Drinking Water Analysis Conditions for USEPA Method 524.3 and 524.4
Applications | | EST AnalyticalInstrumentation
Environmental regulations require sensitive detection of volatile organic compounds in drinking water. USEPA Methods 524.3 and 524.4 outline protocols for purge and trap gas chromatography mass spectrometry analysis. Switching from helium to nitrogen as purge gas can reduce costs and ease supply chain constraints while maintaining method performance.
This study compares helium and nitrogen purge gases using updated USEPA methods. It evaluates method flexibility, sample throughput, and detection performance. The goals include assessing linearity, precision, accuracy, and response factors under identical purge and trap conditions.
The analysis employed an EST Analytical Encon Evolution purge and trap concentrator and Centurion WS autosampler coupled to a Shimadzu QP2010 SE GCMS. A Vocarb 3000 trap was used. Key purge and trap parameters included an 11 min purge time at 40 ml min ambient gas flow and a 1 min desorb at 260 C under 5 psi. The GC program ran from 45 C to 240 C with a 30 to 1 split injection. A seven point quadratic calibration from 0.5 ppb to 40 ppb with 5 ppb internal standard was implemented.
Both purge gases achieved excellent linearity with regression coefficients above 0.995. Compound response factors averaged slightly lower with nitrogen by approximately 7 percent. Precision at low and mid ranges showed recoveries within regulatory limits for both gases. Minor variations in response did not impact quantitation. The system features such as the eight port valve and foam sensor effectively managed moisture and foaming with nitrogen purge.
Replacing helium with high purity nitrogen as purge gas under USEPA Method 524.4 delivers comparable performance in terms of linearity, precision, and accuracy. The slight reduction in response factors is offset by operational benefits and reduced helium dependence.
GC/MSD, GC/SQ, Purge and Trap
IndustriesFood & Agriculture
ManufacturerEST Analytical, Restek, Shimadzu
Summary
Importance of the Topic
Environmental regulations require sensitive detection of volatile organic compounds in drinking water. USEPA Methods 524.3 and 524.4 outline protocols for purge and trap gas chromatography mass spectrometry analysis. Switching from helium to nitrogen as purge gas can reduce costs and ease supply chain constraints while maintaining method performance.
Objectives and Overview
This study compares helium and nitrogen purge gases using updated USEPA methods. It evaluates method flexibility, sample throughput, and detection performance. The goals include assessing linearity, precision, accuracy, and response factors under identical purge and trap conditions.
Methodology and Instrumentation
The analysis employed an EST Analytical Encon Evolution purge and trap concentrator and Centurion WS autosampler coupled to a Shimadzu QP2010 SE GCMS. A Vocarb 3000 trap was used. Key purge and trap parameters included an 11 min purge time at 40 ml min ambient gas flow and a 1 min desorb at 260 C under 5 psi. The GC program ran from 45 C to 240 C with a 30 to 1 split injection. A seven point quadratic calibration from 0.5 ppb to 40 ppb with 5 ppb internal standard was implemented.
Results and Discussion
Both purge gases achieved excellent linearity with regression coefficients above 0.995. Compound response factors averaged slightly lower with nitrogen by approximately 7 percent. Precision at low and mid ranges showed recoveries within regulatory limits for both gases. Minor variations in response did not impact quantitation. The system features such as the eight port valve and foam sensor effectively managed moisture and foaming with nitrogen purge.
Benefits and Practical Applications
- Maintains compliance with USEPA 524 protocols using nitrogen
- Reduces reliance on high purity helium and associated costs
- Shortens sample preparation through flexible purge and trap parameters
- Improves laboratory uptime with foam control and sample handling innovations
Future Trends and Applications
- Broader adoption of nitrogen purge in routine water analysis labs
- Integration of real time moisture and foam monitoring across platforms
- Further optimization of trap materials and cycle times
- Expanding application to trace analysis of emerging contaminants
Conclusion
Replacing helium with high purity nitrogen as purge gas under USEPA Method 524.4 delivers comparable performance in terms of linearity, precision, and accuracy. The slight reduction in response factors is offset by operational benefits and reduced helium dependence.
References
- USEPA Method 524.3 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography MS Version 1 0 June 2009
- USEPA Method 524.4 Draft Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography MS Using Nitrogen Purge Gas Version 1 September 2011
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Drinking Water Analysis Conditions for USEPA Method 524.3 and the Newly Proposed Method 524.4 
|Agilent Technologies|Applications
Drinking Water Analysis Conditions for USEPA Method 524.3 and the Newly Proposed Method 524.4 Anne Jurek Abstract: In June 2009, the United States Environmental Protection Agency (USEPA) promulgated a new drinking water method, 524.3. Due to advances in analytical instrumentation,…
Key words
ether, etherbutylbenzene, butylbenzenemethyl, methylbutyl, butylpurge, purgeregress, regressethyl, ethylamyl, amylchloride, chlorideencon, enconest, estquadratic, quadraticxylene, xylenedesorb, desorbprecision
Drinking Water Analysis Conditions for USEPA Method 524.3 and the Newly Proposed Method 524.4
|Agilent Technologies|Applications
JSB is an authorised partner of Drinking Water Analysis Conditions for USEPA Method 524.3 and the Newly Proposed Method 524.4 Anne Jurek Abstract: In June 2009, the United States…
Key words
ether, etherbutylbenzene, butylbenzenepurge, purgemethyl, methylbutyl, butylregress, regressethyl, ethylchloride, chlorideamyl, amylest, estencon, enconquadratic, quadraticprecision, precisionxylene, xylenedesorb
Shimadzu Guide to US EPA Methods 524.3 and 524.4 for Analysis of Volatile Organic Compounds in Drinking Water
2015|Shimadzu|Applications
Gas Chromatograph Mass Spectrometer Shimadzu Guide to US EPA Methods 524.3 and 524.4 for Analysis of Volatile Organic Compounds in Drinking Water No. GCMS-1502 Anne Jurek, EST Analytical ■ Abstract Due to advances in analytical instrumentation the United States Environmental…
Key words
pass, passether, etherppb, ppbpir, pirpurge, purgebutylbenzene, butylbenzenemethyl, methylethyl, ethylchloride, chlorideamyl, amylbutyl, butyltrap, traprun, runencon, enconcenturion
Analysis of Volatile Organic Compounds Using USEPA Method 524.3
|Agilent Technologies|Applications
Analysis of Volatile Organic Compounds Using USEPA Method 524.3 Eclipse 4760 Purge and Trap and the 4100 Autosampler Introduction Although analysis of VOCs by purge and trap is considered a mature technique, advances in methodology and refinement of instrumentation are…
Key words
tert, tertether, etherstd, stdbutylbenzene, butylbenzeneethyl, ethylmethyl, methylbutyl, butylpurge, purgeamyl, amylppb, ppbsoil, soilchloride, chloridesam, samchlorodifluoromethane, chlorodifluoromethanechlorodibromomethane