The Analysis of Acid Herbicides by RLVI/GC/MS
Applications | | GL SciencesInstrumentation
Acid herbicides such as MCPP, Dicamba, MCPA and 2,4-D have been widely used since the 1940s to control broadleaf weeds. Their water solubility and tendency to form emulsions or readily leach into soil and groundwater raise environmental and health concerns. Sensitive analytical methods are required to monitor trace residues with minimal sample preparation.
Nine replicate 100 μL injections of pentafluorobenzylated herbicides yielded RSD values below 6% for most compounds, demonstrating repeatability. MCPA and 2,4,5-T showed RSDs of 3.37% and 2.82%, respectively, while Bromoxynil reached 10.18% RSD. The total ion chromatogram achieved baseline separation of all analytes within a 16 min runtime.
The RLVI/GC/MS method using the ATAS Optic 2-200 injector significantly improves the analysis of acid herbicides by lowering detection limits and simplifying sample preparation while maintaining precise quantitation, making it suitable for routine environmental monitoring.
GC/MSD
IndustriesFood & Agriculture
ManufacturerGL Sciences
Summary
Importance of the Topic
Acid herbicides such as MCPP, Dicamba, MCPA and 2,4-D have been widely used since the 1940s to control broadleaf weeds. Their water solubility and tendency to form emulsions or readily leach into soil and groundwater raise environmental and health concerns. Sensitive analytical methods are required to monitor trace residues with minimal sample preparation.
Objectives and Study Overview
- Establish a large volume injection method for acid herbicide analysis by RLVI/GC/MS
- Reduce detection limits and sample preparation steps
- Evaluate performance of the ATAS Optic 2-200 injector with an ‘A’ Type liner
Methodology and Instrumentation
Used Instrumentation
- ATAS Optic 2-200 programmable large volume injector
- Varian Star GC/MS system
Method Conditions
- Injection: 100 μL of pentafluorobenzyl derivatives in iso-octane via ATAS ‘A’ Type liner
- Injector program: 55 °C initial, 4 °C/s to 280 °C, split flow 50 mL/min, vent flow 150 mL/min, vent time 35 s
- GC column: HP-5MS 30 m × 0.25 mm × 0.25 μm; oven: 90 °C initial, 18 °C/min to 260 °C, hold 8 min
- MS: EI scan mode, 50–560 m/z range, 0.5 s scan time, 16 min segment
Main Results and Discussion
Nine replicate 100 μL injections of pentafluorobenzylated herbicides yielded RSD values below 6% for most compounds, demonstrating repeatability. MCPA and 2,4,5-T showed RSDs of 3.37% and 2.82%, respectively, while Bromoxynil reached 10.18% RSD. The total ion chromatogram achieved baseline separation of all analytes within a 16 min runtime.
Benefits and Practical Application
- Enhanced sensitivity and lower detection limits for environmental monitoring
- Minimal sample cleanup through direct large volume injection
- High throughput for routine QC/QA testing of water and soil extracts
- Robust performance and reduced carryover with the ATAS ‘A’ Type liner
Future Trends and Possibilities
- Automation of RLVI for even higher throughput
- Application to a wider range of pesticide and pollutant classes
- Coupling with high-resolution MS for comprehensive screening
- Development of portable GC/MS systems for on-site environmental analysis
Conclusion
The RLVI/GC/MS method using the ATAS Optic 2-200 injector significantly improves the analysis of acid herbicides by lowering detection limits and simplifying sample preparation while maintaining precise quantitation, making it suitable for routine environmental monitoring.
References
- GL Sciences B.V., Application Note No. 071, The Analysis of Acid Herbicides by RLVI/GC/MS.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
The Large Volume Aqueous Injection of a Parent Amino Acid Extract
|Agilent Technologies|Applications
Application Note No. 083 The Large Volume Aqueous Injection of a Parent Amino Acid Extract Diane Nicholas Monitors breakdown products of sample at high temperature Removes need for pre-concentration of water extract No derivatisation steps needed prior…
Key words
breakdown, breakdownparent, parentamino, aminoextract, extractdiane, dianenicholas, nicholasacid, acidwater, waterderivatisation, derivatisationlarge, largemonitors, monitorsprinciples, principlesproducts, productsappendix, appendixremoves
Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface
|Agilent Technologies|Applications
Application Note No. 033 Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface Ryoichi Sasano*, Takayuki Hamada, Masahiro Furuno GL Sciences, 237-2 Sayamagahara, Iruma, Saitama, 358-0032 Japan Key Words: On-line SPE-GC;…
Key words
concentrating, concentratingprospekt, prospektinj, injtemp, tempspe, speector, ectormidas, midasctor, ctorline, linemethod, methodholland, hollandnol, nolpressure, pressureconnected, connectedliner
Fast Analysis of Chlorinated Priority Pollutants using Large Volume Injection
|GL Sciences|Applications
Application Note No. 067 Fast Analysis of Chlorinated Priority Pollutants using Large Volume Injection Dr Roman Hirsch, LECO Instrumente GmbH, Moenchengladbach, Germany Introduction The determination of chlorinated compounds such as PCBs, dioxins, etc., has always been of interest in environmental…
Key words
acceleration, accelerationpriority, prioritychlorinated, chlorinatedacknowledgement, acknowledgementpollutants, pollutantsvarying, varyinginjection, injectionmaintaining, maintainingauto, autoallowing, allowingvolumes, volumespowerful, powerfulreal, realanalyses, analysesinstrumentation
The Analysis of High Molecular Weight Polymer Additives Using the Large Volume AT-Column Concentrating Technique
|Agilent Technologies|Applications
Application Note No. 057 The Analysis of High Molecular Weight Polymer Additives Using the Large Volume AT-Column Concentrating Technique Diane Nicholas. Introduction The analysis of high molecular weight compounds by GC is often very difficult. There are many opportunities for…
Key words
volts, voltsconcentrating, concentratinglarge, largeweight, weightmolecular, molecularvolume, volumeadditives, additivestechnique, techniquepolymer, polymerliner, linerautosampling, autosamplingatas, atascolumn, columnminutes, minutespeak