract 250 microliters of effluent via a flowcell connected to a Focus Robotic Sample Processor. Samples were placed in sealed 2 mL vials and headspace was allowed to equilibrate at ambient temperature. A hot split injection introduced the headspace into an Agilent 6890 GC with FID for analysis. Key hardware components:
  • Focus Robotic Sample Processor with Flowcell
  • ATAS Optic 2-200 Programmable Injector (Agilent split/splitless configuration)
  • Agilent 6890 Gas Chromatograph with Flame Ionization Detector

Main Results and Discussion


Calibration curves for toluene, ethyl benzene, o-xylene, m-xylene and p-xylene spiked into effluents from different days (0.4 to 4 ppm) showed linear response and reproducibility. Ambient temperature fluctuations had negligible impact on calibration (two-day checks yielded consistency within 1–4%). A representative chromatogram at 4 ppm resolved all five compounds with baseline separation. Continuous hourly monitoring over a 13-hour period captured real-time concentration trends and enabled prompt identification of sudden concentration rises.

Benefits and Practical Applications


  • True on-line operation reduces operator workload to periodic vial and wash reservoir replenishment
  • Room temperature headspace simplifies sample preparation and lowers thermal stress on components
  • Detection and precision meet requirements for regulatory monitoring and process control
  • System flexibility allows adaptation to diverse effluent sources and target analytes

Future Trends and Potential Applications


Emerging developments may integrate mass spectrometric detectors for enhanced compound identification, implement remote data access via IoT platforms, and apply machine learning for predictive trend analysis. Miniaturized sampling modules and automated maintenance alerts will further improve field deployability and operational reliability.

Conclusion


The presented ambient headspace on-line GC-FID method offers a robust, low-maintenance solution for continuous VOC monitoring in industrial effluents. It delivers reliable trend detection, acceptable sensitivity, and high throughput with minimal user involvement.

Reference


  • Stuart Johnson and Diane Nicholas Application Note No. 086 The On-line VOC Analysis of Industrial Effluents using Ambient Headspace Aventis Crop Science Norfolk UK

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Petrol Range Organics In Soils, Sludges And Wastewaters by Headspace GC-MS
Application Note No. 093 Petrol Range Organics In Soils, Sludges And Wastewaters by Headspace GC-MS Steve Gardner, Alcontrol Technichem, Uxbridge, Middlesex UK Introduction Most PRO methods quantify the petrol range organics by indiscriminately integrating the area between two points, this…
Key words
petrol, petrolether, ethermean, meanlod, lodtert, tertrsd, rsdfocus, focuscrimp, crimpnonane, nonanedecane, decaneoctane, octaneheptane, heptaneeradication, eradicationisopropyl, isopropyltray
EPA TO-15 Analysis Using Hydrogen Carrier Gas and the Agilent HydroInert Source
Application Note Environmental EPA TO-15 Analysis Using Hydrogen Carrier Gas and the Agilent HydroInert Source Ambient air testing using cryogen-free thermal desorption and gas chromatography coupled to a single quadrupole mass spectrometer (GC/MS) with hydrogen gas Authors Abstract Laura Miles,…
Key words
hydrogen, hydrogenrrf, rrfhydroinert, hydroinertmethyl, methylkori, koricanister, canisterair, airnaphthalene, naphthalenepropene, propenexylene, xylenevinyl, vinylketone, ketonecia, ciaisopropanol, isopropanoldichlorodifluoromethane
Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15
Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15 Application Note Environmental, Canister, TO-15, Air Monitoring, Air Toxics Abstract This application note describes the GC/MS analysis of humidified canister air toxics samples at various…
Key words
kori, koricanister, canisterabundance, abundanceair, aircia, ciarrf, rrfretention, retentioncanisters, canisterssampling, samplingmin, minxylene, xylenevinyl, vinylnaphthalene, naphthaleneppbv, ppbvpropene
Improvements to Ambient Air Monitoring (U.S. EPA PAMS) Using a Clarus 690 Gas Chromatograph
APPLICATION Gas Chromatography Author: Miles Snow PerkinElmer, Inc. Shelton, CT Improvements to Ambient Air Monitoring (U.S. EPA PAMS) Using a Clarus 690 Gas Chromatograph Introduction In the United States, the Clean Air Act of 1970 gave the U.S. Environmental Protection…
Key words
ethyltoluene, ethyltolueneppbc, ppbcdiethylbenzene, diethylbenzenexylene, xylenetrap, trapair, aircompounds, compoundszone, zonepams, pamsozone, ozonelri, lrioctane, octanehexane, hexanepropylbenzene, propylbenzeneheptane
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike