Micro GC Analysis of Permanent Gas Impurities in PEM Fuel Cell-Grade Hydrogen
Applications | 2025 | Agilent TechnologiesInstrumentation
This application note presents a rapid and robust analytical approach for monitoring trace permanent gas impurities in proton exchange membrane (PEM) fuel cell-grade hydrogen. Ensuring ultra-high purity hydrogen is essential for fuel cell performance, safety, and longevity in vehicular and stationary applications.
The primary objective was to establish and validate a micro gas chromatography method using the Agilent 990 Micro GC system to quantify helium, argon, oxygen, nitrogen, methane, and neon at low parts-per-million levels. The method was benchmarked against standards SAE J2719, ISO 14687, EN 17124, and ISO 19880-8, with the goal of achieving a total analysis cycle under five minutes to facilitate high-throughput quality control.
A single 20 m MS5A straight channel column operated at 30 °C and pressurized to 120 kPa was employed, with hydrogen as the carrier gas to mask the bulk hydrogen peak. A MEMS-based thermal conductivity detector enabled single-digit ppm detection. Calibration was performed using a single-point standard containing certified concentrations of He, Ar, O2, N2, CH4, and a balance of H2. Key operational conditions included a 50 °C injector, 200 ms injection time, and continuous-flow sampling.
Chromatographic separation achieved baseline resolution for the He/Ne pair and sufficient separation of O2 and Ar at low ppm levels. Method validation demonstrated recoveries of 98.6–100.1%, repeatability of 0.1–0.4% for most gases (1.8% for low-level O2), and site precision up to 2.2% for oxygen. Method detection limits ranged from 0.3 to 1.8 ppm, all below the five ppm threshold specified for oxygen. The five-minute runtime allows 10–12 analyses per hour, supporting efficient quality control workflows.
Advancements in column chemistries and detector technologies may further lower detection limits below 1 ppm, enabling direct monitoring of ammonia, carbon monoxide, carbon dioxide, and formaldehyde without additional instrumentation. Integration with automated sampling and real-time data analytics could support continuous monitoring at hydrogen production and distribution sites.
The Agilent 990 Micro GC method provides a validated, high-throughput solution for quantifying permanent gas impurities in PEM fuel cell-grade hydrogen, meeting or exceeding international standards. Its rapid runtime and robust performance make it well suited for routine quality control in hydrogen fueling and production facilities.
GC
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Significance of the Topic
This application note presents a rapid and robust analytical approach for monitoring trace permanent gas impurities in proton exchange membrane (PEM) fuel cell-grade hydrogen. Ensuring ultra-high purity hydrogen is essential for fuel cell performance, safety, and longevity in vehicular and stationary applications.
Objectives and Study Overview
The primary objective was to establish and validate a micro gas chromatography method using the Agilent 990 Micro GC system to quantify helium, argon, oxygen, nitrogen, methane, and neon at low parts-per-million levels. The method was benchmarked against standards SAE J2719, ISO 14687, EN 17124, and ISO 19880-8, with the goal of achieving a total analysis cycle under five minutes to facilitate high-throughput quality control.
Methodology and Instrumentation
A single 20 m MS5A straight channel column operated at 30 °C and pressurized to 120 kPa was employed, with hydrogen as the carrier gas to mask the bulk hydrogen peak. A MEMS-based thermal conductivity detector enabled single-digit ppm detection. Calibration was performed using a single-point standard containing certified concentrations of He, Ar, O2, N2, CH4, and a balance of H2. Key operational conditions included a 50 °C injector, 200 ms injection time, and continuous-flow sampling.
Instrumentation Used
- Agilent 990 Micro GC system
- 20 m MS5A column, straight channel
- Hydrogen carrier gas
- MEMS-based thermal conductivity detector (µTCD)
Results and Discussion
Chromatographic separation achieved baseline resolution for the He/Ne pair and sufficient separation of O2 and Ar at low ppm levels. Method validation demonstrated recoveries of 98.6–100.1%, repeatability of 0.1–0.4% for most gases (1.8% for low-level O2), and site precision up to 2.2% for oxygen. Method detection limits ranged from 0.3 to 1.8 ppm, all below the five ppm threshold specified for oxygen. The five-minute runtime allows 10–12 analyses per hour, supporting efficient quality control workflows.
Benefits and Practical Applications
- Rapid analysis under five minutes enhances sample throughput.
- Low detection limits meet stringent fuel cell hydrogen standards.
- High accuracy and precision ensure reliable compliance testing.
- Compact µGC system simplifies integration in QA/QC laboratories.
Future Trends and Opportunities
Advancements in column chemistries and detector technologies may further lower detection limits below 1 ppm, enabling direct monitoring of ammonia, carbon monoxide, carbon dioxide, and formaldehyde without additional instrumentation. Integration with automated sampling and real-time data analytics could support continuous monitoring at hydrogen production and distribution sites.
Conclusion
The Agilent 990 Micro GC method provides a validated, high-throughput solution for quantifying permanent gas impurities in PEM fuel cell-grade hydrogen, meeting or exceeding international standards. Its rapid runtime and robust performance make it well suited for routine quality control in hydrogen fueling and production facilities.
References
- Bu T Hydrogen Impurity Analysis Using the Agilent 990 Micro GC Agilent Technologies application note 5994-2138EN 2020
- Bajja M van Loon R Permanent Gas Analysis – Separation of Argon and Oxygen on a MolSieve 5Å Column using the Agilent 490 Micro GC Agilent Technologies application note 5990-8700EN 2011
- van Loon R Permanent Gas Analysis – Separation of Helium Neon and Hydrogen on a MolSieve 5Å Column using the Agilent 490 Micro GC Agilent Technologies application note 5990-8527EN 2011
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Argon and Oxygen Analysis Using the Agilent 990 Micro GC
2020|Agilent Technologies|Applications
Application Brief Argon and Oxygen Analysis Using the Agilent 990 Micro GC Authors Tingting Bu and Fei Jiang Agilent Technologies, Inc. Introduction Argon (Ar) and oxygen (O2) detection is always a challenge for the permanent gas manufacturers in their gas…
Key words
oxygen, oxygenargon, argonneon, neonnitrogen, nitrogenbalance, balancehelium, heliumcomponent, componentmin, minbrief, brieftingting, tingtingfei, feicarrier, carrierclever, cleverchannel, channeljiang
Hydrogen Impurity Analysis Using the Agilent 990 Micro GC
2020|Agilent Technologies|Applications
Application Brief Hydrogen Impurity Analysis Using the Agilent 990 Micro GC Author Tingting Bu Agilent Technologies, Inc. Introduction As a clean energy source, the application of hydrogen (H2) in fuel cell cars is very promising. In Japan, many H2 energy…
Key words
neon, neonhydrogen, hydrogenmethane, methaneargon, argoncarbon, carbonoxygen, oxygenmonoxide, monoxidecomponent, componentdioxide, dioxidebalance, balancebrief, brieftingting, tingtingmin, mingases, gasesppm
Noble Gases Analysis Using the Agilent 990 Micro GC
2022|Agilent Technologies|Applications
Application Note Energy & Chemicals Noble Gases Analysis Using the Agilent 990 Micro GC Authors Fei Jiang and Tingting Bu Agilent Technologies, Inc. Abstract This application note describes the use of the Agilent 990 Micro GC for noble gases analysis,…
Key words
noble, noblegases, gasesneon, neonargon, argonoxygen, oxygenkrypton, kryptonhydrogen, hydrogenxenon, xenonnitrogen, nitrogenmv·s, mv·smethane, methanesmelting, smeltingtingting, tingtingfei, feilod
Helium, Argon, Nitrogen, and Hydrocarbon Impurity Analysis in Hydrogen Using an Agilent 8890 GC and TCD/FID System
2024|Agilent Technologies|Applications
Application Note Energy & Chemicals Helium, Argon, Nitrogen, and Hydrocarbon Impurity Analysis in Hydrogen Using an Agilent 8890 GC and TCD/FID System Author Zhang Jie Agilent Technologies (Shanghai) Co. Ltd. Abstract In this application note, the analysis of helium (He),…
Key words
hcs, hcshydrogen, hydrogenbalance, balancegas, gastcd, tcdmethane, methanefid, fidppm, ppmµmol, µmolimpurities, impuritiesargon, argonair, airalumina, aluminabutane, butaneethane